Glandular dose in breast computed tomography with synchrotron radiation.

G Mettivier1, C Fedon, F Di Lillo, R Longo, A Sarno, G Tromba, P Russo.   

Abstract

The purpose of this work is to provide an evaluation of the mean glandular dose (MGD) for breast computed tomography (CT) with synchrotron radiation in an axial scanning configuration with a partial or total organ volume irradiation, for the in vivo program of breast CT ongoing at the ELETTRA facility (Trieste, Italy). A Geant4 Monte Carlo code was implemented, simulating the photon irradiation from a synchrotron radiation source in the energetic range from 8 to 50 keV with 1 keV intervals, to evaluate the MGD. The code was validated with literature data, in terms of mammographic normalized glandular dose coefficients (DgN) and with ad hoc experimental data, in terms of computed tomography dose index (CTDI). Simulated cylindrical phantoms of different sizes (diameter at phantom base 8, 10, 12, 14 or 16 cm, axial length 1.5 times the radius) and glandular fraction by weight (0%, 14.3%, 25%, 50%, 75% and 100%) were implemented into the code. The validation of the code shows an excellent agreement both with previously published work and in terms of DgN and CDTI measurements. The implemented simulations show a dependence of the glandular dose estimate on the vertical dimension of the irradiated zone when a partial organ irradiation was implemented. Specific normalized coefficients for calculating the MGD to the whole breast or to the single irradiated slice were reported.

Entities:  

Mesh:

Year:  2015        PMID: 26683710     DOI: 10.1088/0031-9155/61/2/569

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  Monochromatic breast computed tomography with synchrotron radiation: phase-contrast and phase-retrieved image comparison and full-volume reconstruction.

Authors:  Luca Brombal; Bruno Golosio; Fulvia Arfelli; Deborah Bonazza; Adriano Contillo; Pasquale Delogu; Sandro Donato; Giovanni Mettivier; Piernicola Oliva; Luigi Rigon; Angelo Taibi; Giuliana Tromba; Fabrizio Zanconati; Renata Longo
Journal:  J Med Imaging (Bellingham)       Date:  2018-11-27

2.  A Monte Carlo model for mean glandular dose evaluation in spot compression mammography.

Authors:  Antonio Sarno; David R Dance; Ruben E van Engen; Kenneth C Young; Paolo Russo; Francesca Di Lillo; Giovanni Mettivier; Kristina Bliznakova; Baowei Fei; Ioannis Sechopoulos
Journal:  Med Phys       Date:  2017-06-13       Impact factor: 4.071

3.  Radiochromic film dosimetry in synchrotron radiation breast computed tomography: a phantom study.

Authors:  Giovanni Mettivier; Marica Masi; Fulvia Arfelli; Luca Brombal; Pasquale Delogu; Francesca Di Lillo; Sandro Donato; Christian Fedon; Bruno Golosio; Piernicola Oliva; Luigi Rigon; Antonio Sarno; Angelo Taibi; Paolo Russo
Journal:  J Synchrotron Radiat       Date:  2020-04-22       Impact factor: 2.616

4.  Comparison of propagation-based CT using synchrotron radiation and conventional cone-beam CT for breast imaging.

Authors:  Seyedamir Tavakoli Taba; Patrycja Baran; Yakov I Nesterets; Serena Pacile; Susanne Wienbeck; Christian Dullin; Konstantin Pavlov; Anton Maksimenko; Darren Lockie; Sheridan C Mayo; Harry M Quiney; Diego Dreossi; Fulvia Arfelli; Giuliana Tromba; Sarah Lewis; Timur E Gureyev; Patrick C Brennan
Journal:  Eur Radiol       Date:  2020-01-23       Impact factor: 5.315

5.  Optimization of the energy for Breast monochromatic absorption X-ray Computed Tomography.

Authors:  Pasquale Delogu; Vittorio Di Trapani; Luca Brombal; Giovanni Mettivier; Angelo Taibi; Piernicola Oliva
Journal:  Sci Rep       Date:  2019-09-11       Impact factor: 4.379

6.  Evaluation of the BreastSimulator software platform for breast tomography.

Authors:  G Mettivier; K Bliznakova; I Sechopoulos; J M Boone; F Di Lillo; A Sarno; R Castriconi; P Russo
Journal:  Phys Med Biol       Date:  2017-07-20       Impact factor: 3.609

7.  Radiation dosimetry of a clinical prototype dedicated cone-beam breast CT system with offset detector.

Authors:  Hsin Wu Tseng; Andrew Karellas; Srinivasan Vedantham
Journal:  Med Phys       Date:  2021-01-26       Impact factor: 4.506

8.  A proposal for a quality control protocol in breast CT with synchrotron radiation.

Authors:  Adriano Contillo; Anna Veronese; Luca Brombal; Sandro Donato; Luigi Rigon; Angelo Taibi; Giuliana Tromba; Renata Longo; Fulvia Arfelli
Journal:  Radiol Oncol       Date:  2018-03-24       Impact factor: 2.991

9.  Experimental optimization of the energy for breast-CT with synchrotron radiation.

Authors:  Piernicola Oliva; Vittorio Di Trapani; Fulvia Arfelli; Luca Brombal; Sandro Donato; Bruno Golosio; Renata Longo; Giovanni Mettivier; Luigi Rigon; Angelo Taibi; Giuliana Tromba; Fabrizio Zanconati; Pasquale Delogu
Journal:  Sci Rep       Date:  2020-10-15       Impact factor: 4.379

10.  Motion artifacts assessment and correction using optical tracking in synchrotron radiation breast CT.

Authors:  Luca Brombal; Lucia Mariel Arana Peña; Fulvia Arfelli; Renata Longo; Francesco Brun; Adriano Contillo; Francesca Di Lillo; Giuliana Tromba; Vittorio Di Trapani; Sandro Donato; Ralf Hendrik Menk; Luigi Rigon
Journal:  Med Phys       Date:  2021-07-29       Impact factor: 4.506

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.