In vitro assessment of artifacts induced by titanium, titanium-zirconium and zirconium dioxide implants in cone-beam computed tomography.

Manuel Sancho-Puchades1, Christoph H F Hämmerle1, Goran I Benic1.   

Abstract

AIM: The aim of this study was to test whether or not the intensity of artifacts around implants in cone-beam computed tomography (CBCT) differs between titanium, titanium-zirconium and zirconium dioxide implants.
MATERIALS AND METHODS: Twenty models of a human mandible, each containing one implant in the single-tooth gap position 45, were cast in dental stone. Five test models were produced for each of the following implant types: titanium 4.1 mm diameter (Ti4.1 ), titanium 3.3 mm diameter (Ti3.3 ), titanium-zirconium 3.3 mm diameter (TiZr3.3 ) and zirconium dioxide 3.5-4.5 mm diameter (ZrO3.5-4.5 ) implants. For control purposes, three models without implants were produced. Each model was scanned using a CBCT device. Gray values (GV) were recorded at eight circumferential positions around the implants at 0.5 mm, 1 mm and 2 mm from the implant surface (GVT est ). GV were assessed in the corresponding volumes of interest (VOI) in the control models without implants (GVC ontrol ). Differences of gray values (ΔGV) between GVT est and GVC ontrol were calculated as percentages. One-way ANOVA and post hoc tests were applied to detect differences between implant types.
RESULTS: Mean ΔGV for ZrO3.5-4.5 presented the highest absolute values, generally followed by TiZr3.3 , Ti4.1 and Ti3.3 implants. The differences of ΔGV between ZrO3.5-4.5 and the remaining groups were statistically significant in the majority of the VOI (P ≤ 0.0167). ΔGV for TiZr3.3 , Ti4.1 and Ti3.3 implants did not differ significantly in the most VOI. For all implant types, ΔGV showed positive values buccally, mesio-buccally, lingually and disto-lingually, whereas negative values were detected mesially and distally.
CONCLUSIONS: Zirconium dioxide implants generate significantly more artifacts as compared to titanium and titanium-zirconium implants. The intensity of artifacts around zirconium dioxide implants exhibited in average the threefold in comparison with titanium implants.
© 2014 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.

Entities:  

Keywords:  artifact; cone-beam computed tomography; dental implant; in vitro; radiology; titanium; titanium-zirconium; zirconia; zirconium dioxide

Mesh:

Substances:

Year:  2014        PMID: 25040484     DOI: 10.1111/clr.12438

Source DB:  PubMed          Journal:  Clin Oral Implants Res        ISSN: 0905-7161            Impact factor:   5.977


  20 in total

1.  Accuracy of linear measurements around dental implants by means of cone beam computed tomography with different exposure parameters.

Authors:  Lauren O L Bohner; Pedro Tortamano; Juliana Marotti
Journal:  Dentomaxillofac Radiol       Date:  2017-03-23       Impact factor: 2.419

2.  Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts.

Authors:  Deborah Queiroz Freitas; Rocharles Cavalcante Fontenele; Eduarda Helena Leandro Nascimento; Taruska Ventorini Vasconcelos; Marcel Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2018-06-27       Impact factor: 2.419

3.  Evaluation of the Metal Artifact Caused by Dental Implants in Cone Beam Computed Tomography Images.

Authors:  Nazmehr Vahdani; Ehsan Moudi; Fatemeh Ghobadi; Ebrahim Mohammadi; Ali Bijani; Sina Haghanifar
Journal:  Maedica (Bucur)       Date:  2020-06

4.  Mapping the expression of beam hardening artefacts produced by metal posts positioned in different regions of the dental arch.

Authors:  Rocharles C Fontenele; Amanda Farias Gomes; Lucas P Lopes Rosado; Frederico S Neves; Deborah Q Freitas
Journal:  Clin Oral Investig       Date:  2020-08-09       Impact factor: 3.573

5.  Effect of anatomical region on the formation of metal artefacts produced by dental implants in cone beam computed tomographic images.

Authors:  Alessiana Helena Machado; Karolina Aparecida Castilho Fardim; Camila Furtado de Souza; Bruno Salles Sotto-Maior; Neuza Maria Souza Picorelli Assis; Karina Lopes Devito
Journal:  Dentomaxillofac Radiol       Date:  2018-01-10       Impact factor: 2.419

6.  Magnitude of cone beam CT image artifacts related to zirconium and titanium implants: impact on image quality.

Authors:  Rocharles C Fontenele; Eduarda Hl Nascimento; Taruska V Vasconcelos; Marcel Noujeim; Deborah Q Freitas
Journal:  Dentomaxillofac Radiol       Date:  2018-05-10       Impact factor: 2.419

7.  Artefacts in multimodal imaging of titanium, zirconium and binary titanium-zirconium alloy dental implants: an in vitro study.

Authors:  Ralf Smeets; Maximilian Schöllchen; Tobias Gauer; Ghazal Aarabi; Alexandre T Assaf; Carsten Rendenbach; Benedicta Beck-Broichsitter; Jan Semmusch; Jan Sedlacik; Max Heiland; Jens Fiehler; Susanne Siemonsen
Journal:  Dentomaxillofac Radiol       Date:  2016-12-02       Impact factor: 2.419

8.  Difference in the artefacts production and the performance of the metal artefact reduction (MAR) tool between the buccal and lingual cortical plates adjacent to zirconium dental implant.

Authors:  Eduarda Helena Leandro Nascimento; Rocharles Cavalcante Fontenele; Gustavo Machado Santaella; Deborah Queiroz Freitas
Journal:  Dentomaxillofac Radiol       Date:  2019-07-10       Impact factor: 2.419

9.  Artefacts at different distances from titanium and zirconia implants in cone-beam computed tomography: effect of tube current and metal artefact reduction.

Authors:  Arthur Xavier Maseti Mancini; Matheus Urias Cruz Santos; Hugo Gaêta-Araujo; Camila Tirapelli; Ruben Pauwels; Christiano Oliveira-Santos
Journal:  Clin Oral Investig       Date:  2021-02-05       Impact factor: 3.573

10.  Influence of windowing and metal artefact reduction algorithms on the volumetric dimensions of five different high-density materials: a cone-beam CT study.

Authors:  Fernanda Coelho-Silva; Luciano Augusto Cano Martins; Daniela Azeredo Braga; Eliana Zandonade; Francisco Haiter-Neto; Sergio Lins de-Azevedo-Vaz
Journal:  Dentomaxillofac Radiol       Date:  2020-06-25       Impact factor: 2.419

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.