| Literature DB >> 31805912 |
Gwang Chul Lee1, Sang Hong Lee1, Sin Wook Kang1, Hyung Seok Park1, Suenghwan Jo2.
Abstract
BACKGROUND: Several methods using simple anteroposterior (AP) radiographs have been suggested for the measurement of anteversion of the cup component after total hip arthroplasty. Herein, we compared six widely used anteversion measurement methods using two different types of AP radiograph, the conventional pelvis AP and hip-centered AP radiographs, to identify the measurement method and the type of radiograph that would provide the highest accuracy and reliability.Entities:
Keywords: Anteversion; Hip-centered anteroposterior; Inclination; Pelvis anteroposterior; Total hip replacement
Mesh:
Year: 2019 PMID: 31805912 PMCID: PMC6896281 DOI: 10.1186/s12891-019-2979-0
Source DB: PubMed Journal: BMC Musculoskelet Disord ISSN: 1471-2474 Impact factor: 2.362
Fig. 1Custom-made bi-planar anteversion measurement device allowing inclination and anteversion of the mounted cup with protractors in two axes. The solid arrow indicates the change in anteversion, while the hollow arrow indicates the change in inclination
Fig. 2Two custom-made bi-planar anteversion measurement models fixed to plexiglass. The distance between two models represents the distance from the center of the triangle formed by both anterior superior iliac spine (ASIS) and symphysis pubis and the hip joint
Fig. 3Methods for measuring anteversion on plane anteroposterior radiographs. a Pradhan et al. (arcsin (p/0.4D)), b Lewinnek et al. (arcsin (D1/D2)), c Widmer et al. (arcsin([S]/[TL]), d Liaw et al. (sin-1 tan β), e Hassan et al. ([arcsin [(h/D) / √([m/D] – [m2/D2])]], f Ackland et al. (arcsin [2y / 2√(2Dx – x2)])
Inter-observer and intra-observer reliabilities of measurement
| Measurement method | Inter-observer reliability | Intra-observer reliability | ||
|---|---|---|---|---|
| ICC | 95% CI | ICC | 95% CI | |
| Pradhan et al. [ | 0.938 | 0.901–0.958 | 0.934 | 0.897–0.952 |
| Lewinnek et al. [ | 0.937 | 0.897–0.967 | 0.916 | 0.845–0.974 |
| Widmer et al. [ | 0.928 | 0.884–0.956 | 0.933 | 0.866–0.966 |
| Liaw et al. [ | 0.887 | 0.835–0.906 | 0.908 | 0.839–0.929 |
| Hassan et al. [ | 0.902 | 0.873–0.937 | 0.899 | 0.879–0.941 |
| Ackland et al. [ | 0.886 | 0.847–0.923 | 0.913 | 0.868–0.933 |
ICC intraclass correlation coefficient, CI confidence interval
Differences between measured and reference anteversions in pelvis and hip-centered AP radiographs
| Measurement method | Mean ± SD difference from reference anteversion | Maximum difference from reference anteversion | ||
|---|---|---|---|---|
| Pradhan et al. [ | Pelvis AP | 1.62 ± 1.54 | 5.43 | 0.845 |
| Hip-centered AP | −3.11 ± 1.14 | −4.83 | 0.629 | |
| Lewinnek et al. [ | Pelvis AP | 2.84 ± 1.59 | 6.61 | 0.358 |
| Hip-centered AP | −2.91 ± 1.56 | −5.34 | 0.521 | |
| Widmer et al. [ | Pelvis AP | −7.10 ± 3.96 | −14.46 | 0.061 |
| Hip-centered AP | −15.66 ± 3.80 | −21.13 | < 0.001 | |
| Liaw et al. [ | Pelvis AP | 2.52 ± 1.59 | 6.42 | 0.763 |
| Hip-centered AP | 1.37 ± 1.73 | 4.31 | 0.692 | |
| Hassan et al. [ | Pelvis AP | − 12.95 ± 9.71 | −33.97 | < 0.001 |
| Hip-centered AP | −18.11 ± 7.59 | −30.9 | < 0.001 | |
| Ackland et al. [ | Pelvis AP | −10.15 ± 6.73 | − 24.24 | < 0.001 |
| Hip-centered AP | −13.87 ± 5.48 | −28.89 | < 0.001 | |
AP anteroposterior, SD standard deviation
Differences between measured and reference anteversions in pelvis and hip-centered AP radiographs within a safe inclination zone (30 °to 50 ° )
| Measurement method | Mean ± SD difference from reference anteversion | Maximum difference from reference anteversion | ||
|---|---|---|---|---|
| Pradhan et al. [ | Pelvis AP | 1.23 ± 0.92 | 2.90 | 0.429 |
| Hip-centered AP | −2.66 ± 1.14 | −4.83 | 0.752 | |
| Lewinnek et al. [ | Pelvis AP | 2.46 ± 0.98 | 3.92 | 0.235 |
| Hip-centered AP | −2.65 ± 1.12 | −4.69 | 0.367 | |
| Widmer et al. [ | Pelvis AP | −7.61 ± 3.74 | −14.46 | < 0.001 |
| Hip-centered AP | −14.99 ± 4.12 | −20.64 | < 0.001 | |
| Liaw et al. [ | Pelvis AP | 2.13 ± 0.96 | 3.35 | 0.523 |
| Hip-centered AP | 1.76 ± 1.55 | 4.31 | 0.847 | |
| Hassan et al. [ | Pelvis AP | −11.65 ± 8.92 | −29.35 | < 0.001 |
| Hip-centered AP | −17.61 ± 8.04 | −29.42 | < 0.001 | |
| Ackland et al. [ | Pelvis AP | −10.84 ± 6.98 | −24.22 | < 0.001 |
| Hip-centered AP | −13.56 ± 5.30 | −28.89 | < 0.001 | |
AP anteroposterior, SD standard deviation
Accuracy of each anteversion measurement method compared to reference anteversion using Pearson’s correlation
| Correlation coefficient | |||
|---|---|---|---|
| Reference | |||
| Pradhan et al. [ | Pelvis AP | 0.958 | < 0.001 |
| Hip-centered AP | 0.973 | < 0.001 | |
| Lewinnek et al. [ | Pelvis AP | 0.962 | < 0.001 |
| Hip-centered AP | 0.968 | < 0.001 | |
| Widmer et al. [ | Pelvis AP | 0.851 | < 0.001 |
| Hip-centered AP | 0.791 | < 0.001 | |
| Liaw et al. [ | Pelvis AP | 0.964 | < 0.001 |
| Hip-centered AP | 0.969 | < 0.001 | |
| Hassan et al. [ | Pelvis AP | 0.575 | < 0.001 |
| Hip-centered AP | 0.610 | < 0.001 | |
| Ackland et al. [ | Pelvis AP | 0.774 | < 0.001 |
| Hip-centered AP | 0.779 | < 0.001 | |
AP anteroposterior
Fig. 4Bland–Altman plots showing differences between anteversion measurements using pelvis AP and hip-centered AP radiographs
Reference radiographs and planes in previous literature measuring anteversion in plain AP radiographs
| Measurement method | Type of radiograph | Reference plane |
|---|---|---|
| Pradhan et al. [ | Hip-centered AP | Planar |
| Lewinnek et al. [ | Not specified | Not specified |
| Widmer et al. [ | Pelvis AP | Planar |
| Liaw et al. [ | Hip-centered AP | Planar, True |
| Hassan et al. [ | Hip-centered AP | Planar |
| Ackland et al. [ | Hip-centered AP | Planar, True |
AP anteroposterior