Artefacts at different distances from titanium and zirconia implants in cone-beam computed tomography: effect of tube current and metal artefact reduction.

Arthur Xavier Maseti Mancini1, Matheus Urias Cruz Santos2, Hugo Gaêta-Araujo3, Camila Tirapelli4, Ruben Pauwels5, Christiano Oliveira-Santos2.   

Abstract

OBJECTIVES: To evaluate the effect of cone-beam computed tomography (CBCT) tube current (mA) on the magnitude of artefacts at different distances from titanium or zirconia implants, with and without activation of a proprietary metal artefact reduction (MAR).
MATERIAL AND METHODS: Human mandibles were scanned on an OP300 Maxio CBCT unit (Instrumentarium, Tuusula, Finland) before and after the installation of dental implants, with four different tube currents (4 mA, 6.3 mA, 8 mA and 10 mA), with and without activation of proprietary MAR. The effect of mA on the standard deviation (SD) of gray values and contrast to noise ratio (CNR) were assessed in regions of interest located 1.5 cm, 2.5 cm, and 3.5 cm from implants.
RESULTS: In the presence of titanium implants, a significant decrease in SD was found by increasing tube current from 4 mA to 6.3 mA or 8 mA. For zirconia implants, 8 mA yielded better results for all distances. MAR improved CNR in the presence of zirconia implants at all distances, whereas no differences were observed with the use of MAR for titanium implants.
CONCLUSION: Increased tube current can improve overall image quality in the presence of implants, at all the distances tested. When a zirconia implant is present, such increase in mA should be higher in comparison to that for examinations with titanium implants. Activation of OP300 Maxio proprietary MAR improved image quality only among examinations with zirconia implants. CLINICAL RELEVANCE: Artefact-generating implants are common in the field of view of CBCT examinations. Optimal exposure parameters, such as tube current, ensure high image quality with lowest possible radiation exposure.
© 2021. The Author(s), under exclusive licence to Springer-Verlag GmbH, DE part of Springer Nature.

Entities:  

Keywords:  Artefacts; Cone-beam computed tomography; Titanium implants; Tube current; Zirconia implants

Mesh:

Substances:

Year:  2021        PMID: 33544197     DOI: 10.1007/s00784-021-03821-y

Source DB:  PubMed          Journal:  Clin Oral Investig        ISSN: 1432-6981            Impact factor:   3.573


  23 in total

1.  Optimization of cone beam CT exposure for pre-surgical evaluation of the implant site.

Authors:  A Dawood; J Brown; V Sauret-Jackson; S Purkayastha
Journal:  Dentomaxillofac Radiol       Date:  2011-12-19       Impact factor: 2.419

2.  Effects of tube current on cone-beam computerized tomography image quality for presurgical implant planning in vitro.

Authors:  Jaideep Sur; Kenji Seki; Hiroshi Koizumi; Koh Nakajima; Tomohiro Okano
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2010-07-02

3.  Image quality produced by different cone-beam computed tomography settings.

Authors:  Jeffrey C Kwong; J Martin Palomo; Michael A Landers; Alex Figueroa; Mark G Hans
Journal:  Am J Orthod Dentofacial Orthop       Date:  2008-02       Impact factor: 2.650

4.  American Academy of Oral and Maxillofacial Radiology executive opinion statement on performing and interpreting diagnostic cone beam computed tomography.

Authors:  Laurie Carter; Allan G Farman; James Geist; William C Scarfe; Christos Angelopoulos; Madhu K Nair; Charles F Hildebolt; Donald Tyndall; Michael Shrout
Journal:  Oral Surg Oral Med Oral Pathol Oral Radiol Endod       Date:  2008-10

5.  On cone-beam computed tomography artifacts induced by titanium implants.

Authors:  Ralf Kurt Willy Schulze; Dorothea Berndt; Bernd d'Hoedt
Journal:  Clin Oral Implants Res       Date:  2009-10-21       Impact factor: 5.977

Review 6.  Artefacts in CBCT: a review.

Authors:  R Schulze; U Heil; D Gross; D D Bruellmann; E Dranischnikow; U Schwanecke; E Schoemer
Journal:  Dentomaxillofac Radiol       Date:  2011-07       Impact factor: 2.419

7.  High-kVp Assisted Metal Artifact Reduction for X-ray Computed Tomography.

Authors:  Yan Xi; Yannan Jin; Bruno De Man; Ge Wang
Journal:  IEEE Access       Date:  2016-09-16       Impact factor: 3.367

8.  In vitro assessment of artifacts induced by titanium, titanium-zirconium and zirconium dioxide implants in cone-beam computed tomography.

Authors:  Manuel Sancho-Puchades; Christoph H F Hämmerle; Goran I Benic
Journal:  Clin Oral Implants Res       Date:  2014-07-08       Impact factor: 5.977

9.  Magnitude of cone beam CT image artifacts related to zirconium and titanium implants: impact on image quality.

Authors:  Rocharles C Fontenele; Eduarda Hl Nascimento; Taruska V Vasconcelos; Marcel Noujeim; Deborah Q Freitas
Journal:  Dentomaxillofac Radiol       Date:  2018-05-10       Impact factor: 2.419

Review 10.  Artifacts: The downturn of CBCT image.

Authors:  Anil Kumar Nagarajappa; Neha Dwivedi; Rana Tiwari
Journal:  J Int Soc Prev Community Dent       Date:  2015 Nov-Dec
View more
  3 in total

1.  Does the size of an object containing dental implant affect the expression of artifacts in cone beam computed tomography imaging?

Authors:  Mahkameh Moshfeghi; Yaser Safi; Ingrid Różyło-Kalinowska; Shiva Gandomi
Journal:  Head Face Med       Date:  2022-06-29       Impact factor: 2.246

2.  CBCT artefact-burden of zirconia-based as compared to titanium implants for different beam energies: an analytical approach.

Authors:  Ralf Schulze
Journal:  Sci Rep       Date:  2022-09-10       Impact factor: 4.996

3.  Influence of kilovoltage-peak and the metal artifact reduction tool in cone-beam computed tomography on the detection of bone defects around titanium-zirconia and zirconia implants.

Authors:  Rocharles Cavalcante Fontenele; Eduarda Helena Leandro Nascimento; Ana Catarina Imbelloni-Vasconcelos; Luciano Augusto Cano Martins; Andrea Dos Anjos Pontual; Flávia Maria Moraes Ramos-Perez; Deborah Queiroz Freitas
Journal:  Imaging Sci Dent       Date:  2022-07-05
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.