Repeated pupil size measurements in refractive surgery candidates.

Christoph Robl1, Ieva Sliesoraityte, Jost Hillenkamp, Philipp Prahs, Chris P Lohmann, Horst Helbig, Wolfgang A Herrmann.   

Abstract

PURPOSE: To evaluate the individual variation in pupil diameter measurements in refractive surgery candidates under standardized conditions over 2 testing sessions using a digital infrared pupillometer.
SETTING: Department of Ophthalmology, University Medical Center Regensburg, Regensburg, Germany.
METHODS: The pupil size in both eyes of consecutive refractive surgery candidates was measured under 2 illumination levels (scotopic at 0.03 lux, low mesopic at 0.82 lux) with a digital infrared pupillometer (Procyon P2000 SA). Before measurement, a standardized dark-adaptation protocol was used. Measurements were taken twice on separate occasions.
RESULTS: The study evaluated 318 eyes of 159 patients with a mean age of 36.3 years +/- 10.2 (SD). The mean interval between measurements was 53 +/- 62.8 days. Despite sufficient power (beta < 0.05), there was no statistically significant difference between 2 consecutive measurements under scotopic and low mesopic illumination.
CONCLUSION: Using a standardized dark-adaptation protocol, refractive surgery candidates showed no significant variation in scotopic or low mesopic pupil diameter over 2 testing sessions.

Entities:  

Mesh:

Year:  2009        PMID: 19969214     DOI: 10.1016/j.jcrs.2009.07.013

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  3 in total

1.  Ocular higher-order aberrations and mesopic pupil size in individuals screened for refractive surgery.

Authors:  Seyed Javad Hashemian; Mohammad Soleimani; Alireza Foroutan; Mahmood Joshaghani; Mohammad Jafar Ghaempanah; Mohammad Ebrahim Jafari; Mehdi Yaseri
Journal:  Int J Ophthalmol       Date:  2012-04-18       Impact factor: 1.779

2.  Can we measure mesopic pupil size with the cobalt blue light slit-lamp biomicroscopy method?

Authors:  Miguel J Maldonado; Alberto López-Miguel; David P Piñero; José R Juberías; Juan C Nieto; Jorge L Alió
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2012-01-08       Impact factor: 3.117

3.  Intrasession repeatability of pupil size measurements under different light levels provided by a multidiagnostic device in healthy eyes.

Authors:  David P Piñero; Dolores de Fez; Inmaculada Cabezos; Alberto López-Navarro; María T Caballero; Vicente J Camps
Journal:  BMC Ophthalmol       Date:  2020-08-31       Impact factor: 2.209

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.