We aimed to assess the usefulness of slice-encoding metal artifact correction (SEMAC) for the evaluation of spinal metallic implants and peripheral soft tissue lesions at 3T magnetic resonance.Twenty-seven patients with spinal metal implants underwent both SEMAC and high bandwidth (HiBW) based sequences scanning for reduction artifacts. The area size and maximum longitude of artifacts, the peri-prosthetic soft tissue, and metal visualization were assessed by 2 independent doctors, as well as the lesions signs were reviewed by 2 senior readers. A paired 2-tailed t-test and McNemar test were used for statistical analysis.The size of artifacts on SEMAC images decreased by 37% and 24%, and the scores are higher than that on HiBW images. T1 weighted (T1W)-SEMAC acquired the highest score in metal prosthesis visualization, while short tau inversion recovery SEMAC showed more signs of lesions than clinical HiBW group.SEMAC effectively reduces the metal artifacts and is useful for assessing soft tissue lesions.
We aimed to assess the usefulness of slice-encoding metal artifact correction (SEMAC) for the evaluation of spinal metallic implants and peripheral soft tissue lesions at 3T magnetic resonance.Twenty-seven patients with spinal metal implants underwent both SEMAC and high bandwidth (HiBW) based sequences scanning for reduction artifacts. The area size and maximum longitude of artifacts, the peri-prosthetic soft tissue, and metal visualization were assessed by 2 independent doctors, as well as the lesions signs were reviewed by 2 senior readers. A paired 2-tailed t-test and McNemar test were used for statistical analysis.The size of artifacts on SEMAC images decreased by 37% and 24%, and the scores are higher than that on HiBW images. T1 weighted (T1W)-SEMAC acquired the highest score in metal prosthesis visualization, while short tau inversion recovery SEMAC showed more signs of lesions than clinical HiBW group.SEMAC effectively reduces the metal artifacts and is useful for assessing soft tissue lesions.
Vertebral fractures are typically caused by major trauma, particularly those that occur in osteoporosis patients. Such fractures can directly lead to secondary spinal cord and periphery soft tissues damage.[ Operation treatment is aimed at decompressing the vertebral canal and stabilizing the disrupted centrums. Thus, metallic implants, such as pedicle screws, are often used in orthopedic surgery to fixate the fractures to align and immobilize the vertebra. However, some complications can still occur after the operation, including infections, neural deficits, and Brown–Séquard syndrome, which may affect the prognosis of the patients.[At present, X-ray plain radiographs and computed tomography scanning are 2 routine tools that are used in the assessment of spinal metal prostheses during the perioperation.[ However, X-ray plain radiographs provide little information regarding physical characteristics, which are based on the natural contrast between the implant metal and human structures. Although computed tomography imaging with the 2-dimensional cross-sectional method and raw data can be reconstructed into 3-D images, which will provide better images, the metal-induced streak/beam-hardening artifacts still contaminate the data around the metal implants. Subsequently, these 2 methods are not sensitive for imaging the soft tissue complications around the implants, such as bone marrow lesions, infections, and edemas.[Magnetic resonance imaging (MRI) is one of the most sensitive tools for detecting lesions of soft tissues due to its higher contrast in different sequences, potentially without any contrast media. However, metallic implants can cause local magnetic field inhomogenous, which cause the MRI images to suffer from severe artifacts, including signal loss and blurring.[ Traditionally, the distortion of MRI with metal implants can be categorized as through-plane or in-plane distortions.[ The use of thin slices and high bandwidth (HiBW), both for radiofrequency pulses and signal readout, has been suggested as an option to partially reduce these distortions. Recently, some MRI acquisition techniques were developed to reduce the artifacts of metal in the body. In 1988, the view-angle tilting (VAT) technique was first introduced to correct in-plane distortions[; however, this technique cannot resolve through-plane distortions.[ To resolve this problem, the advanced magnetic resonance (MR) techniques of slice- encoding metal artifact correction (SEMAC) and multi-acquisition variable-resonance image combination have been proposed to minimize the artifacts near metallic prostheses.[ By extending the VAT turbo spin echo (TSE) with an additional phase-encoding along the slice-selective z-axis, SEMAC uses a TSE acquisition mode to correct the profiles of each excited slice in the region. The resolved profiles of all slices in the region of interest (ROI) are aligned to their actual voxel locations to resolve through-slice distortions.[In this study, we aimed to assess the spinal metallic implants and peripheral soft tissues with SEMAC using a 3 T MR scanner, and to evaluate the clinical value of SEMAC in spinal metal prostheses comparing with clinical HiBW sequences.
