Monte Carlo modeling and simulations of the High Definition (HD120) micro MLC and validation against measurements for a 6 MV beam.

C Borges1, M Zarza-Moreno, E Heath, N Teixeira, P Vaz.   

Abstract

PURPOSE: The most recent Varian(®) micro multileaf collimator (MLC), the High Definition (HD120) MLC, was modeled using the BEAMNRC Monte Carlo code. This model was incorporated into a Varian medical linear accelerator, for a 6 MV beam, in static and dynamic mode. The model was validated by comparing simulated profiles with measurements.
METHODS: The Varian(®) Trilogy(®) (2300C/D) accelerator model was accurately implemented using the state-of-the-art Monte Carlo simulation program BEAMNRC and validated against off-axis and depth dose profiles measured using ionization chambers, by adjusting the energy and the full width at half maximum (FWHM) of the initial electron beam. The HD120 MLC was modeled by developing a new BEAMNRC component module (CM), designated HDMLC, adapting the available DYNVMLC CM and incorporating the specific characteristics of this new micro MLC. The leaf dimensions were provided by the manufacturer. The geometry was visualized by tracing particles through the CM and recording their position when a leaf boundary is crossed. The leaf material density and abutting air gap between leaves were adjusted in order to obtain a good agreement between the simulated leakage profiles and EBT2 film measurements performed in a solid water phantom. To validate the HDMLC implementation, additional MLC static patterns were also simulated and compared to additional measurements. Furthermore, the ability to simulate dynamic MLC fields was implemented in the HDMLC CM. The simulation results of these fields were compared with EBT2 film measurements performed in a solid water phantom.
RESULTS: Overall, the discrepancies, with and without MLC, between the opened field simulations and the measurements using ionization chambers in a water phantom, for the off-axis profiles are below 2% and in depth-dose profiles are below 2% after the maximum dose depth and below 4% in the build-up region. On the conditions of these simulations, this tungsten-based MLC has a density of 18.7 g cm(- 3) and an overall leakage of about 1.1 ± 0.03%. The discrepancies between the film measured and simulated closed and blocked fields are below 2% and 8%, respectively. Other measurements were performed for alternated leaf patterns and the agreement is satisfactory (to within 4%). The dynamic mode for this MLC was implemented and the discrepancies between film measurements and simulations are within 4%.
CONCLUSIONS: The Varian(®) Trilogy(®) (2300 C/D) linear accelerator including the HD120 MLC was successfully modeled and simulated using the Monte Carlo BEAMNRC code by developing an independent CM, the HDMLC CM, either in static and dynamic modes.

Entities:  

Mesh:

Year:  2012        PMID: 22225311     DOI: 10.1118/1.3671935

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  9 in total

1.  Monte Carlo Modeling of the Agility MLC for IMRT and VMAT Calculations.

Authors:  Shingo Ohira; Hideki Takegawa; Masayoshi Miyazaki; Masahiko Koizumi; Teruki Teshima
Journal:  In Vivo       Date:  2020 Sep-Oct       Impact factor: 2.155

2.  Monte Carlo modeling of a Novalis Tx Varian 6 MV with HD-120 multileaf collimator.

Authors:  Luis Alberto Vazquez-Quino; Brian Massingill; Chengyu Shi; Alonso Gutierrez; Carlos Esquivel; Tony Eng; Nikos Papanikolaou; Sotirios Stathakis
Journal:  J Appl Clin Med Phys       Date:  2012-09-06       Impact factor: 2.102

3.  Monte carlo model and output factors of elekta infinity™ 6 and 10 MV photon beam.

Authors:  Sitti Yani; Indra Budiansah; Mohamad Fahdillah Rhani; Freddy Haryanto
Journal:  Rep Pract Oncol Radiother       Date:  2020-04-28

4.  Development of a dose verification system for Vero4DRT using Monte Carlo method.

Authors:  Yoshitomo Ishihara; Akira Sawada; Mitsuhiro Nakamura; Yuki Miyabe; Hiroaki Tanabe; Shuji Kaneko; Kenji Takayama; Takashi Mizowaki; Masaki Kokubo; Masahiro Hiraoka
Journal:  J Appl Clin Med Phys       Date:  2014-11-08       Impact factor: 2.102

5.  Dosimetric characteristics of LinaTech DMLC H multi leaf collimator: Monte Carlo simulation and experimental study.

Authors:  Mikaeil Molazadeh; Ahad Zeinali; Mostafa Robatjazi; Alireza Shirazi; Ghazale Geraily
Journal:  J Appl Clin Med Phys       Date:  2017-03-06       Impact factor: 2.102

6.  Comparison of 3DCRT and IMRT out-of-field doses in pediatric patients using Monte Carlo simulations with treatment planning system calculations and measurements.

Authors:  Ana Cravo Sá; Andreia Barateiro; Bryan P Bednarz; Pedro Almeida; Pedro Vaz; Tiago Madaleno
Journal:  Front Oncol       Date:  2022-08-05       Impact factor: 5.738

7.  AAPM MEDICAL PHYSICS PRACTICE GUIDELINE 5.b: Commissioning and QA of treatment planning dose calculations-Megavoltage photon and electron beams.

Authors:  Mark W Geurts; Dustin J Jacqmin; Lindsay E Jones; Stephen F Kry; Dimitris N Mihailidis; Jared D Ohrt; Timothy Ritter; Jennifer B Smilowitz; Nicholai E Wingreen
Journal:  J Appl Clin Med Phys       Date:  2022-08-10       Impact factor: 2.243

8.  Monte Carlo modeling of HD120 multileaf collimator on Varian TrueBeam linear accelerator for verification of 6X and 6X FFF VMAT SABR treatment plans.

Authors:  Alanah M Bergman; Ermias Gete; Cheryl Duzenli; Tony Teke
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

9.  MLC parameters from static fields to VMAT plans: an evaluation in a RT-dedicated MC environment (PRIMO).

Authors:  Lucia Paganini; Giacomo Reggiori; Antonella Stravato; Valentina Palumbo; Pietro Mancosu; Francesca Lobefalo; Anna Gaudino; Antonella Fogliata; Marta Scorsetti; Stefano Tomatis
Journal:  Radiat Oncol       Date:  2019-12-02       Impact factor: 3.481