| Literature DB >> 32676203 |
Pedro Tañá-Rivero1, Francisco Pastor-Pascual1, Marceliano Crespo1, José L Rodríguez-Prats1, José J Muñoz-Tomás1, Robert Montés-Micó1,2.
Abstract
PURPOSE: To assess the efficacy, safety, and predictability of the Visian Implantable Collamer Lens (ICL) model having a central port in patients over 40 years of age.Entities:
Year: 2020 PMID: 32676203 PMCID: PMC7341401 DOI: 10.1155/2020/7457902
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Preoperative patient demographics and ICL characteristics.
| Mean ± SD | Range [min, max] | |
|---|---|---|
| Age (years) | 43.52 ± 4.49 | [40, 56] |
| Refraction sphere (D) | −8.18 ± 2.66 | [−13.5, −2] |
| Refraction cylinder (D) | −1.55 ± 1.18 | [−4.0, 0] |
| Spherical equivalent (D) | −8.46 ± 3.31 | [−14.88, −1.88] |
| Corneal thickness ( | 535.18 ± 36.66 | [452, 580] |
| CDVA (decimal) | 0.88 ± 0.18 | [0.40, 1.00] |
| ACD (mm) | 3.07 ± 0.26 | [2.67, 3.59] |
| WTW (mm) | 12.16 ± 0.41 | [11.30, 12.70] |
| ATA (mm) | 12.07 ± 0.39 | [11.50, 12.79] |
| ECD (cells/mm2) | 2568 ± 332 | [2004, 3297] |
| IOP (mmHg) | 16.23 ± 3.61 | [ |
| ICL sphere (D) | −9.52 ± 2.65 | [−15.0, −1.0] |
| ICL cylinder (D) | 2.65 ± 1.23 | [1, 4.5] |
| ICL size (mm) | 12.96 ± 0.35 | [12.10, 13.20] |
ICL: implantable Collamer lens; SD: standard deviation; D: diopters; CDVA: corrected-distance visual acuity; ACD: anterior chamber depth; WTW: white-to-white; ATA: angle-to-angle; ECD: endothelial cell density; IOP: intraocular pressure.
Figure 1Cumulative uncorrected distance visual acuity (UDVA), at 1 year after surgery (a), cumulative corrected distance visual acuity (CDVA) before and at 1 year after surgery (b), and postoperative changes in corrected distance visual acuity (CDVA) at 1 year after surgery (c).
Figure 2Postoperative spherical equivalent accuracy (a), attempted-versus-achieved spherical equivalent (b), and scatterplot of preoperative and 1-year postoperative astigmatic vectors (c). Solid line in figure B represents the best-fit line (regression equation and R value was included) and dotted lines ±1.00D. In the third figure, the more central location of postoperative data around 0 represents a reduction in astigmatism after lens implantation (J0 = Jackson cross-cylinder, axes at 180° and 90°; J45 = Jackson cross-cylinder, axes at 45° and 135°).
Figure 3Postoperative changes in intraocular pressure (IOP), measured in mmHg, at 1 year after surgery (a), and distribution of eyes according to the vault, measured in microns, at 1 year after surgery (b).
Studies reporting data for the V4c ICL with 1 year of follow-up: year of publication, sample age, and preoperative spherical equivalent (SE) (mean, standard deviation, and range).