Materials and methods
General information of the subjects
This prospective study was approved by our Institutional Review Board and performed between January 2015 and December 2017. We established the following inclusion criteria: patients with a history of spinal fixation surgery, limb paralysis, lower back pain, and other local symptoms or with a history of tumor in the surgical region; patients also had to provide informed consent. The exclusion criteria were patients with a cardiac pacemaker or who had major complications; who cannot bear to complete the scanning; and who are allergic to the contrast agent.Thirty-three patients with metallic prostheses after spinal surgery were recruited for this research, and 27 patients met the inclusion criteria (15 females and 12 males). The average age was 61.6 ± 11.3 years. The average fixation segments were 2.52 levels: 2 segments of 17 patients, 3 segments of 6 patients, and 4 segments of 4 patients. Postoperative MRI was performed in 6 conditions:back pain without disc herniation recurrence (n = 9),back pain with other-level herniated nucleus pulposus (n = 7),postoperative evaluation of intraspinal tumors (n = 6),back pain with sequelae of compression fracture (n = 3),back pain with a fresh compression fracture (n = 1), andfever evaluation (n = 1).
MRI protocols
All MRI examinations were performed on a 3 T MRI scanner (MAGNETOM Skyra, Siemens Healthcare, Erlangen, Germany) using an integrated spine coil. SEMAC sequences were prototypes provided by Siemens Healthcare. Short tau inversion recovery (STIR) sequence and T1W TSE with and without the advanced SEMAC techniques were prescribed to image metals in the patient spines (see Table 1), and the generalized autocalibrating partially parallel acquisition technique was used in the SEMAC group to reduce the total acquisition time. The parameters of the SEMAC and HiBW sequences are summarized in Table 2.
Table 1
Two groups of different imaging sequences.
Table 2
Imaging parameters of the 2 groups.
Two groups of different imaging sequences.Imaging parameters of the 2 groups.
Data analysis
In the coronal images, 2 fellowship-trained radiologists drew ROIs of the axial pedicle screws, which included the blurring and signal loss regions to measure the size of the metallic artifacts. Two radiologists performed a semi-quantitative assessment of the prosthesis and peripheral soft tissues. The patients’ information and imaging parameters were hidden to minimize the learning bias. Two experienced musculoskeletal radiologists (Prof Ji and Dr Lin, with 35 and 16 years of experience in the interpretation of spinal MRI, respectively) analyzed the different imaging findings of all of the in vivo data, and the positive signs were recorded.We referenced the literature to establish a semi-quantitative standard to assess the effect of the sequences on reducing the metal artifacts and usefulness for clinical diagnosis. Five imaging findings of periprosthetic tissues and prosthetic regions were evaluated, including:the structures in vertebral canal, such as the dural sac, spinal cord, and cauda equine,the intervertebral neural foramens,the implant interface (bone marrow, muscles),the intervertebral disc between the affected intervertebral disc level, andoverall artifacts.They were assessed using 5-point scales as follows[:A grade of 1 indicated the worst quality for interpretation, where the periprosthetic region was barely delineated; a grade of 2 indicated visualization of less than 25% of the aforementioned structures; a grade of 3 indicated visualization of 25% to 50% of the structures; a grade of 4 indicated visualization of 50% to 75% of the structures; and a grade of 5 indicated visualization of more than 75% of the structures. With the cross-reference function, axial images that were obtained at the level of the center of the pedicle screws were selected for evaluation of visibility of the dural sac, visibility of the bone-implant interface and overall artifacts. Axial images just below the pedicles or the coronal images were selected for evaluation on the visibility of neural foramens.Metallic prostheses, which were delineated by a dark signal intensity with clearly marginated prosthetic contours, were graded as follows: a grade of 1 indicated nearly complete non-visualization of the prosthesis; a grade of 2 indicated visualization of less than one third of the prosthesis; a grade of 3 indicated visualization of one to two thirds of the prosthesis; a grade of 4 indicated visualization of more than two thirds of the prosthesis; and a grade of 5 indicated that the entire prosthesis was delineated and free of artifacts.
