Tonometry in adults and children. A manometric evaluation of pneumatonometry, applanation, and TonoPen in vitro and in vivo.

D L Eisenberg1, B G Sherman, C A McKeown, J S Schuman.   

Abstract

OBJECTIVE: The purpose of the study was to determine the accuracy of applanation tonometry, pneumatonometry, and TonoPen tonometry in adults and children and the effect of age on tonometer error.
DESIGN: The design was divided into four parts: part 1 was prospective and cross-sectional, and parts 2 through 4 were prospective, cross-sectional, and masked. PARTICIPANTS: This study contained 72 patients representing 74 data points. INTERVENTION: Tonometry with simultaneous manometry was performed. MAIN OUTCOME MEASURES: Intraocular pressure (i.o.p.) and the tonometric estimate of IOP were obtained.
RESULTS: The normal pediatric IOP follows the line Ta = 0.71 age(years) + 10 up to age 10. Applanation tonometry under anesthesia differs from pneumatonometry by an average of -8.6 mmHg and is age related by the equation Ta = Tpn + 2.6 log(age) - 10.3. The TonoPen was the most accurate instrument for enucleated eyes, and the pneumatonometer was the most accurate in anesthetized living eyes.
CONCLUSIONS: Applanation tonometry markedly underestimated IOP in young eyes. TonoPen tonometry performed well with enucleated eyes but was not adequately accurate for clinical use. The pneumatonometer performed the best clinically and the best overall.

Entities:  

Mesh:

Year:  1998        PMID: 9663218     DOI: 10.1016/S0161-6420(98)97016-6

Source DB:  PubMed          Journal:  Ophthalmology        ISSN: 0161-6420            Impact factor:   12.079


  24 in total

1.  The influence of central corneal thickness and age on intraocular pressure measured by pneumotonometry, non-contact tonometry, the Tono-Pen XL, and Goldmann applanation tonometry.

Authors:  P-A Tonnu; T Ho; T Newson; A El Sheikh; K Sharma; E White; C Bunce; D Garway-Heath
Journal:  Br J Ophthalmol       Date:  2005-07       Impact factor: 4.638

2.  Underestimation of intraocular pressure after photorefractive keratectomy: a biomechanical analysis.

Authors:  Per Hallberg; Anders Eklund; Kenneth Santala; Timo Koskela; Olof Lindahl; Christina Lindén
Journal:  Med Biol Eng Comput       Date:  2006-07-26       Impact factor: 2.602

3.  Symmetric sensor for applanation resonance tomometry of the eye.

Authors:  Per Hallberg; Christina Lindén; Tomas Bäcklund; Anders Eklund
Journal:  Med Biol Eng Comput       Date:  2006-03       Impact factor: 2.602

4.  Intraocular Pressure Measurement Accuracy and Repeatability of a Modified Goldmann Prism: Multicenter Randomized Clinical Trial.

Authors:  Sean McCafferty; Kyle Tetrault; Ann McColgin; Warren Chue; Jason Levine; Melissa Muller
Journal:  Am J Ophthalmol       Date:  2018-09-07       Impact factor: 5.258

5.  Can we trust intraocular pressure measurements in eyes with intracameral air?

Authors:  Gauti Jóhannesson; Christina Lindén; Anders Eklund; Anders Behndig; Per Hallberg
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2014-07-17       Impact factor: 3.117

6.  Age differences in central and peripheral intraocular pressure using a rebound tonometer.

Authors:  J M González-Méijome; J Jorge; A Queirós; P Fernandes; R Montés-Micó; J B Almeida; M A Parafita
Journal:  Br J Ophthalmol       Date:  2006-08-02       Impact factor: 4.638

7.  Comparative evaluation of TonoPen AVIA, Goldmann applanation tonometry and non-contact tonometry.

Authors:  Shibal Bhartiya; Shveta Jindal Bali; Reetika Sharma; Neha Chaturvedi; Tanuj Dada
Journal:  Int Ophthalmol       Date:  2011-07-15       Impact factor: 2.031

8.  Accuracy of dynamic contour tonometry compared with applanation tonometry in human cadaver eyes of different hydration states.

Authors:  Christoph Kniestedt; Michelle Nee; Robert L Stamper
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2004-10-13       Impact factor: 3.117

9.  Evaluation of rebound tonometry in non-human primates.

Authors:  Elizabeth J Elsmo; Julie A Kiland; Paul L Kaufman; Gillian J McLellan
Journal:  Exp Eye Res       Date:  2011-02-16       Impact factor: 3.467

10.  Ocular hypertensive response to topical dexamethasone ointment in children.

Authors:  Yoon Jung Lee; Chan Yi Park; Kyung In Woo
Journal:  Korean J Ophthalmol       Date:  2006-09