Antonia Grimm1, Moritz Winkelmann1, Jakob Weiß1, Georg Gohla1, Gunnar Blumenstock2, Konstantin Nikolaou1, Stephan Clasen1, Rüdiger Hoffmann3.
Abstract
BACKGROUND: We evaluated a magnetic resonance (MR)-conditional high-power microwave ablation system.Entities:
Keywords: Ablation techniques; Cattle; Liver; Magnetic resonance imaging; Microwaves
Year: 2019 PMID: 31549338 PMCID: PMC6757083 DOI: 10.1186/s41747-019-0115-4
Source DB: PubMed Journal: Eur Radiol Exp ISSN: 2509-9280
Fig. 1Cross-section of an ablation zone after 15 min ablation at 120 W. The long axis diameter (LAD) and short-axis diameter (SAD) are shown
Artefact diameters and tip location error of 36 artefact measurements
| Sequence | Angulation | Orientation | Encoding direction | Total | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|
| T1W-VIBE | T1W-FLASH | T2W-TSE | 0° | 45° | 90° | Axial | Coronal | Sagittal | Phase | Frequency | ||
| 3.7 ± 2.4 | 6.3 ± 3.4 | 3.1 ± 1.8 | 1.4 ± 0.6 | 4.9 ± 2.1 | 6.7 ± 2.4 | 4.5 ± 2.3 | 4.0 ± 3.3 | 4.5 ± 3.3 | 4.3 ± 3.0 | 4.4 ± 2.9 | 4.4 ± 2.9 | |
| 1.3 ± 0.8 | 2.0 ± 0.9 | 1.7 ± 0.4 | 1.2 ± 0.8 | 1.6 ± 0.6 | 2.1 ± 0.5 | 2.1 ± 0.7 | 1.4 ± 0.6 | 1.4 ± 0.7 | 1.7 ± 0.8 | 1.6 ± 0.7 | 1.6 ± 0.7 | |
| TLE (mm) | -2.6 ± 1.4 | -2.5 ± 0.2 | -2.6 ± 0.8 | -3.6 ± 0.6 | -2.4 ± 0.5 | -2.1 ± 0.8 | _ | -2.6 ± 0.9 | -2.5 ± 1.0 | -2.7 ± 0.8 | 2.4 ± 1.1 | -2.6 ± 1.0 |
A Artefact diameter at the antenna shaft, A Artefact diameter at the antenna tip, TLE Tip12 location error
Fig. 2Statistical analysis regarding the influence of the sequence type, applicator angulation to B0 and slice orientation on the applicator artefact. Graphs display: Ashaft (artefact diameter at the antenna shaft), Atip (artefact diameter at the antenna tip), and TLE (tip location error) of the microwave applicator in relation to sequence type, applicator angulation to B0 and slice orientation. Depending on < 2 or > 2 parameters Student’s t test (STT) or ANOVA was performed. In case of overall statistical significance, the Student t test was performed according to the closed testing procedure. Whiskers indicate the minimum and maximum extreme values; the box indicates the upper and lower quartile. The line in the box indicates the median value
Fig. 3Artefact of the microwave applicator. Images are acquired under 90° applicator angulation relative to B0 with T1W-VIBE (a), T1W-FLASH (b), and T2W-TSE (c) (see text for sequence details). For all sequences, the applicator shows a similar artefact pattern with a prominent artefact at the shaft (Ashaft) and a smaller artefact at the tip (Atip). PED Phase encoding direction, FED Frequency encoding direction
Fig. 4Artefact of the MW applicator according to the angulation to B0. Images are acquired with T1W-VIBE sequence (see text for sequence details) with an applicator angulation to B0 of 0°(a), 45°(b), and 90°(c). The diameter of shaft artefact was significantly dependent on the angulation with the smallest diameter with an angulation parallel to B0. PED Phase encoding direction, FED Frequency encoding direction
Comparison of the 24 ablation results at 80 W and 120 W
| 80 W | 120 W | ||
|---|---|---|---|
| 5 min | |||
| SAD (mm) | 27.25 ± 2.63 | 32.0 ± 0.82 | 0.0136* |
| Volume (cm3) | 14.75 ± 3.44 | 23.93 ± 2.47 | 0.0049* |
| SI | 0.73 ± 0.08 | 0.72 ± 0.02 | 0.8324 |
| 10 min | |||
| SAD (mm) | 33.5 ± 1.29 | 38.0 ± 1.63 | 0.0050* |
| Volume (cm3) | 26.45 ± 1.76 | 39.03 ± 3.69 | 0.0008* |
| SI | 0.75 ± 0.05 | 0.74 ± 0.03 | 0.7835 |
| 15 min | |||
| SAD (mm) | 38.50 ± 1.91 | 42.0 ± 1.41 | 0.0259* |
| Volume (cm3) | 40.15 ± 5.09 | 56.78 ± 3.08 | 0.0014* |
| SI | 0.75 ± 0.02 | 0.68 ± 0.05 | 0.0480* |
SAD Short axis diameter, SI Sphericity index
*Significant differences