| Literature DB >> 805582 |
Abstract
Intraocular pressure was determined manometrically in owl monkey eyes with normal and edematous corneas. The pressure was measured with Perkins hand-held, Goldmann, and MacKay-Marg tonometers to compare relative accuracy. At monometric levels of 20, 30, and 40 mm Hg, a statistically significant difference in readings was found between normal and edematous corneas with use of the Perkins and Goldmann applanators, whereas no such difference was found with the MacKay-Marg tonometer. The MacKay-Marg tonometer most closely approximated the true intraocular pressure in edematous corneas.Entities:
Mesh:
Substances:
Year: 1975 PMID: 805582 DOI: 10.1001/archopht.1975.01010020434007
Source DB: PubMed Journal: Arch Ophthalmol ISSN: 0003-9950