Statistical analysis
The data were recorded using Excel 2013 (Microsoft Corporation) and analyzed using software IBM SPSS (Statistical Package for Social Sciences) Statistics version 21.0 for Windows. The screw artifact sizes between 2 imaging groups were compared using a paired samples t-test, and the semi-quantitative scores of the 2 group images were obtained using a paired samples t-test. A comparison of the abnormal signs from different sequences was performed using the McNemar test for the matched data.
Results
Quantitative evaluation of metal artifacts
The Dice index for ROIs drawn by the 2 fellowship-trained radiologists was 0.82 ± 0.03. The areas of the pedicle screws are shown in Table 3. The area of the SEMAC group decreased by 36.95% compared to the traditional HiBW group, and the maximum artifact length decreased by 23.8%.
Table 3
Sizes of the axial artifacts of 102 pedicle screws.
Sizes of the axial artifacts of 102 pedicle screws.
Semi-quantitative evaluation of metal artifacts
The kʼs Cohen for the 2 experienced radiologists was 0.84 ± 0.04. The semi-quantitative assessment for peri-prosthetic tissues showed that the STIR-/T1W-SEMAC scores were significantly higher than those of the clinical HiBW group (P < .05) (Table 4). Figure 1 shows the metal prosthesis visualization scores in 4 sequences from 2 radiologists. The scores in descending order were T1W-SEMAC > STIR-SEMAC > T1W-HiBW > STIR-HiBW, and the scores of T1W-SEMAC were significantly higher than other groups.
Table 4
Semi-quantitative assessment for peri-prosthetic tissues.
Figure 1
shows a comparison of the semi-quantitative evaluation scores of metal prostheses; T1 weighted slice-encoding metal artifact correction has a higher score than the other sequences.
Semi-quantitative assessment for peri-prosthetic tissues.shows a comparison of the semi-quantitative evaluation scores of metal prostheses; T1 weighted slice-encoding metal artifact correction has a higher score than the other sequences.
Positive signs
Table 5 provides a comparison of the positive signs between 2 different STIR sequences from 2 observers. Significantly, more bone marrow lesions, herniated intervertebral discs, intervertebral foramen and radiculitis, and soft tissue edemas were detected in the STIR-SEMAC group than in the clinical HiBW group. In addition, there were more positive signs in the STIR-SEMAC group than in the STIR-HiBW group.
Table 5
Comparison of positive signs between 2 different STIR sequences.
Comparison of positive signs between 2 different STIR sequences.Table 6 shows that there were more instances of peri-prosthetic osteolysis and soft tissue edemas in T1W-SEMAC than in the T1W-HiBW group, and there were significant differences between the 2 groups. Moreover, there were more positive signs in the T1W-SEMAC group than in the T1W-HiBW group.
Table 6
Comparison of the positive signs between 2 different T1W sequences.
Comparison of the positive signs between 2 different T1W sequences.
Case interpretation
A 46-year-old male with spinal fixation is shown in Figure 2. Picture A was imaged with fat saturation prepared T2 weighted (T2W) TSE, and the surgical area shows ripple-like metal artifacts blurring the peripheral soft tissues, which makes it difficult to evaluate the condition. Picture B was imaged with STIR-HiBW, and picture C was imaged with STIR-SEMAC. We can observe a large range of exudation and effusion around the cross-link device in picture C, and the metal artifacts were decreased compared to pictures A and B. Pictures D–F are the subsequent weekly follow-up images with STIR-SEMAC. In this 4-week period, antibiotics were used moderately, and the exudation surrounding the implanted metals, bone marrow edema, and soft tissue swelling were successively alleviated as reflected in the MRI. The white arrows in pictures G and H indicate the nerve roots running out from the unilateral intervertebral foramens, with slight visible effusion in the nerve root sheath sleeve shown; the green arrows indicate the pedicle screws.
