Lucia Paganini1, Giacomo Reggiori2, Antonella Stravato1, Valentina Palumbo1, Pietro Mancosu1, Francesca Lobefalo1, Anna Gaudino1, Antonella Fogliata1, Marta Scorsetti1,3, Stefano Tomatis1.
Abstract
BACKGROUND: PRIMO is a graphical environment based on PENELOPE Monte Carlo (MC) simulation of radiotherapy beams able to compute dose distribution in patients, from plans with different techniques. The dosimetric characteristics of an HD-120 MLC (Varian), simulated using PRIMO, were here compared with measurements, and also with Acuros calculations (in the Eclipse treatment planning system, Varian).Entities:
Mesh:
Year: 2019 PMID: 31791355 PMCID: PMC6889207 DOI: 10.1186/s13014-019-1421-y
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Tests for static and dynamic MLC used for this work. Main geometrical characteristics and relative measurements are reported
| Test | Jaws (cm) | MLC | Measurements | Analysis | |
|---|---|---|---|---|---|
| Static | 10 × 10 | Square 2,3,5 and 10 cm side | microDiamond | Point dose | |
| 10 × 10 | Fully closed on the central axis | Film | Profiles | ||
| 10 × 20 | Film | Profiles | |||
| Dynamic | DLG-test | 10 × 10 | Gap: 2,4,6,10,14,16, 20 mm moving from left to right Transmitted dose (bank A closed and then bank B closed) | microDiamond | Point dose |
| a-SG | 10 × 22 | Gap: 5, 10, 20, 30 mm TG: 0, 1 | Film | Profiles | |
| a-OSG | 6 × 8 | Gap: 10, 30 mm TG: 0, 1 | Film | Profiles |
Fig. 1Configuration for the MLC bank a and bank b in the MLC_alternate test
Fig. 2Example of MLC configuration in four of the a-SG tests where it is possible to see different gap values and different TG fractions [15]
HU and mass density ranges used in PRIMO and Acuros computations
| Material | PRIMO HU range | PRIMO mass density range (g/cm3) | Acuros HU range | Acuros mass density range (g/cm3) |
|---|---|---|---|---|
| Air | − 1000, − 957 | 0, 0.0204 | −1000, − 957 | 0, 0.0204 |
| Lung | −957, − 400 | 0.0204, 0.594 | − 967, − 374 | 0.0104, 0.624 |
| Adipose Tissue | − 400, − 59 | 0.594, 0.969 | − 434, 1 | 0.551, 1.001 |
| Muscle, Skeletal | − 59, 88 | 0.969, 1.075 | −59, 117 | 0.969, 1.093 |
| Cartilage | 88, 298 | 1.075, 1.199 | 57, 971 | 1.056, 1.600 |
| Bone | 298, 2832 | 1.199, 2.833 | 128, 2832 | 1.100, 2.830 |
OF relative to the MLC_square test
| Field size (cm2) | Measured OF | PRIMO OF (diff.) | Acuros OF (diff.) |
|---|---|---|---|
| 2 × 2 | 0.852 | 0.857 (0.5%) | 0.862 (1.2%) |
| 3 × 3 | 0.902 | 0.905 (0.3%) | 0.907 (0.6%) |
| 5 × 5 | 0.946 | 0.944 (−0.3%) | 0.947 (0.0%) |
Fig. 3Profiles for the MLC_trasm test extracted under a leaf at 0.1 cm parallel to the leaves direction
Fig. 4Dose profiles for the MLC_alternate test for the dose distributions obtained with PRIMO (blue), film measurements (green) and Acuros (red). All profiles are plotted in absolute dose
DLG values obtained with experimental measurements, with TPS calculations and with PRIMO and relative uncertainties
| DLG (cm) | |
|---|---|
| MicroDiamond | 0.032 ± 0.006 |
| PRIMO | 0.033 ± 0.003 |
| Acuros | 0.038 ± 0.004 |
Fig. 5Point measurements for the different plans used in the procedure for the evaluation of DLG. Measured points are fitted with a line, whose intercept on the x axes gives the DLG value. The plot region where the intercept is visible is zoomed in the top box
Fig. 6Inline dose profiles for different gap sizes and different TG fractions for PRIMO (blue), Acuros (red) and measurements with Gafchromic films (green)
Fig. 7Gamma Agreement Index between dose distributions relative to the a-SG test calculated with PRIMO and Acuros inside the entire MULTIcube volume for gap sizes of 5, 10, 20 and 30 and TG fractions of 0, 0.4 and 1
Fig. 8Dose distributions for the a-OSG test with the 30 mm gap for PRIMO at the isocenter level for TG fraction = 0 is illustrated in the left plane (a). The dose profiles along the straight line depicted in (b) are given for TG fractions 0 and 1