Virtual monoenergetic images from dual-energy CT: systematic assessment of task-based image quality performance.

Davide Cester1, Matthias Eberhard1, Hatem Alkadhi1, André Euler1.   

Abstract

BACKGROUND: To compare task-based image quality (TB-IQ) among virtual monoenergetic images (VMI) and linear-blended images (LBI) from dual-energy CT as a function of contrast task, radiation dose, size, and lesion diameter.
METHODS: A TB-IQ phantom (Mercury Phantom 4.0, Sun Nuclear Corporation) was imaged on a third-generation dual-source dual-energy CT with 100/Sn150 kVp at three volume CT dose levels (5, 10, 15 mGy). Three size sections (diameters 16, 26, 36 cm) with subsections for image noise and spatial resolution analysis were used. High-contrast tasks (e.g., calcium-containing stone and vascular lesion) were emulated using bone and iodine inserts. A low-contrast task (e.g., low-contrast lesion or hematoma) was emulated using a polystyrene insert. VMI at 40-190 keV and LBI were reconstructed. Noise power spectrum (NPS) determined the noise magnitude and texture. Spatial resolution was assessed using the task-transfer function (TTF) of the three inserts. The detectability index (d') served as TB-IQ metric.
RESULTS: Noise magnitude increased with increasing phantom size, decreasing dose, and decreasing VMI-energy. Overall, noise magnitude was higher for VMI at 40-60 keV compared to LBI (range of noise increase, 3-124%). Blotchier noise texture was found for low and high VMIs (40-60 keV, 130-190 keV) compared to LBI. No difference in spatial resolution was observed for high contrast tasks. d' increased with increasing dose level or lesion diameter and decreasing size. For high-contrast tasks, d' was higher at 40-80 keV and lower at high VMIs. For the low-contrast task, d' was higher for VMI at 70-90 keV and lower at 40-60 keV.
CONCLUSIONS: Task-based image quality differed among VMI-energy and LBI dependent on the contrast task, dose level, phantom size, and lesion diameter. Image quality could be optimized by tailoring VMI-energy to the contrast task. Considering the clinical relevance of iodine, VMIs at 50-60 keV could be proposed as an alternative to LBI. 2022 Quantitative Imaging in Medicine and Surgery. All rights reserved.

Entities:  

Keywords:  Dual-energy; Phantoms, Imaging; image reconstruction; task-based image quality assessment; virtual monoenergetic images (VMI)

Year:  2022        PMID: 34993114      PMCID: PMC8666791          DOI: 10.21037/qims-21-477

Source DB:  PubMed          Journal:  Quant Imaging Med Surg        ISSN: 2223-4306


  43 in total

1.  Performance evaluation of computed tomography systems: Summary of AAPM Task Group 233.

Authors:  Ehsan Samei; Donovan Bakalyar; Kirsten L Boedeker; Samuel Brady; Jiahua Fan; Shuai Leng; Kyle J Myers; Lucretiu M Popescu; Juan Carlos Ramirez Giraldo; Frank Ranallo; Justin Solomon; Jay Vaishnav; Jia Wang
Journal:  Med Phys       Date:  2019-09-11       Impact factor: 4.071

2.  Impact of dose reduction and the use of an advanced model-based iterative reconstruction algorithm on spectral performance of a dual-source CT system: A task-based image quality assessment.

Authors:  Joël Greffier; Djamel Dabli; Aymeric Hamard; Philippe Akessoul; Asmaa Belaouni; Jean-Paul Beregi; Julien Frandon
Journal:  Diagn Interv Imaging       Date:  2021-04-02       Impact factor: 4.026

3.  Comparison of low-contrast detectability between two CT reconstruction algorithms using voxel-based 3D printed textured phantoms.

Authors:  Justin Solomon; Alexandre Ba; François Bochud; Ehsan Samei
Journal:  Med Phys       Date:  2016-12       Impact factor: 4.071

4.  Improving iodine contrast to noise ratio using virtual monoenergetic imaging and prior-knowledge-aware iterative denoising (mono-PKAID).

Authors:  Shengzhen Tao; Kishore Rajendran; Wei Zhou; Joel G Fletcher; Cynthia H McCollough; Shuai Leng
Journal:  Phys Med Biol       Date:  2019-05-16       Impact factor: 3.609

5.  Principles and applications of multienergy CT: Report of AAPM Task Group 291.

Authors:  Cynthia H McCollough; Kirsten Boedeker; Dianna Cody; Xinhui Duan; Thomas Flohr; Sandra S Halliburton; Jiang Hsieh; Rick R Layman; Norbert J Pelc
Journal:  Med Phys       Date:  2020-05-28       Impact factor: 4.071

6.  Can dual-energy CT replace perfusion CT for the functional evaluation of advanced hepatocellular carcinoma?

Authors:  Sébastien Mulé; Frédéric Pigneur; Ronan Quelever; Arthur Tenenhaus; Laurence Baranes; Philippe Richard; Vania Tacher; Edouard Herin; Hugo Pasquier; Maxime Ronot; Alain Rahmouni; Valérie Vilgrain; Alain Luciani
Journal:  Eur Radiol       Date:  2017-11-22       Impact factor: 5.315

7.  Technical Note: Quantitative accuracy evaluation for spectral images from a detector-based spectral CT scanner using an iodine phantom.

Authors:  Xinhui Duan; Gary Arbique; Jeffrey Guild; Yin Xi; Jon Anderson
Journal:  Med Phys       Date:  2018-03-24       Impact factor: 4.071

8.  Estimating detectability index in vivo: development and validation of an automated methodology.

Authors:  Taylor Brunton Smith; Justin Solomon; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2017-12-11

9.  Myocardial Delayed Enhancement CT for the Evaluation of Heart Failure: Comparison to MRI.

Authors:  Yasutoshi Ohta; Shinichiro Kitao; Hiroto Yunaga; Shinya Fujii; Natsuko Mukai; Kazuhiro Yamamoto; Toshihide Ogawa
Journal:  Radiology       Date:  2018-07-10       Impact factor: 11.105

10.  Iodine and Fat Quantification for Differentiation of Adrenal Gland Adenomas From Metastases Using Third-Generation Dual-Source Dual-Energy Computed Tomography.

Authors:  Simon S Martin; Sebastian Weidinger; Rouben Czwikla; Benjamin Kaltenbach; Moritz H Albrecht; Lukas Lenga; Thomas J Vogl; Julian L Wichmann
Journal:  Invest Radiol       Date:  2018-03       Impact factor: 6.016

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.