| Literature DB >> 29542277 |
Zaiyang Long1, Donald J Tradup1, Pengfei Song1, Scott F Stekel1, Shigao Chen1, Katrina N Glazebrook1, Nicholas J Hangiandreou1.
Abstract
Because of the rapidly growing use of ultrasound shear wave elastography (SWE) in clinical practices, there is a significant need for development of clinical physics performance assessment methods for this technology. This study aims to report two clinical medical physicists' tasks: (a) acceptance testing (AT) of SWE function on ten commercial ultrasound systems for clinical liver application and (b) comparison of SWE measurements of targets across vendors for clinical musculoskeletal application. For AT, ten GE LOGIQ E9 XDclear 2.0 scanners with ten C1-6-D and ten 9L-D transducers were studied using two commercial homogenous phantoms. Five measurements were acquired at two depths for each scanner/transducer pair by two operators. Additional tests were performed to access effects of different coupling media, phantom locations and operators. System deviations were less than 5% of group mean or three times standard deviation; therefore, all systems passed AT. A test protocol was provided based on results that no statistically significant difference was observed between using ultrasound gel and salt water for coupling, among different phantom locations, and that interoperator and intraoperator coefficient of variation was less than 3%. For SWE target measurements, two systems were compared - a Supersonic Aixplorer scanner with a SL10-2 and a SL15-4 transducer, and an abovementioned GE scanner with 9L-D transducer. Two stepped cylinders with diameters of 4.05-10.40 mm were measured both longitudinally and transaxially. Target shear wave speed quantification was performed using an in-house MATLAB program. Using the target shear wave speed deduced from phantom specs as a reference, SL15-4 performed the best at the measured depth. However, it was challenging to reliably measure a 4.05 mm target for either system. The reported test methods and results could provide important information when dealing with SWE-related tasks in the clinical environment.Entities:
Keywords: zzm321990GEzzm321990; Supersonic; acceptance testing; shear wave elastography; target measurements; ultrasound
Mesh:
Year: 2018 PMID: 29542277 PMCID: PMC5978380 DOI: 10.1002/acm2.12310
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Schematic representations of the experiment setup for (a) investigating the effects of different coupling media, operators, and phantom locations (homogeneity) for the acceptance testing protocol, and (b) longitudinal measurements of the cylindrical test objects for small target task.
Mean and standard deviation (SD) of the shear wave speed measurements (m/s) in the acceptance test of ten GE LOGIQ E9 XDclear 2.0 systems. Shallower and deeper depths were 1 and 4 cm, and 3 and 7 cm, for the 9L‐D and C1‐6‐D transducers, respectively
| Soft phantom | Hard phantom | |||
|---|---|---|---|---|
| Shallower depth | Deeper depth | Shallower depth | Deeper depth | |
| Systems with C1‐6‐D transducers | 0.97 ± 0.01 | 1.00 ± 0.01 | 3.74 ± 0.03 | 3.83 ± 0.10 |
| Systems with 9L‐D transducers | 0.98 ± 0.01 | 0.97 ± 0.02 | 3.56 ± 0.02 | 3.66 ± 0.07 |
Maximum deviation of any individual shear wave speed measurement from the group mean in the acceptance test of ten GE LOGIQ E9 XDclear 2.0 systems (expressed as percentages of the group mean and as multiples of the standard deviation (SD)). Shallower and deeper depths were 1 and 4 cm, and 3 and 7 cm, for the 9L‐D and C1‐6‐D transducers, respectively
| Soft phantom | Hard phantom | |||
|---|---|---|---|---|
| Shallower depth | Deeper depth | Shallower depth | Deeper depth | |
| Systems with C1‐6‐D transducers | 1.2% (1.5 SD) | 1.6% (2.3 SD) | 1.4% (1.5 SD) | 5.3% (2.0 SD) |
| Systems with 9L‐D transducers | 1.3% (2.2 SD) | 5.4% (2.3 SD) | 1.1% (2.0 SD) | 3.7% (2.0 SD) |
Figure 2Shear wave speed measurements made by using different coupling mediums (ultrasound gel or salt water), three operators (O1, O2, and O3), and five fixed probe locations in the homogeneous phantom, in order to assess these factors for the acceptance testing protocol. Operator 1 (O1) also repeated the same dataset after 1 week (O1‐repeat).
