PURPOSE: This study was performed to evaluate the magnitude of artifacts produced by gutta-percha and metal posts on cone-beam computed tomography (CBCT) scans obtained with different tube currents and with or without metal artifact reduction (MAR).Entities:
Keywords: Artifacts; Cone-Beam Computed Tomography; Gutta-Percha; Metals
Year: 2020 PMID: 32206614 PMCID: PMC7078413 DOI: 10.5624/isd.2020.50.1.1
Source DB: PubMed Journal: Imaging Sci Dent ISSN: 2233-7822
Fig. 1A. A line was drawn perpendicular to the mandibular body and through the center of the artifact-generating object. B. 20°-step lines were drawn above and below the perpendicular line, and 5 concentric semicircles were created with different radii (0.5-cm intervals). C. In the regions of intersection between the lines and the semicircles, 24 square regions of interest (ROIs) of the same size were established. A square ROI of the same size as the others was established in the silicone block, serving as the control area.
Fig. 2Axial images represent intracanal materials (control in the first row, gutta-percha in the second row, and metal post in the third row) with different tube currents (4 mA, 8 mA, and 10 mA) and without and with the application of a metal artifact reduction tool.
Mean standard deviation of gray values by distance for each intracanal material with 3 different tube currents (mA) and with or without the application of a metal artifact reduction (MAR) tool
Different uppercase superscript letters represent statistically significant differences between the SDs of gray values for the tube currents tested within each intracanal material group and considering all distances tested and with or without MAR. Different lowercase superscript letters shown with metal post values represent statistically significant differences in SD values, without MAR, among the distances tested.
*: statistically significant differences in SD values compared to the same distance without MAR.
Contrast-to-noise ratio (CNR) values by distance for each intracanal material with 3 different tube currents (mA) and with or without the application of a metal artifact reduction (MAR) tool
Different uppercase superscript letters represent statistically significant differences between CNR values for the different tube currents tested within each intracanal material group and considering all distances tested with or without MAR. Different lowercase superscript letters shown with metal post values represent statistically significant differences in CNR values, without MAR, among the distances tested.
*: statistically significant differences in CNR values compared to the same distance without MAR.