A Universal Protocol for Abdominal CT Examinations Performed on a Photon-Counting Detector CT System: A Feasibility Study.

Wei Zhou1, Gregory J Michalak, Jayse M Weaver, Hao Gong, Lifeng Yu, Cynthia H McCollough, Shuai Leng.   

Abstract

OBJECTIVE: The aims of this study were to investigate the feasibility of using a universal abdominal acquisition protocol on a photon-counting detector computed tomography (PCD-CT) system and to compare its performance to that of single-energy (SE) and dual-energy (DE) CT using energy-integrating detectors (EIDs).
METHODS: Iodine inserts of various concentrations and sizes were embedded into different sizes of adult abdominal phantoms. Phantoms were scanned on a research PCD-CT and a clinical EID-CT with SE and DE modes. Virtual monoenergetic images (VMIs) were generated from PCD-CT and DE mode of EID-CT. For each image type and phantom size, contrast-to-noise ratio (CNR) was measured for each iodine insert and the area under the receiver operating characteristic curve (AUC) for iodine detectability was calculated using a channelized Hotelling observer. The optimal energy (in kiloelectrovolt) of VMIs was determined separately as the one with highest CNR and the one with the highest AUC. The PCD-CT VMIs at the optimal energy were then compared with DE VMIs and SE images in terms of CNR and AUC.
RESULTS: Virtual monoenergetic image at 50 keV had both the highest CNR and highest AUC for PCD-CT and DECT. For 1.0 mg I/mL iodine and 35 cm phantom, the CNRs of 50 keV VMIs from PCD-CT (2.01 ± 0.67) and DE (1.96 ± 0.52) were significantly higher (P < 0.001, Wilcoxon signed-rank test) than SE images (1.11 ± 0.35). The AUC of PCD-CT (0.98 ± 0.01) was comparable to SE (0.98 ± 0.01), and both were slightly lower than DE (0.99 ± 0.01, P < 0.01, Wilcoxon signed-rank test). A similar trend was observed for other phantom sizes and iodine concentrations.
CONCLUSIONS: Virtual monoenergetic images at a fixed energy from a universal acquisition protocol on PCD-CT demonstrated higher iodine CNR and comparable iodine detectability than SECT images, and similar performance compared with DE VMIs.

Entities:  

Mesh:

Substances:

Year:  2020        PMID: 32049691      PMCID: PMC7241672          DOI: 10.1097/RLI.0000000000000634

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   10.065


  49 in total

1.  Virtual monochromatic imaging in dual-source dual-energy CT: radiation dose and image quality.

Authors:  Lifeng Yu; Jodie A Christner; Shuai Leng; Jia Wang; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2011-12       Impact factor: 4.071

Review 2.  CT of the Abdomen with Reduced Tube Voltage in Adults: A Practical Approach.

Authors:  Adeel R Seyal; Atilla Arslanoglu; Samir F Abboud; Azize Sahin; Jeanne M Horowitz; Vahid Yaghmai
Journal:  Radiographics       Date:  2015-10-16       Impact factor: 5.333

3.  Scout-Based Automated Tube Potential Selection Technique (kV Assist) in Enhanced Chest Computed Tomography: Effects on Radiation Exposure and Image Quality.

Authors:  Meng Li; Shichao Feng; Ning Wu; Li Zhang
Journal:  J Comput Assist Tomogr       Date:  2017 May/Jun       Impact factor: 1.826

4.  Virtual monoenergetic dual-layer, dual-energy CT enterography: optimization of keV settings and its added value for Crohn's disease.

Authors:  Sang Min Lee; Se Hyung Kim; Su Joa Ahn; Hyo-Jin Kang; Ji Hee Kang; Joon Koo Han
Journal:  Eur Radiol       Date:  2018-01-02       Impact factor: 5.315

5.  Eighty-peak kilovoltage 16-channel multidetector computed tomography and reduced contrast-medium doses tailored to body weight to diagnose pulmonary embolism in azotaemic patients.