Figure 2
A 46-year-old male with spinal fixation subjected to magnetic resonance imaging (MRI) scanning 4 times for fever evaluation. Pictures A–C are the first MRI examination. Picture A is an image obtained with the traditional turbo spin echo sequence combined with selected fat saturation, where the surgical area shows ripple-like metal artifacts. Picture B was imaged with short tau inversion recovery -high bandwidth, and picture C was imaged with short tau inversion recovery-slice-encoding metal artifact correction. There was considerable exudation and effusion around the cross-link device in picture C (red arrows), and the metal artifacts were decreased compared to A and B. Pictures D–F are the subsequent weekly follow-up images with short tau inversion recovery-slice-encoding metal artifact correction. In this 4-week period, antibiotics were used moderately, and the exudation surrounding the implanted metals, bone marrow edema, and soft tissue swelling were successively alleviated as reflected in the MRI. The white arrows in pictures G and H indicate the nerve roots running out from the unilateral intervertebral foramens. Slight visible effusion in the nerve root sheath sleeve can be observed. The green arrows indicate the pedicle screws.
A 46-year-old male with spinal fixation subjected to magnetic resonance imaging (MRI) scanning 4 times for fever evaluation. Pictures A–C are the first MRI examination. Picture A is an image obtained with the traditional turbo spin echo sequence combined with selected fat saturation, where the surgical area shows ripple-like metal artifacts. Picture B was imaged with short tau inversion recovery -high bandwidth, and picture C was imaged with short tau inversion recovery-slice-encoding metal artifact correction. There was considerable exudation and effusion around the cross-link device in picture C (red arrows), and the metal artifacts were decreased compared to A and B. Pictures D–F are the subsequent weekly follow-up images with short tau inversion recovery-slice-encoding metal artifact correction. In this 4-week period, antibiotics were used moderately, and the exudation surrounding the implanted metals, bone marrow edema, and soft tissue swelling were successively alleviated as reflected in the MRI. The white arrows in pictures G and H indicate the nerve roots running out from the unilateral intervertebral foramens. Slight visible effusion in the nerve root sheath sleeve can be observed. The green arrows indicate the pedicle screws.A 73-year-old male patient is shown in Figure 3. Pictures A and B were preoperative sagittal fat-suppressed T2W and axial fat-suppressed T1W with gadolinium diethylene triamine pentacetic acid (Gd-DTPA enhancement images obtained approximately 3 years ago respectively. Picture A shows bone destruction in several cervical vertebral bodies and accessories at the level of C2–4. In addition, it illustrates spinal stenosis and soft tissue in the epidural space. Picture B shows lesions in the vertebral body, accessories, intraspinal epidural soft tissues, and the paravertebral soft tissue were significantly enhanced, with expansion of the right intervertebral foramen, and compressed spinal cord. We considered a diagnosis of a malignant tumor (i.e., lymphoma, metastasis). The lesions were observed, as indicated by the red arrows.
Figure 3
A diffuse large B cell lymphoma patient. Pictures A and B were preoperative sagittal fat-suppressed T2 weighted and axial fat-suppressed T1 weighted with gadolinium diethylene triamine pentacetic acid (Gd-DTPA) enhancement images taken approximately 3 years ago. Picture A shows bone destruction in several cervical vertebral bodies and accessories at the level of C2–4. Spinal stenosis and soft tissue in the epidural space are also observed. Picture B shows that lesions in the vertebral body, accessories, intraspinal epidural soft tissues, and the paravertebral soft tissue were significantly enhanced, with expansion of the right intervertebral foramen and compressed spinal cord (lesions pointed by red arrows). Pictures D–F are images of enhanced T1 weighted with slice-encoding metal artifact correction and view-angle tilting, where the metal artifacts (blue arrows) are narrower than in Picture C. The spinal cord is indicated by the green arrows.
