| Literature DB >> 32699978 |
Moritz T Winkelmann1, Georg Gohla1, Jens Kübler1, Jakob Weiß1, Stephan Clasen1, Konstantin Nikolaou1, Rüdiger Hoffmann2.
Abstract
PURPOSE: Evaluation of technique effectiveness, patient safety and ablation parameters of MR-guided microwave ablation in hepatic malignancies using an MR-conditional high-power microwave ablation system.Entities:
Keywords: Ablation techniques; Duration of therapy; Interventional magnetic resonance imaging; Liver neoplasms; Microwave ablation
Mesh:
Year: 2020 PMID: 32699978 PMCID: PMC7591402 DOI: 10.1007/s00270-020-02584-5
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740
Fig. 1Sixty-one-year-old female patient with hepatic metastases from a neuroendocrine tumour. Planning imaging with unenhanced T1-weighted VIBE A reveals a 12-mm metastasis in segment VIII (arrow). Coronal T1-weighted VIBE B depicts the small tip artefact of the microwave applicator in the subcutaneous tissue (small arrow) pointing towards the target lesion (larger arrow). Near-real-time imaging with MR fluoroscopy C shows the microwave antenna (dotted arrows) in three slice angulations and enables an accurate targeting of the tumour (arrows). Coronal therapy monitoring with T1-weighted VIBE D shows a hyperintense ablation zone covering the target tumour with an adequate safety margin
Sequence details
| Sequence | TE (ms) | TR (ms) | Slice thickness (mm) | Matrix (mm2) | Flip angle (°) | Bandwidth (Hz/pixel) |
|---|---|---|---|---|---|---|
| T2 HASTE (coronal) | 94 | 1100 | 3 | 256 × 120 | 160 | 490 |
| T2 TSE | 95 | 5800 | 6 | 384 × 384 | 158 | 590 |
| T1 Dixon VIBE | 2.4 | 6.7 | 3 | 320 × 182 | 10 | 520 |
| Diffusion* | 59 | 3000 | 6 | 192 × 144 | 90 | 1860 |
| CE dynamic T1 Dixon VIBEa,* | 2.4 | 6.7 | 3 | 320 × 182 | 10 | 520 |
| BEAT-Multislice | 3.2 | 464 | 8 | 128 × 128 | 20 | 500 |
| T1 VIBE | 1.3 | 3.5 | 2 | 320 × 189 | 10 | 400 |
| T1 Flash 2D | 4.1 | 193 | 6 | 256 × 140 | 70 | 140 |
| T2 TSE | 95 | 5800 | 6 | 384 × 384 | 158 | 590 |
| CE dynamic T1 Dixon VIBEb | 2.4 | 6.7 | 3 | 320 × 182 | 10 | 520 |
HASTE half acquisition single shot turbo spin echo, TSE turbo spin echo, VIBE volumetric interpolated breath-hold imaging, BEAT-Multislice multiplanar interactive real-time sequence, Flash fast low angle shot gradient echo, TrueFISP True fast imaging with steady state free precession, Diffusion echoplanar imaging with b-values of 0, 400, 800 mm/s2, CE contrast enhanced
aAfter i.v. injection of 0.025 mmol/kg body weight of gadoxetate disodium
bAfter i.v. injection of 0.1 mmol/kg body weight of gadobutrol
*Optional
Patient und tumour characteristics
| Patient no./age (years) | Tumour no. | Liver segment | Tumour entity | Maximal tumour diameter (mm) |
|---|---|---|---|---|
| 1/47 | 1.1a | V | Metastasis uveal melanoma | 17 |
| 1.2a | VIII | Metastasis uveal melanoma | 13 | |
| 2/66 | 2 | VIIIc | HCCb | 20 |
| 3/66 | 3 | VIII | HCCb | 15 |
| 4/64 | 4 | IV | Metastasis CRC | 21 |
| 5/57 | 5 | VIIIc | Metastasis CRC | 12 |
| 6/72 | 6.1 | VIII | HCCb | 15 |
| 6.2 | V | HCCb | 12 | |
| 7/48 | 7 | VIIc | HCCb | 25 |
| 8/54 | 8 | VIIIc | HCCb | 21 |
| 9/65 | 9 | VI | Metastasis uveal melanoma | 10 |
| 10/62 | 10 | VIII | Metastasis CRC | 8 |
| 11/61 | 11.1 | VIII | Metastasis NET | 12 |
| 11.2 | VIII | Metastasis NET | 6 | |
| 12/64 | 12.1 | VIIIc | Metastasis CRC | 21 |
| 12.2 | V | Metastasis CRC | 6 | |
| 13/61 | 13.1 | IV | Metastasis CRC | 7 |
| 13.2 | VII | Metastasis CRC | 8 | |
| 14/72 | 14 | VIIIc | HCC | 14 |
| 15/40 | 15 | VIII | Metastasis pancreas carcinoma | 24 |
| 16/77 | 16 | VIII | Metastasis uveal melanoma | 13 |
| 17/72 | 17.1 | IVc | HCCb | 17 |
| 17.2 | VIII | HCCb | 15 | |
| 18/79 | 18 | V | Metastasis CRC | 24 |
| 19/74 | 19 | VI | Metastasis CRC | 12 |
| 20/60 | 20.1 | IIc | Metastasis CRC | 16 |
| 20.2 | VII | Metastasis CRC | 14 | |
| 21/77 | 21 | V | Metastasis CRC | 10 |
HCC hepatocellular carcinoma, CRC colorectal carcinoma, NET neuroendocrine tumour
atumours treated in two procedures, bunderlying cirrhosis, csubcapsular tumour location