| Literature DB >> 31780707 |
Luca Brombal1,2, Fulvia Arfelli1,2, Pasquale Delogu3,4, Sandro Donato1,2, Giovanni Mettivier5,6, Koen Michielsen7, Piernicola Oliva8,9, Angelo Taibi10,11, Ioannis Sechopoulos7,12, Renata Longo13,14, Christian Fedon2,7.
Abstract
In this study we compared the image quality of a synchrotron radiation (SR) breast computed tomography (BCT) system with a clinical BCT in terms of contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), noise power spectrum (NPS), spatial resolution and detail visibility. A breast phantom consisting of several slabs of breast-adipose equivalent material with different embedded targets (i.e., masses, fibers and calcifications) was used. Phantom images were acquired using a dedicated BCT system installed at the Radboud University Medical Center (Nijmegen, The Netherlands) and the SR BCT system at the SYRMEP beamline of Elettra SR facility (Trieste, Italy) based on a photon-counting detector. Images with the SR setup were acquired mimicking the clinical BCT conditions (i.e., energy of 30 keV and radiation dose of 6.5 mGy). Images were reconstructed with an isotropic cubic voxel of 273 µm for the clinical BCT, while for the SR setup two phase-retrieval (PhR) kernels (referred to as "smooth" and "sharp") were alternatively applied to each projection before tomographic reconstruction, with voxel size of 57 × 57 × 50 µm3. The CNR for the clinical BCT system can be up to 2-times higher than SR system, while the SNR can be 3-times lower than SR system, when the smooth PhR is used. The peak frequency of the NPS for the SR BCT is 2 to 4-times higher (0.9 mm-1 and 1.4 mm-1 with smooth and sharp PhR, respectively) than the clinical BCT (0.4 mm-1). The spatial resolution (MTF10%) was estimated to be 1.3 lp/mm for the clinical BCT, and 5.0 lp/mm and 6.7 lp/mm for the SR BCT with the smooth and sharp PhR, respectively. The smallest fiber visible in the SR BCT has a diameter of 0.15 mm, while for the clinical BCT is 0.41 mm. Calcification clusters with diameter of 0.13 mm are visible in the SR BCT, while the smallest diameter for the clinical BCT is 0.29 mm. As expected, the image quality of the SR BCT outperforms the clinical BCT system, providing images with higher spatial resolution and SNR, and with finer granularity. Nevertheless, this study assesses the image quality gap quantitatively, giving indications on the benefits associated with SR BCT and providing a benchmarking basis for its clinical implementation. In addition, SR-based studies can provide a gold-standard in terms of achievable image quality, constituting an upper-limit to the potential clinical development of a given technique.Entities:
Mesh:
Year: 2019 PMID: 31780707 PMCID: PMC6882794 DOI: 10.1038/s41598-019-54131-z
Source DB: PubMed Journal: Sci Rep ISSN: 2045-2322 Impact factor: 4.379
Figure 1(a) Photograph of the phantom and (b) phantom dimensions [in mm]. (c) Details embedded in slab #9: calcifications (CaCO3) in red circles, masses in blue stars and fibers in green rectangles. The drawings are not to scale.
Figure 2(a) Region of interest position for the contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) evaluation. (b) Region of interest position for the noise power spectrum (NPS) evaluation in a homogeneous background.
Figure 3(a) Contrast-to-noise ratio and (b) signal-to-noise ratio as a function of mass diameter for the clinical breast CT (red solid line) and synchrotron breast CT with smooth (blue dashed lines) and sharp (blue solid lines) phase-retrieval kernels. Of note, in both plots, the point relative to the largest mass scanned with the clinical system produces a higher-than-expected CNR/SNR due to a reconstruction artifact (see Supplementary Materials Fig. S3).
Figure 4Bi-dimensional normalized noise power spectra (nNPS) for (a) the clinical BCT system, synchrotron BCT with (b) smooth and (c) sharp phase-retrieval algorithm. Of note, the range of frequency axes in (a) is different from (b,c). The inset on the top-left corner of each image represent the same homogeneous ROI with an area of 20×20 mm2. (d) Radial averaged nNPS for the clinical system (dashed red line), and SR breast CT with smooth (dashed blue line) and sharp (solid blue line) phase-retrieval algorithm. Of note, the left y-axis refers to the nNPS of the clinical system, while the right y-axis to the synchrotron breast CT. The shaded region around each line represents one standard deviation uncertainty.
Figure 5Evaluation of the spatial resolution for the clinical system (red circles), and synchrotron breast CT with smooth (blue squares) and sharp (blue-white triangles) phase-retrieval algorithm. The logarithm of the squared absolute value of the image Fourier transform (on y-axis) is plotted as function of the square of the spatial frequency (on x-axis). The linear fit for each dataset is shown with black lines. The inset displays a zoom at lower spatial frequencies.
Summary of the comparison analysis between the two systems: clinical BCT and synchrotron radiation (SR) datasets (with smooth and sharp phase-retrieval (PhR) kernel).
| CNR | SNR | nNPS peak frequency (mm−1) | FWHM (mm) | MTF10% (lp/mm) | |
|---|---|---|---|---|---|
| Clinical BCT | 5.2 | 48 | 0.4 | 0.61 | 1.3 |
| SR smooth PhR | 2.3 (1 slice) | 105 | 0.9 | 0.16 | 5.0 |
| 3.0 (5 slices) | 135 | ||||
| SR sharp PhR | 1.2 (1 slice) | 55 | 1.4 | 0.12 | 6.7 |
| 1.7 (5 slices) | 76 |
The table reports the values for contrast-to-noise ratio (CNR) and signal-to-noise ratio (SNR) for the 4.76 mm diameter mass, the peak frequency of the normalized noise power spectrum (nNPS), full width at half maximum (FWHM) of the estimated point spread function (PSF) and modulation transfer function at 10% (MTF10%). CNR and SNR are measured by selecting 1 slice and the average of 5 slices, while the other values can be found in Supplementary Materials (Tab. S1).
Figure 6Details of the fibers reconstructed (a–d) with the clinical BCT system, (e–h) smooth and (i–l) sharp phase-retrieval (PhR) kernels for the synchrotron BCT. The scale reference is reported on the upper right corner of image (a).
Figure 7Details of the calcification clusters reconstructed (a–d) with the clinical BCT system, (e–h) smooth and (i–l) phase-retrieval (PhR) kernels for the synchrotron BCT.