A diffuse large B cell lymphoma patient. Pictures A and B were preoperative sagittal fat-suppressed T2 weighted and axial fat-suppressed T1 weighted with gadolinium diethylene triamine pentacetic acid (Gd-DTPA) enhancement images taken approximately 3 years ago. Picture A shows bone destruction in several cervical vertebral bodies and accessories at the level of C2–4. Spinal stenosis and soft tissue in the epidural space are also observed. Picture B shows that lesions in the vertebral body, accessories, intraspinal epidural soft tissues, and the paravertebral soft tissue were significantly enhanced, with expansion of the right intervertebral foramen and compressed spinal cord (lesions pointed by red arrows). Pictures D–F are images of enhanced T1 weighted with slice-encoding metal artifact correction and view-angle tilting, where the metal artifacts (blue arrows) are narrower than in Picture C. The spinal cord is indicated by the green arrows.Next, the patient underwent a hollow needle biopsy at Fudan University Shanghai Cancer Center, and a pathological smear was observed. A large number of lymphocytes and some atypical cells were observed, and non-Hodgkin lymphoma was considered. On January 5, 2013, under general anesthesia, the patient was subjected to a posterior tumor resection with cervical fixation surgery, and postoperative biopsy was prompted: malignant lymphoma and diffuse large B cell lymphoma invaded the striated muscle. The immunohistochemistry results were NSE (–), S100 (–), GFAP (+), BCL-2 (+), BCL-6 (+), CD20 (++), CD79a (scattered positive), CD45LCA (++), CD8 (scattered positive), ki67 (20% positive), CD5 (+), Pax-5 (+), CD19 (+), CD45RA (part-positive), and MUM-1 (–).The patient was then treated with chemotherapy in our hospital and followed-up. Pictures C–F are the latest postoperative follow-up images: Enhanced T1W image with HiBW (C) shows some metal artifacts on the level of C3 vertebral, as indicated by the blue arrow. In addition, the spinal cord is observed. However, the metal artifacts in the images D–F, which were imaged with T1W using SEMAC and VAT, were narrower than in the images taken without these techniques, such as image C. The spinal cord and other intraspinal structures (green arrow), paraspinal soft tissues can be more clearly imaged.Images A and B in Figure 4 are 2 T1W images acquired using a HiBW acquisition sequence with or without SEMAC. They are located at the field of view margin slice of the metal implants, as indicated by the yellow line marked in picture C. The SEMAC factor was still 8, that is, there were 8 slices in addition to the center, which was used mathematically to correct the distortion artifacts through the plane. However, the area of artifacts in image B is not significantly reduced compared to A. Nevertheless, SEMAC can display more details of the prosthesis, such as the pedicle screw (green arrows), which is indicated clearly in B, whereas the traditional common HiBW technology does not show these details.
Figure 4
Pictures A and B are 2 T1 weighted images acquired using the high-bandwidth acquisition sequence with or without slice-encoding metal artifact correction. They are located at the field of view margin slice of the metal implants, as indicated by the yellow line marked in picture C. The area of artifacts in B is not significantly reduced compared to A, but the slice-encoding metal artifact correction better displayed more details of the prosthesis, as shown by the green arrows.
Pictures A and B are 2 T1 weighted images acquired using the high-bandwidth acquisition sequence with or without slice-encoding metal artifact correction. They are located at the field of view margin slice of the metal implants, as indicated by the yellow line marked in picture C. The area of artifacts in B is not significantly reduced compared to A, but the slice-encoding metal artifact correction better displayed more details of the prosthesis, as shown by the green arrows.Figure 5 presents a case of a plasma lymphoma patient after surgery. Picture A is an optimized HiBW image, and images B and C are T2W and T1W images, respectively, which were all acquired using SEMAC. There were decreased artifacts in images B and C compared to A, and an irregular-shaped tumor was observed in the sacral canal. Picture F is the mask of the T1W image, after gadolinium injection. The subtraction of E and F can observe such that the tumor is significantly enhanced, and adjacent soft tissues of internal fixation showed enhancement considering the tumor recurrence and invasion paraspinal soft tissues.
