PURPOSE: The aims of this study were to develop a single-scan dual-contrast protocol for biphasic liver imaging with 2 intravenous contrast agents (iodine and gadolinium) and to evaluate its effectiveness in an exploratory swine study using a photon-counting detector computed tomography (PCD-CT) system.
MATERIALS AND METHODS: A dual-contrast CT protocol was developed for PCD-CT to simultaneously acquire 2 phases of liver contrast enhancement, with the late arterial phase enhanced by 1 contrast agent (iodine-based) and the portal venous phase enhanced by the other (gadolinium-based). A gadolinium contrast bolus (gadobutrol: 64 mL, 8 mL/s) and an iodine contrast bolus (iohexol: 40 mL, 5 mL/s) were intravenously injected in the femoral vein of a healthy domestic swine, with the second injection initiated after 17 seconds from the beginning of the first injection; PCD-CT image acquisition was performed 12 seconds after the beginning of the iodine contrast injection. A convolutional neural network (CNN)-based denoising technique was applied to PCD-CT images to overcome the inherent noise magnification issue in iodine/gadolinium decomposition task. Iodine and gadolinium material maps were generated using a 3-material decomposition method in image space. A set of contrast samples (mixed iodine and gadolinium) was attached to the swine belly; quantitative accuracy of material decomposition in these inserts between measured and true concentrations was calculated using root mean square error. An abdominal radiologist qualitatively evaluated the delineation of arterial and venous vasculatures in the swine liver using iodine and gadolinium maps obtained using the dual-contrast PCD-CT protocol.
RESULTS: The iodine and gadolinium samples attached to the swine were quantified with root mean square error values of 0.75 mg/mL for iodine and 0.45 mg/mL for gadolinium from the contrast material maps derived from the denoised PCD-CT images. Hepatic arteries containing iodine and veins containing gadolinium in the swine liver could be clearly visualized. Compared with the original images, better distinctions between 2 liver phases were achieved using CNN denoising, with approximately 60% to 80% noise reduction in contrast material maps acquired with the denoised PCD-CT images compared with the original images.
CONCLUSIONS: Simultaneous biphasic liver imaging in a single multienergy PCD-CT acquisition using a dual-contrast (iodine and gadolinium) injection protocol and CNN denoising was demonstrated in a swine study, where the enhanced hepatic arteries (containing iodine) and the enhanced hepatic veins (containing gadolinium) could be clearly visualized and delineated in the swine liver.
Copyright © 2021 Wolters Kluwer Health, Inc. All rights reserved.

Entities:  

Mesh:

Substances:

Year:  2022        PMID: 34411033      PMCID: PMC8732294          DOI: 10.1097/RLI.0000000000000815

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  34 in total

Review 1.  Helical biphasic contrast-enhanced CT of the liver: technique, indications, interpretation, and pitfalls.

Authors:  J H Oliver; R L Baron
Journal:  Radiology       Date:  1996-10       Impact factor: 11.105

2.  Impact of prior information on material decomposition in dual- and multienergy computed tomography.

Authors:  Liqiang Ren; Shengzhen Tao; Kishore Rajendran; Cynthia H McCollough; Lifeng Yu
Journal:  J Med Imaging (Bellingham)       Date:  2019-03-14

3.  Material decomposition with prior knowledge aware iterative denoising (MD-PKAID).

Authors:  Shengzhen Tao; Kishore Rajendran; Cynthia H McCollough; Shuai Leng
Journal:  Phys Med Biol       Date:  2018-09-21       Impact factor: 3.609

4.  Radiation dose efficiency of multi-energy photon-counting-detector CT for dual-contrast imaging.

Authors:  Liqiang Ren; Kishore Rajendran; Cynthia H McCollough; Lifeng Yu
Journal:  Phys Med Biol       Date:  2019-12-13       Impact factor: 3.609

5.  Comparison of a Photon-Counting-Detector CT with an Energy-Integrating-Detector CT for Temporal Bone Imaging: A Cadaveric Study.

Authors:  W Zhou; J I Lane; M L Carlson; M R Bruesewitz; R J Witte; K K Koeller; L J Eckel; R E Carter; C H McCollough; S Leng
Journal:  AJNR Am J Neuroradiol       Date:  2018-08-09       Impact factor: 3.825

6.  Future of Diagnostic Computed Tomography: An Update on Physicochemical Properties, Safety, and Development of X-ray Contrast Media.

Authors:  Imran Shahid; Eric Lancelot; Pierre Desché
Journal:  Invest Radiol       Date:  2020-09       Impact factor: 6.016

7.  How Low Can We Go in Radiation Dose for the Data-Completion Scan on a Research Whole-Body Photon-Counting Computed Tomography System.

Authors:  Zhicong Yu; Shuai Leng; Zhoubo Li; Ahmed F Halaweish; Steffen Kappler; Erik L Ritman; Cynthia H McCollough
Journal:  J Comput Assist Tomogr       Date:  2016 Jul-Aug       Impact factor: 1.826

8.  Deep-learning-based direct inversion for material decomposition.

Authors:  Hao Gong; Shengzhen Tao; Kishore Rajendran; Wei Zhou; Cynthia H McCollough; Shuai Leng
Journal:  Med Phys       Date:  2020-10-30       Impact factor: 4.071

9.  Improved Peritoneal Cavity and Abdominal Organ Imaging Using a Biphasic Contrast Agent Protocol and Spectral Photon Counting Computed Tomography K-Edge Imaging.

Authors:  Salim Si-Mohamed; Arnaud Thivolet; Pierre-Emmanuel Bonnot; Daniel Bar-Ness; Vahan Képénékian; David P Cormode; Philippe Douek; Pascal Rousset
Journal:  Invest Radiol       Date:  2018-10       Impact factor: 6.016

10.  Multicolour imaging with spectral photon-counting CT: a phantom study.

Authors:  Salim Si-Mohamed; Daniel Bar-Ness; Monica Sigovan; Valérie Tatard-Leitman; David P Cormode; Pratap C Naha; Philippe Coulon; Lucie Rascle; Ewald Roessl; Michal Rokni; Ami Altman; Yoad Yagil; Loic Boussel; Philippe Douek
Journal:  Eur Radiol Exp       Date:  2018-10-17
View more
  3 in total

1.  Simultaneous dual-contrast imaging using energy-integrating detector multi-energy CT: An in vivo feasibility study.

Authors:  Zhongxing Zhou; Liqiang Ren; Kishore Rajendran; Felix E Diehn; Joel G Fletcher; Cynthia H McCollough; Lifeng Yu
Journal:  Med Phys       Date:  2022-01-27       Impact factor: 4.071

Review 2.  An introduction to photon-counting detector CT (PCD CT) for radiologists.

Authors:  Yuko Nakamura; Toru Higaki; Shota Kondo; Ikuo Kawashita; Isao Takahashi; Kazuo Awai
Journal:  Jpn J Radiol       Date:  2022-10-18       Impact factor: 2.701

3.  Extracellular Volume Quantification With Cardiac Late Enhancement Scanning Using Dual-Source Photon-Counting Detector CT.

Authors:  Victor Mergen; Thomas Sartoretti; Ernst Klotz; Bernhard Schmidt; Lisa Jungblut; Kai Higashigaito; Robert Manka; André Euler; Markus Kasel; Matthias Eberhard; Hatem Alkadhi
Journal:  Invest Radiol       Date:  2022-01-21       Impact factor: 10.065

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.