Towards breast tomography with synchrotron radiation at Elettra: first images.

R Longo1, F Arfelli, R Bellazzini, U Bottigli, A Brez, F Brun, A Brunetti, P Delogu, F Di Lillo, D Dreossi, V Fanti, C Fedon, B Golosio, N Lanconelli, G Mettivier, M Minuti, P Oliva, M Pinchera, L Rigon, P Russo, A Sarno, G Spandre, G Tromba, F Zanconati.   

Abstract

The aim of the SYRMA-CT collaboration is to set-up the first clinical trial of phase-contrast breast CT with synchrotron radiation (SR). In order to combine high image quality and low delivered dose a number of innovative elements are merged: a CdTe single photon counting detector, state-of-the-art CT reconstruction and phase retrieval algorithms. To facilitate an accurate exam optimization, a Monte Carlo model was developed for dose calculation using GEANT4. In this study, high isotropic spatial resolution (120 μm)(3) CT scans of objects with dimensions and attenuation similar to a human breast were acquired, delivering mean glandular doses in the range of those delivered in clinical breast CT (5-25 mGy). Due to the spatial coherence of the SR beam and the long distance between sample and detector, the images contain, not only absorption, but also phase information from the samples. The application of a phase-retrieval procedure increases the contrast-to-noise ratio of the tomographic images, while the contrast remains almost constant. After applying the simultaneous algebraic reconstruction technique to low-dose phase-retrieved data sets (about 5 mGy) with a reduced number of projections, the spatial resolution was found to be equal to filtered back projection utilizing a four fold higher dose, while the contrast-to-noise ratio was reduced by 30%. These first results indicate the feasibility of clinical breast CT with SR.

Entities:  

Mesh:

Year:  2016        PMID: 26836274     DOI: 10.1088/0031-9155/61/4/1634

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  21 in total

1.  Monochromatic breast computed tomography with synchrotron radiation: phase-contrast and phase-retrieved image comparison and full-volume reconstruction.

Authors:  Luca Brombal; Bruno Golosio; Fulvia Arfelli; Deborah Bonazza; Adriano Contillo; Pasquale Delogu; Sandro Donato; Giovanni Mettivier; Piernicola Oliva; Luigi Rigon; Angelo Taibi; Giuliana Tromba; Fabrizio Zanconati; Renata Longo
Journal:  J Med Imaging (Bellingham)       Date:  2018-11-27

2.  Development of 3D patient-based super-resolution digital breast phantoms using machine learning.

Authors:  Marco Caballo; Christian Fedon; Luca Brombal; Ritse Mann; Renata Longo; Ioannis Sechopoulos
Journal:  Phys Med Biol       Date:  2018-11-12       Impact factor: 3.609

3.  Dose and diagnostic performance comparison between phase-contrast mammography with synchrotron radiation and digital mammography: a clinical study report.

Authors:  Christian Fedon; Luigi Rigon; Fulvia Arfelli; Diego Dreossi; Elisa Quai; Maura Tonutti; Giuliana Tromba; Maria Assunta Cova; Renata Longo
Journal:  J Med Imaging (Bellingham)       Date:  2018-02-01

4.  Radiochromic film dosimetry in synchrotron radiation breast computed tomography: a phantom study.

Authors:  Giovanni Mettivier; Marica Masi; Fulvia Arfelli; Luca Brombal; Pasquale Delogu; Francesca Di Lillo; Sandro Donato; Christian Fedon; Bruno Golosio; Piernicola Oliva; Luigi Rigon; Antonio Sarno; Angelo Taibi; Paolo Russo
Journal:  J Synchrotron Radiat       Date:  2020-04-22       Impact factor: 2.616

5.  Comparison of propagation-based CT using synchrotron radiation and conventional cone-beam CT for breast imaging.

Authors:  Seyedamir Tavakoli Taba; Patrycja Baran; Yakov I Nesterets; Serena Pacile; Susanne Wienbeck; Christian Dullin; Konstantin Pavlov; Anton Maksimenko; Darren Lockie; Sheridan C Mayo; Harry M Quiney; Diego Dreossi; Fulvia Arfelli; Giuliana Tromba; Sarah Lewis; Timur E Gureyev; Patrick C Brennan
Journal:  Eur Radiol       Date:  2020-01-23       Impact factor: 5.315

6.  Optimization of the energy for Breast monochromatic absorption X-ray Computed Tomography.

Authors:  Pasquale Delogu; Vittorio Di Trapani; Luca Brombal; Giovanni Mettivier; Angelo Taibi; Piernicola Oliva
Journal:  Sci Rep       Date:  2019-09-11       Impact factor: 4.379

7.  Evaluation of the BreastSimulator software platform for breast tomography.

Authors:  G Mettivier; K Bliznakova; I Sechopoulos; J M Boone; F Di Lillo; A Sarno; R Castriconi; P Russo
Journal:  Phys Med Biol       Date:  2017-07-20       Impact factor: 3.609

Review 8.  Review of quantitative multiscale imaging of breast cancer.

Authors:  Michael A Pinkert; Lonie R Salkowski; Patricia J Keely; Timothy J Hall; Walter F Block; Kevin W Eliceiri
Journal:  J Med Imaging (Bellingham)       Date:  2018-01-22

Review 9.  Dedicated breast CT: state of the art-Part I. Historical evolution and technical aspects.

Authors:  Yueqiang Zhu; Avice M O'Connell; Yue Ma; Aidi Liu; Haijie Li; Yuwei Zhang; Xiaohua Zhang; Zhaoxiang Ye
Journal:  Eur Radiol       Date:  2021-08-03       Impact factor: 7.034

10.  Motion artifacts assessment and correction using optical tracking in synchrotron radiation breast CT.

Authors:  Luca Brombal; Lucia Mariel Arana Peña; Fulvia Arfelli; Renata Longo; Francesco Brun; Adriano Contillo; Francesca Di Lillo; Giuliana Tromba; Vittorio Di Trapani; Sandro Donato; Ralf Hendrik Menk; Luigi Rigon
Journal:  Med Phys       Date:  2021-07-29       Impact factor: 4.506

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.