Authors:  Fredrik Holmquist; Ulf Nyman
Journal:  Eur Radiol       Date:  2006-01-27       Impact factor: 5.315

6.  Impact of Photon Counting Detector Technology on kV Selection and Diagnostic Workflow in CT.

Authors:  Wei Zhou; Dilbar Abdurakhimova; Michael Bruesewitz; Ahmed Halaweish; Cynthia H McCollough; Shuai Leng
Journal:  Proc SPIE Int Soc Opt Eng       Date:  2018-03

7.  Lowering kilovoltage to reduce radiation dose in contrast-enhanced abdominal CT: initial assessment of a prototype automated kilovoltage selection tool.

Authors:  David M Hough; Joel G Fletcher; Katharine L Grant; Jeff L Fidler; Lifeng Yu; Jennifer R Geske; Rickey E Carter; Rainer Raupach; Bernhard Schmidt; Thomas Flohr; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2012-11       Impact factor: 3.959

8.  Distinguishing enhancing from nonenhancing renal masses with dual-source dual-energy CT: iodine quantification versus standard enhancement measurements.

Authors:  Giorgio Ascenti; Achille Mileto; Bernhard Krauss; Michele Gaeta; Alfredo Blandino; Emanuele Scribano; Nicola Settineri; Silvio Mazziotti
Journal:  Eur Radiol       Date:  2013-03-12       Impact factor: 5.315

9.  Correlation between model observer and human observer performance in CT imaging when lesion location is uncertain.

Authors:  Shuai Leng; Lifeng Yu; Yi Zhang; Rickey Carter; Alicia Y Toledano; Cynthia H McCollough
Journal:  Med Phys       Date:  2013-08       Impact factor: 4.071

10.  Evaluation of conventional imaging performance in a research whole-body CT system with a photon-counting detector array.

Authors:  Zhicong Yu; Shuai Leng; Steven M Jorgensen; Zhoubo Li; Ralf Gutjahr; Baiyu Chen; Ahmed F Halaweish; Steffen Kappler; Lifeng Yu; Erik L Ritman; Cynthia H McCollough
Journal:  Phys Med Biol       Date:  2016-02-02       Impact factor: 3.609

View more
  6 in total

1.  Virtual monoenergetic images from dual-energy CT: systematic assessment of task-based image quality performance.

Authors:  Davide Cester; Matthias Eberhard; Hatem Alkadhi; André Euler
Journal:  Quant Imaging Med Surg       Date:  2022-01

2.  Dual-Contrast Biphasic Liver Imaging With Iodine and Gadolinium Using Photon-Counting Detector Computed Tomography: An Exploratory Animal Study.

Authors:  Liqiang Ren; Nathan Huber; Kishore Rajendran; Joel G Fletcher; Cynthia H McCollough; Lifeng Yu
Journal:  Invest Radiol       Date:  2022-02-01       Impact factor: 6.016

Review 3.  An introduction to photon-counting detector CT (PCD CT) for radiologists.

Authors:  Yuko Nakamura; Toru Higaki; Shota Kondo; Ikuo Kawashita; Isao Takahashi; Kazuo Awai
Journal:  Jpn J Radiol       Date:  2022-10-18       Impact factor: 2.701

4.  Quantum Iterative Reconstruction for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lung.

Authors:  Thomas Sartoretti; Damien Racine; Victor Mergen; Lisa Jungblut; Pascal Monnin; Thomas G Flohr; Katharina Martini; Thomas Frauenfelder; Hatem Alkadhi; André Euler
Journal:  Diagnostics (Basel)       Date:  2022-02-18

5.  Simultaneous Dual-Contrast Imaging of Small Bowel With Iodine and Bismuth Using Photon-Counting-Detector Computed Tomography: A Feasibility Animal Study.

Authors:  Liqiang Ren; Kishore Rajendran; Joel G Fletcher; Cynthia H McCollough; Lifeng Yu
Journal:  Invest Radiol       Date:  2020-10       Impact factor: 10.065

6.  Extracellular Volume Quantification With Cardiac Late Enhancement Scanning Using Dual-Source Photon-Counting Detector CT.

Authors:  Victor Mergen; Thomas Sartoretti; Ernst Klotz; Bernhard Schmidt; Lisa Jungblut; Kai Higashigaito; Robert Manka; André Euler; Markus Kasel; Matthias Eberhard; Hatem Alkadhi
Journal:  Invest Radiol       Date:  2022-01-21       Impact factor: 10.065

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.