Changes in corneal hysteresis after clear corneal cataract surgery.

Annette Hager1, Kristina Loge, Marc-Oliver Füllhas, Bernd Schroeder, Martin Grossherr, Wolfgang Wiegand.   

Abstract

PURPOSE: To assess the changes in corneal hysteresis (CH) as measured by the Ocular Response Analyzer (ORA; Reichert Ophthalmic Instruments, Buffalo, New York, USA) to describe the influence of clear corneal cataract surgery on corneal viscoelastic properties and intraocular pressure (IOP) measured by noncontact tonometry (NCT) and Goldmann applanation tonometry (GAT).
DESIGN: Retrospective, interventional, comparative study.
METHODS: One hundred and one eyes of 101 consecutive patients who underwent routine clear corneal cataract surgery were evaluated. CH, NCT, and central corneal thickness (CCT) were measured by ORA before surgery and at postoperative day 1. A control group of 48 pseudophakic eyes (surgery >3 months previously) was included.
RESULTS: CCT increased from 556.82 +/- 32.5 microm before surgery to 580.26 +/- 45.5 microm after surgery (P < .001; control, 555.16 +/- 42.33 microm). Mean CH decreased from 10.35 +/- 2.5 mm Hg before surgery to 9.20 +/- 1.9 mm Hg after surgery (P < .001; control, 10.47 +/- 1.63 mm Hg). NCT values rose from 17.85 +/- 3.8 mm Hg before surgery to 20.10 +/- 6.3 mm Hg after surgery. GAT values were 14.85 +/- 2.8 mm Hg before surgery and 15.24 +/- 4.1 mm Hg after surgery (P = .52). There was no significant difference of CCT or CH between the preoperative values and the values of the control group (CCT, P = .986; CH, P = .166), in contrast to the difference between postoperative values and the values of the control group (CCT, P = .005; CH, P = .031).
CONCLUSIONS: At day 1 after clear corneal cataract surgery, CH is diminished, whereas CCT is increased significantly. Postoperative corneal edema leads to a change of corneal viscoelastic properties, resulting in a lower damping capacity of the cornea. It is supposed that GAT and NCT measurements are significantly different because of postoperative changes in viscoelastic properties of the cornea.

Entities:  

Mesh:

Year:  2007        PMID: 17631265     DOI: 10.1016/j.ajo.2007.05.023

Source DB:  PubMed          Journal:  Am J Ophthalmol        ISSN: 0002-9394            Impact factor:   5.258


  14 in total

1.  Corneal biomechanical properties changes after coaxial 2.2-mm microincision and standard 3.0-mm phacoemulsification.

Authors:  Zhe Zhang; Hua Yu; Hui Dong; Li Wang; Ya-Ding Jia; Su-Hua Zhang
Journal:  Int J Ophthalmol       Date:  2016-02-18       Impact factor: 1.779

2.  Combining corneal hysteresis with central corneal thickness and intraocular pressure for glaucoma risk assessment.

Authors:  D Pensyl; M Sullivan-Mee; M Torres-Monte; K Halverson; C Qualls
Journal:  Eye (Lond)       Date:  2012-08-10       Impact factor: 3.775

3.  Repeatability of Corneal Hysteresis Measurements in Glaucoma Patients During Routine Follow Up and After Cataract Surgery.

Authors:  Tzukit Zehavi-Dorin; Anne E Kutzscher; Mai Badr; Marta Mora; Yennie Shyu; Travis C Porco; Robert L Stamper
Journal:  J Glaucoma       Date:  2022-03-01       Impact factor: 2.290

4.  [Influence of corneal biomechanical properties on myopic regression after laser in situ keratomileusis].

Authors:  D Uthoff; K Hebestedt; G I W Duncker; E Spörl
Journal:  Ophthalmologe       Date:  2013-01       Impact factor: 1.059

5.  [Measurement of viscoelastic corneal parameters (corneal hysteresis) in patients with primary open angle glaucoma].

Authors:  B Schroeder; A Hager; A Kutschan; W Wiegand
Journal:  Ophthalmologe       Date:  2008-10       Impact factor: 1.059

6.  [Methods of measuring intraocular pressure independently of central corneal thickness].

Authors:  A Hager; W Wiegand
Journal:  Ophthalmologe       Date:  2008-09       Impact factor: 1.059

7.  Changes of extracellular matrix of the cornea in diabetes mellitus.

Authors:  A Hager; K Wegscheider; W Wiegand
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2009-05-13       Impact factor: 3.117

8.  [Biomechanical condition of the cornea as a new indicator for pathological and structural changes].

Authors:  E Spörl; N Terai; M Haustein; A G Böhm; F Raiskup-Wolf; L E Pillunat
Journal:  Ophthalmologe       Date:  2009-06       Impact factor: 1.059

9.  Corneal Biomechanical Changes Following Trabeculectomy, Phaco-trabeculectomy, Ahmed Glaucoma Valve Implantation and Phacoemulsification.

Authors:  Mohammad Pakravan; Mohsen Afroozifar; Shahin Yazdani
Journal:  J Ophthalmic Vis Res       Date:  2014-01

10.  Impaired corneal biomechanical properties and the prevalence of keratoconus in mitral valve prolapse.

Authors:  Emine Kalkan Akcay; Murat Akcay; Betul Seher Uysal; Pinar Kosekahya; Abdullah Nabi Aslan; Mehtap Caglayan; Cemal Koseoglu; Fatma Yulek; Nurullah Cagil
Journal:  J Ophthalmol       Date:  2014-04-17       Impact factor: 1.909

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.