Literature DB >> 18498892

Influence of instrument size on the accuracy of different apex locators: an in vitro study.

Benjamín Briseño-Marroquín1, Santiago Frajlich, Fernando Goldberg, Brita Willershausen.   

Abstract

The aim of this in vitro investigation was to determine the accuracy of 4 different electronic apex locators (EALs) with 3 different instrument sizes. For this study 146 roots were embedded in an agar solution. Electronic measurements were made to the physiologic foramen (apical constriction) with the Elements Apex Locator, Justy II, Raypex 5, and ProPex II and K-type files sizes 08, 10, and 15. Statistical significances were calculated with the sign test (P < .001). Exact measurements to the physiologic foramen were made with the Elements Apex Locator, 36.99%, 39.04%, and 44.93%; Justy II, 38.62%, 32.41%, and 43.41%; Raypex 5, 42.76%, 39.31%, and 39.06%; and ProPex II, 38.62%, 43.45%, and 40.63% of the time with instrument sizes 08, 10, and 15, respectively. No significant differences were found between the actual working length and EALs/instrument size. A nonsignificant higher number of unstable measurements were observed in all EALs with instrument size 15.

Entities:  

Mesh:

Year:  2008        PMID: 18498892     DOI: 10.1016/j.joen.2008.02.019

Source DB:  PubMed          Journal:  J Endod        ISSN: 0099-2399            Impact factor:   4.171


  15 in total

1.  Consistency of electronic measurements of endodontic working length when using multiple devices from the same manufacturer-an in vitro study.

Authors:  Franziska Haupt; M Hülsmann
Journal:  Clin Oral Investig       Date:  2018-03-07       Impact factor: 3.573

2.  Effect of pre-flaring and file size on the accuracy of two electronic apex locators.

Authors:  Manoel Brito-Júnior; Carla Cristina Camilo; Gil Moreira-Júnior; Jesus Djalma Pecora; Manoel Damião Sousa-Neto
Journal:  J Appl Oral Sci       Date:  2012 Sep-Oct       Impact factor: 2.698

3.  Accuracy of electronic apex locators to detect root canal perforations with inserted metallic posts: an ex vivo study.

Authors:  Benjamín Briseño Marroquín; Claudia Cortazar Fernández; Irene Schmidtmann; Brita Willershausen; Fernando Goldberg
Journal:  Head Face Med       Date:  2014-12-23       Impact factor: 2.151

4.  In vitro evaluation of Root ZX and Raypex 6 in teeth with different apical diameters.

Authors:  Ugur Aydin; Emrah Karataslioglu; Fatih Aksoy; Cihan Yildirim
Journal:  J Conserv Dent       Date:  2015 Jan-Feb

5.  In vitro comparison of working length determination using three different electronic apex locators.

Authors:  Alper Kuştarci; Dilara Arslan; Demet Altunbaş
Journal:  Dent Res J (Isfahan)       Date:  2014-09

6.  The Effect of File Size and Type and Irrigation Solutions on the Accuracy of Electronic Apex Locators: An In Vitro Study on Canine Teeth.

Authors:  Maciej Janeczek; Piotr Kosior; Dagmara Piesiak-Pańczyszyn; Krzysztof Dudek; Aleksander Chrószcz; Agnieszka Czajczyńska-Waszkiewicz; Małgorzata Kowalczyk-Zając; Aleksandra Gabren-Syller; Karol Kirstein; Aleksandra Skalec; Ewelina Bryła; Maciej Dobrzyński
Journal:  Biomed Res Int       Date:  2016-09-26       Impact factor: 3.411

7.  Influence of calcium hydroxide residues after using different irrigants on the accuracy of two electronic apex locators: An in vitro study.

Authors:  Nooshin Sadat Shojaee; Zahra Zaeri; Mohammad Mehdi Shokouhi; Fereshteh Sobhnamayan; Alireza Adl
Journal:  Dent Res J (Isfahan)       Date:  2020-01-21

8.  An In Vitro Comparison of Propex II Apex Locator to Standard Radiographic Method.

Authors:  Kalyan Vinayak Chakravarthy Pishipati
Journal:  Iran Endod J       Date:  2013-08-01

9.  Effects of dissolving solutions on the accuracy of an electronic apex locator-integrated endodontic handpiece.

Authors:  Yakup Ustun; Ozgur Uzun; Ozgur Er; Murat Maden; Fatma Yalpı; Burhan Can Canakci
Journal:  ScientificWorldJournal       Date:  2013-11-26

10.  An in vivo comparison of accuracy of two electronic apex locators in determining working length using stainless steel and nickel titanium files.

Authors:  Sanjeeb Chaudhary; Archana Gharti; Bhawana Adhikari
Journal:  Clin Cosmet Investig Dent       Date:  2018-05-22