Literature DB >> 33625608

Do the number of basis images and metal artifact reduction affect the production of artifacts near and far from zirconium dental implants in CBCT?

Eduarda Helena Leandro Nascimento1,2, Hugo Gaêta-Araujo1, Rocharles Cavalcante Fontenele3, Nicolly Oliveira-Santos1, Christiano Oliveira-Santos4, Deborah Queiroz Freitas1.   

Abstract

OBJECTIVES: This study aimed to evaluate the influence of the number of basis images and the metal artifact reduction (MAR) tool on the production of artifacts near and far from a zirconium implant in cone-beam computed tomography (CBCT).
MATERIALS AND METHODS: CBCT scans of a mandible were acquired before and after insertion of an implant, using 450 and 720 basis images, with and without MAR activation. The mean and standard deviation (SD) of the gray values of the regions of interest (ROIs) located on the cortices adjacent to the implant and at different distances from it (in the soft tissue) were calculated. The mean of the gray values was used to calculate the absolute contrast difference (ACD) between the control and implant scans.
RESULTS: In general, the number of basis images did not affect the SD and the ACD values of the buccal and lingual ROIs (p > 0.05). The implant increased the SD in the lingual cortical plate (p < 0.05). In this case, MAR activation decreased SD (p < 0.05). All ROIs located at different distances from the implant showed higher SD on scans acquired with 450 basis images (p < 0.05), regardless of MAR condition.
CONCLUSIONS: A higher number of basis images reduces the magnitude of artifacts but does not influence the image quality in bone cortical plates. MAR improves the image in the areas most affected by artifacts. CLINICAL RELEVANCE: The number of basis images is known as a factor capable of influencing the image quality and radiation dose for the patient. Therefore, it is important to investigate its effect on the expression of artifacts in the CBCT images.

Entities:  

Keywords:  Artifacts; Cone-beam computed tomography; Dental implant; Zirconium

Year:  2021        PMID: 33625608     DOI: 10.1007/s00784-021-03836-5

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  26 in total

1.  Evaluation of a metal artefact reduction tool on different positions of a metal object in the FOV.

Authors:  Polyane M Queiroz; Gustavo M Santaella; Thais D J da Paz; Deborah Q Freitas
Journal:  Dentomaxillofac Radiol       Date:  2017-02-17       Impact factor: 2.419

2.  Influence of acquisition parameters on the magnitude of cone beam computed tomography artifacts.

Authors:  Deborah Queiroz Freitas; Rocharles Cavalcante Fontenele; Eduarda Helena Leandro Nascimento; Taruska Ventorini Vasconcelos; Marcel Noujeim
Journal:  Dentomaxillofac Radiol       Date:  2018-06-27       Impact factor: 2.419

3.  Magnitude of cone beam CT image artifacts related to zirconium and titanium implants: impact on image quality.

Authors:  Rocharles C Fontenele; Eduarda Hl Nascimento; Taruska V Vasconcelos; Marcel Noujeim; Deborah Q Freitas
Journal:  Dentomaxillofac Radiol       Date:  2018-05-10       Impact factor: 2.419

4.  Influence of the Intracanal Material and Metal Artifact Reduction Tool in the Detection of the Second Mesiobuccal Canal in Cone-beam Computed Tomographic Examinations.

Authors:  Lucas de Paula Lopes Rosado; Fernanda Bulhões Fagundes; Deborah Queiroz Freitas; Matheus Lima Oliveira; Frederico Sampaio Neves
Journal:  J Endod       Date:  2020-05-08       Impact factor: 4.171

5.  Evaluation of the efficacy of a metal artifact reduction algorithm in different cone beam computed tomography scanning parameters.

Authors:  Polyane Mazucatto Queiroz; Francisco Carlos Groppo; Matheus Lima Oliveira; Francisco Haiter-Neto; Deborah Queiroz Freitas
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol       Date:  2017-03-09

6.  Difference in the artefacts production and the performance of the metal artefact reduction (MAR) tool between the buccal and lingual cortical plates adjacent to zirconium dental implant.

Authors:  Eduarda Helena Leandro Nascimento; Rocharles Cavalcante Fontenele; Gustavo Machado Santaella; Deborah Queiroz Freitas
Journal:  Dentomaxillofac Radiol       Date:  2019-07-10       Impact factor: 2.419

7.  Effect of exposure parameters of cone beam computed tomography on metal artifact reduction around the dental implants in various bone densities.

Authors:  Abbas Shokri; Mohammad Reza Jamalpour; Atefeh Khavid; Zeinab Mohseni; Masoud Sadeghi
Journal:  BMC Med Imaging       Date:  2019-04-29       Impact factor: 1.930

Review 8.  Cone beam computed tomography in implant dentistry: recommendations for clinical use.

Authors:  Reinhilde Jacobs; Benjamin Salmon; Marina Codari; Bassam Hassan; Michael M Bornstein
Journal:  BMC Oral Health       Date:  2018-05-15       Impact factor: 2.757

9.  Effect of object position in the field of view and application of a metal artifact reduction algorithm on the detection of vertical root fractures on cone-beam computed tomography scans: An in vitro study.

Authors:  Ava Nikbin; Zahra Dalili Kajan; Mehran Taramsari; Negar Khosravifard
Journal:  Imaging Sci Dent       Date:  2018-12-20

10.  Magnitude of beam-hardening artifacts produced by gutta-percha and metal posts on cone-beam computed tomography with varying tube current.

Authors:  Hugo Gaêta-Araujo; Eduarda Helena Leandro Nascimento; Rocharles Cavalcante Fontenele; Arthur Xavier Maseti Mancini; Deborah Queiroz Freitas; Christiano Oliveira-Santos
Journal:  Imaging Sci Dent       Date:  2020-03-17
View more
  2 in total

1.  CBCT artefact-burden of zirconia-based as compared to titanium implants for different beam energies: an analytical approach.

Authors:  Ralf Schulze
Journal:  Sci Rep       Date:  2022-09-10       Impact factor: 4.996

2.  Influence of kilovoltage-peak and the metal artifact reduction tool in cone-beam computed tomography on the detection of bone defects around titanium-zirconia and zirconia implants.

Authors:  Rocharles Cavalcante Fontenele; Eduarda Helena Leandro Nascimento; Ana Catarina Imbelloni-Vasconcelos; Luciano Augusto Cano Martins; Andrea Dos Anjos Pontual; Flávia Maria Moraes Ramos-Perez; Deborah Queiroz Freitas
Journal:  Imaging Sci Dent       Date:  2022-07-05
  2 in total