Scout-Based Automated Tube Potential Selection Technique (kV Assist) in Enhanced Chest Computed Tomography: Effects on Radiation Exposure and Image Quality.

Meng Li1, Shichao Feng, Ning Wu, Li Zhang.   

Abstract

OBJECTIVE: The aim of our study was to assess radiation dose reduction and image quality for enhanced chest CT examinations with a scout-based automated tube potential selection technique (kV Assist) compared with a standard 120-kV protocol.
METHODS: Prospective study of enhanced chest CT examinations was performed in 100 consecutive patients with kV Assist and in 100 consecutive patients with conventional 120-kV protocol on a multislice CT (Discovery CT750 HD). The body mass index, CT dose index volume, and dose length product were recorded from the examination protocol. Image noise and CT value was measured on region of interest, signal-to-noise ratio, and contrast-to-noise ratio was calculated. The subjective image quality was assessed by two radiologists blinded to the respective protocol with the use of a 3-grade scale (3, superior quality; 2, moderate quality; 1, inferior quality).
RESULTS: With kV Assist, the percentages of patients being scanned using 80, 100, and 120 kV were 12.0%, 80.0%, and 8.0%, respectively. The kilovolt setting was related with body mass index (r = 0.565, P = 0.000). Compared with the conventional 120 kV protocol, kV Assist allowed for an overall average decrease of 30.6% in CT dose index volume (kV Assist, 11.05 ± 4.78 mGy; 120 kV, 15.92 ± 6.89 mGy) (P < 0.001) and 32.3% in dose length product (kV Assist, 386.41 ± 184.02 mGy cm; 120 kV, 571.14 ± 286.68 mGy cm) (P < 0.001). In the kV Assist, mean attenuation of regions of interest inside the aorta was significantly higher than that in 120-kV protocols (kV Assist, 310.27 ± 73.70 HU; 120 kV, 239.44 ± 47.65 HU) (P < 0.001), resulting in increased contrast-to-noise ratio (kV Assist, 26.69 ± 7.78; 120 kV, 21.38 ± 6.05) (P < 0.001). There was no significant difference in subjective image quality scores between the 2 groups.
CONCLUSIONS: The use of attenuation-based kV Assist technique enables significant dose reduction in enhanced chest CT scan while improving arterial enhancement and preserving image quality at adequate levels.

Entities:  

Mesh:

Year:  2017        PMID: 27759600     DOI: 10.1097/RCT.0000000000000523

Source DB:  PubMed          Journal:  J Comput Assist Tomogr        ISSN: 0363-8715            Impact factor:   1.826


  3 in total

1.  Improving image quality with model-based iterative reconstruction algorithm for chest CT in children with reduced contrast concentration.

Authors:  Jihang Sun; Di Hu; Yun Shen; Haiming Yang; Chenghao Chen; Jie Yin; Yun Peng
Journal:  Radiol Med       Date:  2019-02-09       Impact factor: 3.469

2.  Automatic Tube Current Modulation and Tube Voltage Selection in Pediatric Computed Tomography: A Phantom Study on Radiation Dose and Image Quality.

Authors:  Antonios E Papadakis; John Damilakis
Journal:  Invest Radiol       Date:  2019-05       Impact factor: 6.016

3.  A Universal Protocol for Abdominal CT Examinations Performed on a Photon-Counting Detector CT System: A Feasibility Study.

Authors:  Wei Zhou; Gregory J Michalak; Jayse M Weaver; Hao Gong; Lifeng Yu; Cynthia H McCollough; Shuai Leng
Journal:  Invest Radiol       Date:  2020-04       Impact factor: 10.065

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.