Figure 3Sample images of transaxial and longitudinal shear wave speed for the 4.05 mm soft target using the Supersonic Aixplorer scanner with SL15‐4 and SL10‐2 transducers, as well as the GE LOGIQ E9 XDclear 2.0 scanner with the 9L‐D transducer.
Figure 4Transaxial and longitudinal measurements of shear wave speed of the soft and hard targets with three diameters using all scanner/transducer pairs. The reference shear wave speed of the targets is indicated with solid lines and that of the background is indicated with dashed lines.
Transaxial shear wave speed measurements (m/s) of all targets of different diameters using the Supersonic Aixplorer scanner with the SL10‐2 and SL15‐4 transducers, as well as the GE LOGIQ E9 scanner with the 9L‐D transducer
| Soft target | Hard target | |||||
|---|---|---|---|---|---|---|
| 10.40 mm | 6.49 mm | 4.05 mm | 10.40 mm | 6.49 mm | 4.05 mm | |
| Supersonic SL10‐2 | 1.78 ± 0.01 | 1.90 ± 0.04 | 2.10 ± 0.01 | 3.39 ± 0.05 | 3.14 ± 0.03 | 2.83 ± 0.01 |
| Supersonic SL15‐4 | 1.70 ± 0.01 | 1.82 ± 0.01 | 1.99 ± 0.02 | 3.70 ± 0.07 | 3.26 ± 0.02 | 2.89 ± 0.02 |
| GE 9L‐D | 1.90 ± 0.01 | 2.12 ± 0.01 | 2.33 ± 0.02 | 3.52 ± 0.01 | 3.15 ± 0.01 | 2.88 ± 0.06 |
Longitudinal shear wave speed measurements (m/s) of all targets of different diameters using the Supersonic Aixplorer scanner with the SL10‐2 and SL15‐4 transducers, as well as the GE LOGIQ E9 scanner with the 9L‐D transducer
| Soft target | Hard target | |||||
|---|---|---|---|---|---|---|
| 10.40 mm | 6.49 mm | 4.05 mm | 10.40 mm | 6.49 mm | 4.05 mm | |
| Supersonic SL10‐2 | 1.99 ± 0.01 | 2.11 ± 0.004 | 2.35 ± 0.003 | 3.27 ± 0.04 | 2.84 ± 0.01 | 2.66 ± 0.01 |
| Supersonic SL15‐4 | 1.95 ± 0.01 | 2.16 ± 0.01 | 2.34 ± 0.003 | 3.37 ± 0.10 | 3.10 ± 0.03 | 2.73 ± 0.01 |
| GE 9L‐D | 2.14 ± 0.01 | 2.54 ± 0.02 | 2.58 ± 0.01 | 3.54 ± 0.01 | 3.01 ± 0.02 | 2.74 ± 0.01 |
Shear wave speed ratios between the hard and soft targets from all scanner/transducer pair measurements and all target diameters. Both transaxial and longitudinal planes are included. The reference ratio from the phantom specification was 2.37
| Transaxial | Longitudinal | |||||
|---|---|---|---|---|---|---|
| 10.40 mm | 6.49 mm | 4.05 mm | 10.40 mm | 6.49 mm | 4.05 mm | |
| Supersonic SL10‐2 | 1.90 | 1.65 | 1.35 | 1.64 | 1.35 | 1.13 |
| Supersonic SL15‐4 | 2.18 | 1.79 | 1.45 | 1.73 | 1.44 | 1.17 |
| GE 9L‐D | 1.85 | 1.49 | 1.24 | 1.65 | 1.19 | 1.06 |