Figure 5
This is another case of a patient with plasmablastic lymphoma who was subjected to magnetic resonance imaging scanning for recurrence assessment. Picture A is an optimized clinical high bandwidth image. Pictures B and C are T2 weighted and T1 weighted images, respectively, which were all acquired using magnetic resonance imaging. Vertebral canal and caudal equine can be observed in Picture D. After injection with gadolinium diethylene triamine pentacetic acid, the subtraction of E and F (as shown in G) indicates that the tumors and adjacent soft tissues were significantly enhanced, which should be used to determine the tumor recurrence and invasion.
This is another case of a patient with plasmablastic lymphoma who was subjected to magnetic resonance imaging scanning for recurrence assessment. Picture A is an optimized clinical high bandwidth image. Pictures B and C are T2 weighted and T1 weighted images, respectively, which were all acquired using magnetic resonance imaging. Vertebral canal and caudal equine can be observed in Picture D. After injection with gadolinium diethylene triamine pentacetic acid, the subtraction of E and F (as shown in G) indicates that the tumors and adjacent soft tissues were significantly enhanced, which should be used to determine the tumor recurrence and invasion.
Discussion
The spinal fixation technique is now widely used in the treatment of spinal degenerative disease, tumors, traumas, and other infections, and the most common used is the pedicle screw fixation system.[ However, in practice, many factors may lead to a poor fixed position or even failure and may lead to some complications.[ Thus, a more accurate assessment tool is urgently needed to properly evaluate the internal fixation and peripheral soft tissues, thereby providing more accurate imaging data for the determination of successive treatment. MRI is 1 technique with a large potential for the assessment of the metal and surrounding tissue conditions.Spinal fixation techniques have achieved significant progress with the development of spinal biomechanics and materials science. Titanium material is widely used as an implantation prosthesis due to its good biocompatibility. Furthermore, there are no significant ferromagnetism and thermal magnetic effects, and patients with titanium prostheses can be safely examined in clinical MR scanners, which may not cause the implantation to move in the body and may not produce heat that injures the tissues. In our study, 27 patients had no complaints on discomfort during the entire MRI examination. Thus, the fixed titanium alloy is safe for humans under the 3 T scanner, as concluded by previous reports on 3 T scanners.In recent years, a number of teams have developed novel MRI techniques to reduce metal artifacts,[ and many achievements had been expected to put into clinical use.[ Sequence-based methods, such as SEMAC and VAT, combined with HiBW acquisition methods are aimed at controlling and resolving the distortions through-plane and in-plane. Starting from traditional high-bandwidth acquisition technology for reducing artifacts, we applied SEMAC and VAT to 27 cases of patients, the results of which were compared to those obtained with conventional sequences with HiBW, but without SEMAC and VAT techniques. According to these results, the pedicle screw artifacts with SEMAC and VAT have been reduced significantly compared to those with conventional technologies.The visibility of the metal and structures surrounding the implant, such as the dura, spinal cord, intervertebral foramens, nerve roots, bone-implant surface and paraspinal soft tissues, was increased using the SEMAC/VAT correction. In particular, the distinct visualization of the nerve roots between the upper and lower screws can fully explain the role of VAT and SEMAC in the reduction of artifacts in-plane and through-plane. The 2 radiologists performed a semi-quantitative assessment and showed that the novel techniques of SEMAC-VAT may be a predominant clinical MRI tool in the evaluation of the spinal implant and peripheral soft tissues.Fat suppression is frequently used in MR imaging to suppress adipose tissue signal and enhance the detection of pathological tissues. Among various fat suppression methods, chemical shift-selective fat saturation is recommended in the presence of large amounts of adipose tissue and is commonly used in spinal MR imaging. Lee[ found that SEMAC correction in fat-saturated T2-W MR images led to significantly greater reductions in the metallic artifacts. However, the signal pile-up could not be completely corrected. The main limitation of spectral fat saturation is the sensitivity to the magnetic field in homogeneity in particular when caused by susceptibility artifacts near