| Author | Year | Eyes (patients) | Age (ys) | SE (D) |
|---|---|---|---|---|
| Lisa et al. [ | 2015 | 147 (80) | 30.4 ± 4.8 (20 to 40) | −8.80 ± 2.60 (−2.75 to −17.50 sph 0 to −3.00 cyl) |
| Karandikar et al. [ | 2015 | 34 (34) | 26.13 ± 3.8 (NR) | −9.24 ± 2.4 (NR) |
| Ganesh et al. [ | 2017 | NR (NR) | 26.4 ± 2.4 (NR) | −5.98 ± 1.15 (NR) |
| Pjano et al. [ | 2017 | 28 (16) | 28.21 ± 4.06 (21 to 35) | −9.52 ± 3.69 (NR) |
| Kamiya et al. [ | 2018 | |||
| Low-to-moderate myopia | 57 (57) | 34.8 ± 7.4 (20 to 57) | −4.29 ± 1.31 (−0.50 to −5.88) | |
| High myopia | 294 (294) | 33.6 ± 7.3 (18 to 54) | −10.13 ± 2.64 (−6.00 to −18.63) | |
| García-de la Rosa et al. [ | 2018 | 76 (42) | 27.4 ± 5.14 (20 to 39) | −11.94 ± 3.51 (−7.50 to −22.88) |
| Niu et al. [ | 2019 | 40 (31) | 32.45 ± 6.85 (20 to 42) | −14.03 ± 4.46 (−7.50 to −25.75) |
| Sachdev et al. [ | 2020 | 203 (NR) | 24 ± NR (22 to 28) | −9.44 ± 4.20 (−1.75 to −21) |
| Current study | 2020 | 33 (21) | 43.52 ± 4.49 (40 to 56) | −8.46 ± 3.31 (−1.88 to −14.88) |
Visual and refractive outcomes, intraocular pressure (IOP), endothelial cell density (ECD), and vault of the different studies considered (mean, standard deviation, and range) NR: not reported.
| Author | UCVA (logMAR) | CDVA (logMAR) | Safety index | Efficacy index | ±0.50D (% eyes) | ±1.00D (% eyes) | SE (D) | IOP (mmHg) (eyes >21) | ECD (cells/m2) (% loss) | Vault ( |
|---|---|---|---|---|---|---|---|---|---|---|
| Lisa et al. [ | 0.028 ± 0.055 | 0.003 ± 0.013 | 1.04 | 1.00 | 93.9 | 100 | −0.14 ± 0.26 | 12.4 ± 1.4 (0 > 20) | 2650 ± 438 (1.7) | 405.5 ± 184.7 (100 to 980) |
| Karandikar et al. [ | NR | NR | 1.15 | 1.6 | 57.12 | 98.12 | −0.19 ± 1.18 | 19.1 ± 1.3 | NR (7.1) | 628.2 ± 300.1 (NR) |
| Ganesh et al. [ | −0.022 ± 0.021 | −0.071 ± 0.079 | 1.24 | 1.12 | 90 | 100 | −0.164 ± 0.20 | NR | 2808 ± 315 (9.0) | NR |
| Pjano et al. [ | 0.76 ± 0.16 (decimal) | 0.79 ± 0.14 (decimal) | 1.25 | 1.2 | NR | NR | −0.21 ± 0.27 | 14.96 ± 1.7 | 2512 ± 127 (5.5) | NR |
| Kamiya et al. [ | ||||||||||
| Low-to-moderate myopia | −0.17 ± 0.14 | −0.21 ± 0.10 | NR | NR | 93 | 98 | −0.12 | 13.1 (0) | NR (0.1) | NR |
| High myopia | −0.16 ± 0.09 | −0.21 ± 0.08 | NR | NR | 94 | 99 | 0.02 | 13.6 (0) | NR (0.1) | NR |
| García-de la Rosa et al. [ | 0.12 ± 0.12 | 0.05 ± 0.08 | NR | NR | 85 | 86 | −0.06 ± 0.77 | NR | NR | 449 ± 180 (NR) |
| Niu et al. [ | 0.89 ± 0.30 (decimal) | 1.00 ± 0.27 (decimal) | 1.33 | 1.14 | 69 | 92 | −0.67 ± 1.29 | 15.15 ± 2.57 (0) | 2963 ± 396 (8.38) | 380 ± 152 (90 to 700) |
| Sachdev et al. [ | 0 ± NR | 0 ± NR | NR | NR | 94.09 | 96.09 | −0.12 ± NR | NR | NR | NR |
| Current study | 0.88 ± 0.16 (decimal) | 0.96 ± 0.09 (decimal) | 1.09 | 1.00 | 90.6 | 93.9 | −0.09 ± 0.47 | 15.27 ± 3.03 (0) | 2516 ± 234 (2.04) | 320 ± 136 (96 to 603) |
UCVA: uncorrected distance visual acuity, CDVA: corrected distance visual acuity, SE: spherical equivalent, and NR: not reported.