| Literature DB >> 32577607 |
George A Kontadakis1, Alexandros Pennos2, Iro Pentari2, George D Kymionis2, Ioannis G Pallikaris2, Harilaos Ginis3.
Abstract
PURPOSE: To investigate in vitro the accuracy of dynamic contour tonometry, Goldmann applanation tonometry, and Tono-Pen XL in edematous corneas.Entities:
Keywords: corneal edema; glaucoma; ocular hypertension; tonometry
Year: 2020 PMID: 32577607 PMCID: PMC7288812 DOI: 10.1177/2515841420923190
Source DB: PubMed Journal: Ther Adv Ophthalmol ISSN: 2515-8414
Mean results and comparisons of each tonometry result with the corresponding true IOP.
| Mean values (SD) | IOP 17 mm Hg | IOP 33 mm Hg | IOP 50 mm Hg | |||
|---|---|---|---|---|---|---|
| GAT | 7.7 (5.9) | <0.001 | 14.6 (6.8) | <0.001 | 17.4 (6.33) | <0.001 |
| DCT | 10.6 (7.5) | 0.001 | 21.6 (7.7) | <0.001 | 26.8 (7.35) | <0.001 |
| TXL | 16.6 (8.5) |
| 27.9 (8.8) |
| 35.6 (8.4) | <0.001 |
DCT, dynamic contour tonometry; GAT, Goldmann applanation tonometry; IOP, intraocular pressure; SD, standard deviation; TXL, Tono-Pen XL.
p values noted in plain text indicate statistical significance, p values noted in bold text indicate non- significance.
Results of correlations among measurements of each experimental eye with the three different methods.
| True IOP | 17 mm Hg | 33 mm Hg | 50 mm Hg | ||||
|---|---|---|---|---|---|---|---|
| GAT | DCT | GAT | DCT | GAT | DCT | ||
| TXL | Pearson | 0.693 | 0.659 | 0.576 | 0.769 | 0.010 | 0.190 |
| <0.001 | 0.002 | 0.006 | <0.001 | 0.965 | 0.364 | ||
| DCT | Pearson | 0.980 | 0.70 | 0.296 | |||
| <0.001 | <0.001 | 0.171 | |||||
DCT, dynamic contour tonometry; GAT, Goldmann applanation tonometry; IOP, intraocular pressure; TXL, Tono-Pen XL.