Abdulaziz Ibrahim Alrashed1, Abdulrahman M Alfuraih2.
Abstract
PURPOSE: This study was aimed to investigate the reproducibility of shear wave elastography (SWE) among operators, machines, and probes in a phantom, and to evaluate the effect of depth of the embedded inclusions and the accuracy of the measurements.Entities:
Keywords: Elasticity imaging techniques; Phantoms; Reproducibility of results; Transducers; Ultrasonography
Year: 2020 PMID: 32660213 PMCID: PMC7758105 DOI: 10.14366/usg.20011
Source DB: PubMed Journal: Ultrasonography ISSN: 2288-5919
Fig. 1.Elastography phantom.
A. The image shows the external shape of the quality assurance phantom and the position of the ultrasound probe. B. The yellow circles represent the position, stiffness, and shape of different elasticity inclusions in the phantom (reprinted with permission from Computerized Imaging Reference Systems [22]).
Descriptive summary of the mean elasticity (kPa) and differences between operators
| Machine | Inclusion | Operator A | Operator B | ||||
|---|---|---|---|---|---|---|---|
| Mean±SD (kPa) | CV (%) | Mean±SD (kPa) | CV (%) | Mean difference (95% LOA) | Difference (%) | ||
| SSI probe 1 | 10 kPa at 1.5 cm | 18.3±0.5 | 2.6 | 18.1±0.2 | 1.2 | 0.2 (-0.8 to 1.1) | 1.0 |
| 40 kPa at 1.5 cm | 27.2±0.7 | 2.4 | 24.4±0.6 | 2.4 | 2.8 (1.5 to 4.1) | 10.2 | |
| 60 kPa at 1.5 cm | 31.4±0.6 | 2.0 | 29.1±0.6 | 2.0 | 2.3 (1.1 to 3.6) | 7.5 | |
| 10 kPa at 5 cm | 13.5±0.6 | 4.6 | 13.9±0.5 | 3.4 | -0.4 (-1.7 to 0.8) | -3.3 | |
| 40 kPa at 5 cm | 24.5±0.7 | 2.8 | 22.5±0.3 | 2.9 | 2.0 (0.6 to 3.3) | 8.0 | |
| 60 kPa at 5 cm | 29.9±0.7 | 2.4 | 28.1±0.8 | 2.9 | 1.8 (0.4 to 3.2) | 6.0 | |
| SSI probe 2 | 10 kPa at 1.5 cm | 11.0±0.4 | 3.3 | 10.8±0.3 | 3.1 | 0.3 (-0.4 to 1.0) | 2.5 |
| 40 kPa at 1.5 cm | 24.5±0.9 | 3.7 | 23.2±0.5 | 2.1 | 1.3 (-0.5 to 3.1) | 5.4 | |
| 60 kPa at 1.5 cm | 33.5±0.8 | 2.3 | 34.4±1.1 | 3.3 | -1.0 (-2.5 to 0.5) | -2.9 | |
| 10 kPa at 5 cm | 10.7±1.0 | 9.1 | 10.4±1.4 | 13.6 | 0.3 (-1.6 to 2.2) | 3.0 | |
| 40 kPa at 5 cm | 37.9±8.1 | 21.3 | 26.9±7.0 | 26.2 | 11.0 (-4.8 to 26.8) | 29.1 | |
| 60 kPa at 5 cm | 59.6±8.2 | 13.8 | 33.9±4.8 | 14.1 | 25.6 (9.5 to 41.8) | 43.0 | |
| SSI probe 3 | 10 kPa at 1.5 cm | 10.3±0.5 | 4.8 | 9.2±0.2 | 2.2 | 1.1 (0.1 to 2.1) | 10.6 |
| 40 kPa at 1.5 cm | 24.3±0.4 | 1.5 | 23.2±0.5 | 2.0 | 1.1 (0.4 to 1.8) | 4.6 | |
| 60 kPa at 1.5 cm | 35.3±0.5 | 1.5 | 34.9±0.3 | 0.9 | 0.4 (-0.6 to 1.4) | 1.2 | |
| 10 kPa at 5 cm | - | - | - | - | - | - | |
| 40 kPa at 5 cm | - | - | - | - | - | - | |
| 60 kPa at 5 cm | - | - | - | - | - | - | |
| GE | 10 kPa at 1.5 cm | 12.7±0.7 | 5.4 | 10.8±0.2 | 1.9 | 1.9 (0.6 to 3.3) | 15.2 |
| 40 kPa at 1.5 cm | 26.8±0.5 | 1.7 | 26.1±1.3 | 4.9 | 0.7 (-0.2 to 1.6) | 2.6 | |
| 60 kPa at 1.5 cm | 36.7±1.4 | 3.7 | 35.9±1.5 | 4.1 | 0.9 (-1.8 to 3.5) | 2.3 | |
| 10 kPa at 5 cm | 6.6±2.3 | 35.5 | 5.9±1.2 | 20.8 | 0.7 (-3.9 to 5.3) | 10.7 | |
| 40 kPa at 5 cm | 7.6±2.4 | 31.3 | 6.4±2.2 | 34.5 | 1.2 (-3.4 to 5.9) | 15.9 | |
| 60 kPa at 5 cm | 10.2±8.2 | 80.6 | 6.9±1.8 | 26.9 | 3.4 (-12.8 to 19.5) | 32.9 | |
(-) indicates failed to acquire shear wave elastography measurements.