metallic implants. In this study, we used a STIR-prepared T2-W TSE sequences for fat suppression to improve the visibility of implants and soft tissue lesions. According to our results, STIR-prepared fat suppression provides more robust reduction of fat signal and a high-quality contrast for the evaluation of bone marrow and other soft tissue lesions surrounding the prosthesis. In Figure 2, after the use of antibiotics, the exudation around the cross connection was significantly reduced. In Figures 3–5, lesions, such as tumors, bone marrow edema, and nerve root sleeve effusion, were clearly observed in STIR with SEMAC-VAT images. Thus, the STIR fat-suppression TW images with SEMAC-VAT correction is expected to play an important role in the diagnosis of lesions, such as tumors, infections, and fresh fractures, which can also be used to assess the lesion progression and therapeutic effect among patients with metallic prosthesis.However, among patients with spinal primary or metastasized malignant tumors, the most important assessment after receiving a metallic prosthesis transplantation is to make a diagnosis regarding whether the tumor is a recurrent or new metastasis. As we previously described above, the STIR fat suppression technique is typically used to detect the presence of lesions, but there are still some difficulties regarding making qualitative diagnoses. Thus, it is necessary to inject the contrast agent Gd-DTPA to observe the blood supply of the mass and to perform a comprehensive assessment of recurrence. In our study, Figure 3 shows that the patient was subjected to a posterior cervical vertebral surgery for the invasion of non-Hodgkin lymphoma, and STIR and T1W enhanced MRI with SEMAC-VAT correction showed no clear soft tissue mass in the postoperative evaluation. Figure 5 shows a patient with a symptom of lower extremity debilitation 6 months later after spinal surgery for plasmablastic lymphoma invasion of the lumbosacral canal. STIR-T2W with SEMAC prompted sacral canal mass, with the T1W-SEAMC subtraction of the pre- and post- Gd-DTPA injection, and the tumor was significantly enhanced. Although we found that in this study, T1W-SEMAC is best for metal prosthesis visualization, even in the margin of the field of view, and it can provide more details compared to the control group. However, when T1W-SEMAC was used in contrast-enhancement imaging for cancer diagnosis, it was likely to miss the peak time in these hyper-vascular tumors because its scan time was excessively long. And the fat suppression techniques such as short inversion recovery or Dixon fat and water separation method on T1W-SEMAC may be beneficial to increase the contrast between the tumors and the peri-prosthesis soft tissues.There are some limitations to our research. First, the artifact area was reduced by the application of 2 strategies, but the biases on the qualitative and quantitative evaluation may be unavoidable. Second, the total scanning time of SEMAC was relatively long due to the additional phase-encoding on the slice direction, and this longer scanning time must be compensated for with acceleration strategies (parallel imaging, partial sampling, and increased echo train length) to ensure clinically acceptable scan times. The small number of study objects is another limitation of our study. A subsequent study with more cases will be conducted to obtain a more reliable statistically result.In conclusion, optimized pulse sequences equipped with SEMAC and VAT were applied for the assessment of spinal metallic endoprostheses, and clinical benefits were demonstrated. These sequences significantly reduced metal artifacts in the qualitative and quantitative analyses and may thus facilitate the follow-up of patients who have undergone spine surgery.
Acknowledgments
We thank Siemens (Siemens Healthcare GmbH, Erlangen, Germany) for providing the work-in-progress software package, including the TSE metal-artifact-reducing sequences.
Author contributions
Conceptualization: Shihong Li.Date curation: Chun Xin, Houdong Liu, Shihong Li, Guangwu Lin.Formal analysis: Guangwu Lin.Methodology: Chun Xin, Houdong Liu.Funding acquisition: Shihong Li.Software: Shihong Li, Guangwu Lin.Writing – original draft: Houdong Liu.Writing – review & editing: Chun Xin.
Authors: K M Koch; A C Brau; W Chen; G E Gold; B A Hargreaves; M Koff; G C McKinnon; H G Potter; K F King Journal: Magn Reson Med Date: 2011-01 Impact factor: 4.668
Authors: Steven J Atlas; Robert B Keller; Yen A Wu; Richard A Deyo; Daniel E Singer Journal: Spine (Phila Pa 1976) Date: 2005-04-15 Impact factor: 3.468
Authors: Steven D Glassman; Leah Y Carreon; John R Dimar; Mitchell J Campbell; Rolando M Puno; John R Johnson Journal: Spine J Date: 2007-01-09 Impact factor: 4.166