In vitro evaluation of the reliability of the Root ZX electronic apex locator.

H F Ounsi1, A Naaman.   

Abstract

AIM: An experimental comparison of the Root ZX apex locator with the real measurement of the root canal length was carried out using a saline gel to simulate the periodontium.
METHODOLOGY: Lengths were taken when the needle reached the 0.5 mark and the Apex mark on the meter. These measurements were then compared with a calculated reference length representing the real length of the root canal. Reliability was assessed using Student's t-test. Precision was assessed using the mean and standard deviation of differences.
RESULTS: The Root ZX gave measurements within a range of 0.5 mm in 84.72% of the cases. The intra-operator and inter-operator variabilities were not statistically significant (P > 0.05).
CONCLUSIONS: The Root ZX is not capable of detecting the '0.5 mm from the foramen' position and thus, should only be used to detect the foramen (major diameter).

Entities:  

Mesh:

Year:  1999        PMID: 10371907     DOI: 10.1046/j.1365-2591.1999.00202.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  10 in total

1.  Clinical evaluation of Root ZX II electronic apex locator in primary teeth.

Authors:  S Saritha; K S Uloopi; C Vinay; R Chandra Sekhar; V V Rao
Journal:  Eur Arch Paediatr Dent       Date:  2012-02

2.  Comparison of two electronic apex locators on human cadavers.

Authors:  Luca Marigo; Gian Luca Gervasi; Francesco Somma; Giuseppe Squeo; Raffaella Castagnola
Journal:  Clin Oral Investig       Date:  2015-11-04       Impact factor: 3.573

3.  An in vitro evaluation of the accuracy of four electronic apex locators using stainless-steel and nickel-titanium hand files.

Authors:  Paras Mull Gehlot; Vinutha Manjunath; Mysore Krishnaswamy Manjunath
Journal:  Restor Dent Endod       Date:  2016-01-04

4.  A comparative assessment of the accuracies of Raypex 5, Raypex 6, iPex and iPex II electronic apex locators: An in vitro study.

Authors:  Mugem Asli Gurel; Bagdagul Helvacioglu Kivanc; Adil Ekici
Journal:  J Istanb Univ Fac Dent       Date:  2017-01-02

5.  Influence of embedding media on the accuracy of working length determination by means of apex locator: an ex vivo study.

Authors:  Thomas Gerhard Wolf; Anna Krauß-Mironjuk; Richard Johannes Wierichs; Benjamín Briseño-Marroquín
Journal:  Sci Rep       Date:  2021-02-08       Impact factor: 4.379

6.  The effect of file size on the accuracy of the raypex 5 apex locator: an in vitro study.

Authors:  Shiva Sadeghi; Masoomeh Abolghasemi
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2008-05-15

7.  Comparison between two methods of working length determination and its effect on radiographic extent of root canal filling: a clinical study [ISRCTN71486641].

Authors:  L Smadi
Journal:  BMC Oral Health       Date:  2006-02-11       Impact factor: 2.757

8.  Working Length Determination of Root Canal of Young Permanent Tooth: An In vitro Study.

Authors:  A Diwanji; As Rathore; R Arora; V Dhar; A Madhusudan; J Doshi
Journal:  Ann Med Health Sci Res       Date:  2014-07

9.  Influence of Root Canal Curvature on the Accuracy of Root ZX Electronic Foramen Locator: An In Vitro Study.

Authors:  Masoud Saatchi; Shiva Iravani; Mehrdad Akhavan Khaleghi; Amin Mortaheb
Journal:  Iran Endod J       Date:  2017

10.  Comparison of Accuracy in Determining the Root Canal Working Length by Using Two Generations of Apex Locators - An In Vitro Study.

Authors:  Kapil Golvankar; Mohammed Abdul Kader; Abhilash Abdul Latheef; Ahamed Babiker Mohammed Ali; Shahabe Saquib Abullais; Mohammed Sibagathullah
Journal:  Open Access Maced J Med Sci       Date:  2019-09-11
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.