Artifact reduction from metallic dental materials in T1-weighted spin-echo imaging at 3.0 tesla.

Sang-Young Zho1, Min-Oh Kim, Keun-Woo Lee, Dong-Hyun Kim.   

Abstract

PURPOSE: To investigate and propose a method of artifact reduction arising from metallic dental materials by applying a slice-encoding for metal artifact correction (SEMAC) technique on T1-weighted spin-echo (SE) imaging at 3 Tesla.
MATERIALS AND METHODS: The view angle tilting (VAT) technique was adapted to conventional T1-weighted spin-echo (SE) sequence to correct the in-plane distortion, and the SEMAC technique was used for correcting the remaining through-plane distortions. Fourier transform based B0 field simulations were performed to estimate the amount of field perturbation and a scout imaging method was developed which guide in selecting the number of slice-encodings needed in SEMAC sequences. Phantoms of six different dental materials with various shapes and sizes that are used in practice (amalgam; titanium implant; gold and Ni-Cr crowns; Ni-Ti and stainless steel orthodontic wires) were imaged. In vivo images of two subjects were also acquired. The amounts of artifact reduction were quantified in phantom studies.
RESULTS: Compared with conventional SE imaging in phantom studies, in-plane artifacts were reduced by up to 43% in the VAT SE images and 80% in the SEMAC images. Through-plane artifacts were reduced by up to 65% in SEMAC images. In vivo SEMAC images also showed reduced artifacts.
CONCLUSION: The SEMAC technique can mitigate artifact caused by metallic dental materials for T1w-SE imaging.
Copyright © 2012 Wiley Periodicals, Inc.

Entities:  

Mesh:

Substances:

Year:  2012        PMID: 22941956     DOI: 10.1002/jmri.23785

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  13 in total

1.  Experience with magnetic resonance imaging of human subjects with passive implants and tattoos at 7 T: a retrospective study.

Authors:  Yacine Noureddine; Andreas K Bitz; Mark E Ladd; Markus Thürling; Susanne C Ladd; Gregor Schaefers; Oliver Kraff
Journal:  MAGMA       Date:  2015-09-26       Impact factor: 2.310

2.  MRI with intraoral orthodontic appliance-a comparative in vitro and in vivo study of image artefacts at 1.5 T.

Authors:  C Zachriat; P Asbach; K I Blankenstein; I Peroz; F H Blankenstein
Journal:  Dentomaxillofac Radiol       Date:  2015-03-03       Impact factor: 2.419

3.  PETRA, MSVAT-SPACE and SEMAC sequences for metal artefact reduction in dental MR imaging.

Authors:  Tim Hilgenfeld; Marcel Prager; Alexander Heil; Franz Sebastian Schwindling; Mathias Nittka; David Grodzki; Peter Rammelsberg; Martin Bendszus; Sabine Heiland
Journal:  Eur Radiol       Date:  2017-07-11       Impact factor: 5.315

4.  MRI artefacts after Bonebridge implantation.

Authors:  C Steinmetz; I Mader; S Arndt; A Aschendorff; R Laszig; F Hassepass
Journal:  Eur Arch Otorhinolaryngol       Date:  2014-03-18       Impact factor: 2.503

5.  Evaluation of magnetic resonance imaging for diagnostic purposes in operative dentistry-a systematic review.

Authors:  Silwan Mendes; Carin A Rinne; Julia C Schmidt; Dorothea Dagassan-Berndt; Clemens Walter
Journal:  Clin Oral Investig       Date:  2019-12-10       Impact factor: 3.573

Review 6.  Principles of Simultaneous PET/MR Imaging.

Authors:  Ciprian Catana
Journal:  Magn Reson Imaging Clin N Am       Date:  2017-05       Impact factor: 2.266

7.  Magnetic resonance imaging artefacts caused by orthodontic appliances and/or implant-supported prosthesis: a systematic review.

Authors:  Katrine Mølgaard Johannsen; João Marcus de Carvalho E Silva Fuglsig; Brian Hansen; Ann Wenzel; Rubens Spin-Neto
Journal:  Oral Radiol       Date:  2022-09-30       Impact factor: 1.882

8.  Hexagonal undersampling for faster MRI near metallic implants.

Authors:  Bragi Sveinsson; Pauline W Worters; Garry E Gold; Brian A Hargreaves
Journal:  Magn Reson Med       Date:  2014-02-18       Impact factor: 4.668

9.  Novel Metal Artifact Reduction Techniques with Use of Slice-Encoding Metal Artifact Correction and View-Angle Tilting MR Imaging for Improved Visualization of Brain Tissue near Intracranial Aneurysm Clips.

Authors:  B Friedrich; M Wostrack; F Ringel; Y-M Ryang; A Förschler; S Waldt; C Zimmer; M Nittka; C Preibisch
Journal:  Clin Neuroradiol       Date:  2014-08-01       Impact factor: 3.649

10.  Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences.

Authors:  Tim Hilgenfeld; Marcel Prager; Franz S Schwindling; Johann M E Jende; Peter Rammelsberg; Martin Bendszus; Sabine Heiland; Alexander Juerchott
Journal:  J Vis Exp       Date:  2018-05-17       Impact factor: 1.355

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.