| Literature DB >> 31949530 |
Kapil Golvankar1, Mohammed Abdul Kader1, Abhilash Abdul Latheef1, Ahamed Babiker Mohammed Ali1, Shahabe Saquib Abullais1, Mohammed Sibagathullah1.
Abstract
AIM: The present in-vitro study aims to compare the accuracy of root canal working length determination between the third generation and fourth generation electronic apex locators.Entities:
Keywords: Apex locators; Apical foramen; Root canal; Working length
Year: 2019 PMID: 31949530 PMCID: PMC6953924 DOI: 10.3889/oamjms.2019.696
Source DB: PubMed Journal: Open Access Maced J Med Sci ISSN: 1857-9655
Figure 1Fifty single-rooted teeth with definite coronal plane
Figure 2Actual working length determination under the stereomicroscope
Figure 3Apparatus for working length determination using apex locators
Mean ± SD values of actual length and length determined by apex locators
| GROUP | N | Mean ± SD (mm) | Minimum (mm) | Maximum (mm) |
|---|---|---|---|---|
| Group AL | 50 | 14.76 ± 1.84 | 11.64 | 20.74 |
| Group EL1 | 50 | 14.72 ± 1.85 | 11.51 | 20.72 |
| Group EL2 | 50 | 14.66 ± 1.84 | 11.62 | 20.70 |
*AL = Actual length determination under stereomicroscope;
EL1=Electronic working length determination by third-generation electronic apex locator;
EL2=Electronic working length determination by fourth-generation electronic apex locator.
Results of One-way ANOVA test comparing the three groups
| GROUP | N | Mean ± SD | F | p-value |
|---|---|---|---|---|
| Group AL | 50 | 14.76 + 1.84 | 0.0370 | p = 0.963 |
| Group EL1 | 50 | 14.72 + 1.85 | ||
| Group EL2 | 50 | 14.66 + 1.85 |
€Statistically significant difference p ≤ 0.05;
AL = Actual length determination under stereomicroscope;
EL1=Electronic working length determination by third-generation electronic apex locator;
EL2=Electronic working length determination by fourth-generation electronic apex locator.
Results of Independent ‘t’ test comparing the individual groups
| Comparing groups | p-value | |
|---|---|---|
| Group EL1 and Group AL | 0.0989 | p = 0.9214 |
| Group EL2 and Group AL | 0.269 | p = 0.7882 |
| Group EL1 and Group EL2 | 0.170 | p = 0.8656 |
€Statistically significant difference p < 0.05; *AL = Actual length determination under stereomicroscope; **EL1=Electronic working length determination by third generation electronic apex locator; ***EL2=Electronic working length determination by fourth generation electronic apex locator.