Literature DB >> 30855094

Five-Year Follow-up of Correction of Myopia: Posterior Chamber Phakic Intraocular Lens With a Central Port Design.

José F Alfonso, Luis Fernández-Vega-Cueto, Belén Alfonso-Bartolozzi, Robert Montés-Micó, Luis Fernández-Vega.   

Abstract

PURPOSE: To assess the long-term correction of moderate to high myopia using a posterior chamber phakic intraocular lens with a central port design.
METHODS: Uncorrected (UDVA) and corrected (CDVA) distance visual acuities, refraction, intraocular pressure (IOP), endothelial cell density (ECD), vault, and adverse events over a 5-year period were evaluated retrospectively.
RESULTS: A total of 147 eyes (83 patients) were evaluated. Mean postoperative UDVA and CDVA were 0.05 ± 0.11 and 0.02 ± 0.08 logMAR at 1 year and 0.13 ± 0.18 and 0.02 ± 0.09 logMAR at 5 years, respectively. More than 95% of eyes achieved CDVA of 20/25 or better at both follow-up periods. CDVA was unchanged or improved from one to three or more lines in all eyes. Preoperatively, mean spherical equivalent (SE) was -9.20 ± 3.02 diopters (D). At 1 year, the mean SE was -0.17 ± 0.26 D, with 91.53% of eyes within ±0.50 D and 100% of eyes within ±1.00 D of the target. At 5 years, the mean SE was -0.44 ± 0.47 D, with 67.4% of eyes within ±0.50 D and 90.1% of eyes within ±1.00 D of the target. Mean IOP was 12.74 ± 1.65 and 13.0 ± 2.03 mm Hg, at 1 and 5 years, respectively. No significant rise in IOP (> 20 mm Hg) occurred during the follow-up period. Mean ECD was 2,696 ± 358 and 2,645 ± 359 cells/mm2 at 1 and 5 years, respectively, representing a non-significant loss of 0.43% from preoperative values (P = .304). Mean vault changed significantly from 398 ± 187 µm at 1 year to 340 ± 163 µm at 5 years (P < .001). No intraoperative or postoperative complications or adverse events occurred during the follow-up period.
CONCLUSIONS: The good long-term outcomes found in this study support the use of this lens with a central port design for the correction of moderate to high myopia. [J Refract Surg. 2019;35(3):169-176.]. Copyright 2019, SLACK Incorporated.

Entities:  

Mesh:

Year:  2019        PMID: 30855094     DOI: 10.3928/1081597X-20190118-01

Source DB:  PubMed          Journal:  J Refract Surg        ISSN: 1081-597X            Impact factor:   3.573


  16 in total

1.  Long-term endothelial cell loss with the iris-claw intraocular phakic lenses (Artisan®).

Authors:  Virgilio Galvis; John F Villamil; María Fernanda Acuña; Paul A Camacho; Jesús Merayo-Lloves; Alejandro Tello; Sandra Lizeth Zambrano; Juan José Rey; Juan Vicente Espinoza; Angélica María Prada
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2019-10-28       Impact factor: 3.117

Review 2.  Analysis of perioperative problems related to intraocular Implantable Collamer Lens (ICL) implantation.

Authors:  Hao Zhang; Rui Gong; Xiaolan Zhang; Yingping Deng
Journal:  Int Ophthalmol       Date:  2022-06-22       Impact factor: 2.029

3.  Inter-eye and postoperative prediction of vault after implantation of EVO + Visian phakic implantable collamer lens.

Authors:  Elena Martínez-Plaza; Alberto López-Miguel; Alberto López-de la Rosa; Miguel J Maldonado
Journal:  Int Ophthalmol       Date:  2022-10-19       Impact factor: 2.029

4.  Posterior-Chamber Phakic Intraocular Lens Implantation in Patients over 40 Years of Age.

Authors:  Pedro Tañá-Rivero; Francisco Pastor-Pascual; Marceliano Crespo; José L Rodríguez-Prats; José J Muñoz-Tomás; Robert Montés-Micó
Journal:  J Ophthalmol       Date:  2020-06-29       Impact factor: 1.909

5.  Safety and Efficacy of a New Posterior Chamber Phakic Intraocular Lens in Cases of High Myopia: Early Results.

Authors:  Praveen Subudhi; Sweta Patro; Prateek Agarwal; Zahiruddin Khan; B Nageswar Rao Subudhi; Chandan Mekap; Asutosh Padhi
Journal:  Clin Ophthalmol       Date:  2020-10-30

6.  Short-term changes in and preoperative factors affecting vaulting after posterior chamber phakic Implantable Collamer Lens implantation.

Authors:  Qiu-Jian Zhu; Wen-Jing Chen; Wei-Jian Zhu; Hai-Xiang Xiao; Man-Hui Zhu; Lie Ma; You Yuan; E Song
Journal:  BMC Ophthalmol       Date:  2021-05-06       Impact factor: 2.209

Review 7.  Phakic intraocular lenses: An overview.

Authors:  Soraya M R Jonker; Tos T J M Berendschot; Isabelle E Y Saelens; Noël J C Bauer; Rudy M M A Nuijts
Journal:  Indian J Ophthalmol       Date:  2020-12       Impact factor: 1.848

8.  Eight-Year Outcomes of Implantation of Posterior Chamber Phakic Intraocular Lens With a Central Port for Moderate to High Ametropia.

Authors:  Kazutaka Kamiya; Kimiya Shimizu; Masahide Takahashi; Wakako Ando; Hideki Hayakawa; Nobuyuki Shoji
Journal:  Front Med (Lausanne)       Date:  2021-12-16

9.  Correlation between intraoperative and postoperative vaulting of the EVO implantable Collamer lens: a retrospective study of real-time observations of vaulting using the RESCAN 700 system.

Authors:  Nian Guan; Xiao-Nong Zhang; Wan-Jun Zhang
Journal:  BMC Ophthalmol       Date:  2022-01-03       Impact factor: 2.209

10.  Atomic force microscopy comparative analysis of the surface roughness of two posterior chamber phakic intraocular lens models: ICL versus IPCL.

Authors:  Juan Gros-Otero; Samira Ketabi; Rafael Cañones-Zafra; Montserrat Garcia-Gonzalez; Cesar Villa-Collar; Santiago Casado; Miguel A Teus
Journal:  BMC Ophthalmol       Date:  2021-07-14       Impact factor: 2.209

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.