SD, standard deviation; CV, coefficient of variation; 95% LOA, Bland-Altman 95% limits of agreement; SSI, Supersonic Imagine machine; GE, General Electric machine.
Fig. 2.Shear wave elastography (SWE) measurements obtained using the different systems and probes.
All SWE images acquired at a depth of 5 cm. A. The image shows SuperSonic Imagine Aixplorer (SSI) probe 1 for the 60 kPa inclusion. B. The image shows SSI probe 2 for the 10 kPa inclusion. C. The image shows SSI probe 3 for the 40 kPa inclusion which demonstarets a failed SWE acquisition showing a saturation artifact. D. The image shows GE probe 9L-D for the 60 kPa inclusion.
Descriptive summary of the mean elasticity (kPa) and differences between machines
| Machine | Inclusion | SSI | GE | ||||
|---|---|---|---|---|---|---|---|
| Mean±SD (kPa) | CV (%) | Mean±SD (kPa) | CV (%) | Mean difference (95% LOA) | Percentage difference (%) | ||
| Operator A | 10 kPa at 1.5 cm | 11.0±0.4 | 3.3 | 12.7±0.7 | 5.4 | -1.7 (-2.4 to -0.9) | 15.0 |
| 40 kPa at 1.5 cm | 24.5±0.9 | 3.7 | 26.8±0.5 | 1.7 | -2.3 (-4.1 to -0.5) | 9.6 | |
| 60 kPa at 1.5 cm | 33.5±0.8 | 2.3 | 36.7±1.4 | 3.7 | -3.3 (-4.7 to -1.8) | 9.7 | |
| 10 kPa at 5 cm | 10.7±1.0 | 9.1 | 6.6±2.3 | 35.5 | 4.1 (2.2 to 6.0) | -38.6 | |
| 40 kPa at 5 cm | 37.9±8.1 | 21.3 | 7.6±2.4 | 31.3 | 30.3 (14.5 to 46.1) | -79.9 | |
| 60 kPa at 5 cm | 59.6±8.2 | 13.8 | 10.2±8.2 | 80.6 | 49.3 (33.2 to 65.5) | -82.9 | |
| Operator B | 10 kPa at 1.5 cm | 10.8±0.3 | 3.1 | 10.8±0.2 | 1.9 | 0 (-0.6 to 0.6) | 0 |
| 40 kPa at 1.5 cm | 23.2±0.5 | 2.1 | 26.1±1.3 | 4.9 | -2.9 (-3.9 to -2.0) | 12.7 | |
| 60 kPa at 1.5 cm | 34.4±1.1 | 3.3 | 35.9±1.5 | 4.1 | -1.4 (-3.6 to 0.8) | 4.2 | |
| 10 kPa at 5 cm | 10.4±1.4 | 13.6 | 5.9±1.2 | 20.8 | 4.5 (1.7 to 7.3) | -43.4 | |
| 40 kPa at 5 cm | 26.9±7.0 | 26.2 | 6.4±2.2 | 34.5 | 20.5 (6.7 to 34.3) | -76.2 | |
| 60 kPa at 5 cm | 33.9±4.8 | 14.1 | 6.9±1.8 | 26.9 | 27.1 (17.7 to 36.5) | -79.8 | |
(-) indicates failed to acquire shear wave elastography measurements.
SSI, Supersonic Imagine machine; GE, General Electric machine; SD, standard deviation; CV, coefficient of variation; 95% LOA, Bland-Altman 95% limits of agreement.
Fig. 3.Comparison of measurements between machines.
The error bars represent mean elasticity and 95% confidence intervals (CIs) for both machines.
Fig. 4.Comparison of measurements among probes.
The error bars represent mean elasticity and 95% confidence intervals (CIs) for the three probes.