Literature DB >> 30506218

Liver microwave ablation: a systematic review of various FDA-approved systems.

Simeon J S Ruiter1, Wouter J Heerink2,3, Koert P de Jong4,3.   

Abstract

OBJECTIVES: The aim of the present study is to analyze preclinical and clinical data on the performance of the currently US Food and Drug Administration (FDA)-approved microwave ablation (MWA) systems.
METHODS: A review of the literature, published between January 1, 2005, and December 31, 2016, on seven FDA-approved MWA systems, was conducted. Ratio of ablation zone volume to applied energy R(AZ:E) and sphericity indices were calculated for ex vivo and in vivo experiments.
RESULTS: Thirty-four studies with ex vivo, in vivo, and clinical data were summarized. In total, 14 studies reporting data on ablation zone volume and applied energy were included for comparison R(AZ:E). A significant correlation between volume and energy was found for the ex vivo experiments (r = 0.85, p < 0.001) in contrast to the in vivo experiments (r = 0.54, p = 0.27).
CONCLUSION: Manufacturers' algorithms on microwave ablation zone sizes are based on preclinical animal experiments with normal liver parenchyma. Clinical data reporting on ablation zone volume in relation to applied energy and sphericity index during MWA are scarce and require more adequate reporting of MWA data. KEY POINTS: • Clinical data reporting on the ablation zone volume in relation to applied energy during microwave ablation are scarce. • Manufacturers' algorithms on microwave ablation zone sizes are based on preclinical animal experiments with normal liver parenchyma. • Preclinical data do not predict actual clinical ablation zone volumes in patients with liver tumors.

Entities:  

Keywords:  Ablation techniques; Electromagnetic radiation; Liver diseases; Microwaves; Tumor volume

Mesh:

Year:  2018        PMID: 30506218      PMCID: PMC6611060          DOI: 10.1007/s00330-018-5842-z

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


Introduction

Thermal ablation such as microwave ablation (MWA) is widely applied for the treatment of liver tumors. Thermal ablation alone or in combination with partial hepatectomy increases the number of intentionally curative treatments in patients in whom partial hepatectomy alone is not an option because of anatomical or functional reasons. Especially in patients with recurrent colorectal liver metastases (CRLM) after previous partial hepatectomy, thermal ablation increases the number of patients who could benefit from repeated procedures [1, 2]. However, the major problem of thermal ablation is incomplete ablation leading to ablation site recurrences (ASR), for which a clear definition should be used [3]. ASR is shown in 60% of patients with a lesion > 5 cm compared to 26% for 3–5 cm lesions and 16% for lesions < 3 cm [4]. Shape and volume of the ablation zone after MWA are depending on physical parameters as thermal conductivity, perfusion rate of the liver parenchyma. These parameters can be different in human liver tissue due to fibrosis, cirrhosis or steatosis [5, 6]. Planning for ablation is partially based on manufacturer-initiated working algorithms in combination with personal experience of the operator. These algorithms, which try to predict the three-dimensional diameter of the ablation zone in relation to the amount of applied energy, are often based on experiments which have serious shortcomings preventing a reliable translation to daily clinical practice. These shortcomings are the result of studies performed in (a) porcine or bovine liver (as opposed to human liver), (b) liver parenchyma (as opposed to tumors), (c) normal liver parenchyma (as opposed to cirrhotic, steatosis, or otherwise non-normal liver parenchyma in humans), and (d) non-perfused ex vivo livers (as opposed to perfused in vivo human livers with variable arterial and portal blood flow). These differences affect the way in which the applied energy is transferred into heat, resulting in highly unpredictable ablation zone volumes [6]. Additionally, despite several individual papers reporting on these shortcomings, a systematic review on this topic is lacking. The aim of this review is to analyze preclinical and clinical data on the performance of the US Food and Drug Administration (FDA)–approved MWA systems. [7]

Methods

Literature search and collected data

A systematic review of seven FDA-approved microwave ablation systems was performed in accordance with the PRISMA statement (Table 1) [13]. Literature published between 1 January 2005 and 1 January 2017, was searched  on Scopus including MEDLINE and EMBASE database, using the keywords “microwave ablation” AND “liver.” Retrieved studies were assessed for eligibility based on title and abstract—full papers were obtained and assessed in detail. Studies were included if (a) data on diameter or volume of the ablation zone—based on imaging or pathology—was reported, (b) ablation procedures were performed with FDA-approved MWA systems (Table 1), and (c) the amount of applied energy was reported and (d) were published in English. A data extraction form was used to collect relevant information including type of study (ex vivo, in vivo, or clinical), subject (porcine, bovine, sheep, or human), malignancy (none, primary, or secondary), device, parameters/outcomes, and measurement of ablation zone dimensions (on imaging or gross pathology). Data collection and analysis of unequivocal literature was performed by one researcher. Equivocal papers or data were discussed with co-authors until consensus was obtained. The included studies were categorized in preclinical (animal) and clinical (patient) studies. Preclinical studies were subdivided in two subcategories: ex vivo and in vivo studies (Table 1).
Table 1

Number of published preclinical and clinical studies using various MWA devices

AngioDynamics, Acculis MTAHS Medical, Amica™Ethicon, NeuWaveMedWaves, AveCure™Medtronic, Evident™Medtronic, Emprint™Perseon, MicroThermX™Total
ABCDEFG
US FDA clearanceJanuary 2006September 2009October 2010January 2008December 2008April 2014August 2010
frequency (MHz)2450245024509159152450915
Ex vivo (perfused)3 (0)8 (0)5 (1)3 (1)3 (1)0 (0)1 (1)23 (4)
In vivo433010011
Clinical studies04110309
Total7159443143a

aThe number of total studies is higher than the number of included published papers. Three papers published ex vivo and in vivo data [8–10]. One paper published ex vivo, in vivo, and clinical data [11]. One paper published ex vivo data of four MWA devices [12]. These are counted as individual studies

Number of published preclinical and clinical studies using various MWA devices aThe number of total studies is higher than the number of included published papers. Three papers published ex vivo and in vivo data [8-10]. One paper published ex vivo, in vivo, and clinical data [11]. One paper published ex vivo data of four MWA devices [12]. These are counted as individual studies

Ablation zone volume and applied energy

Ablation zone volume as reported in the selected papers was recorded. If only long-axis diameter (LAD) and short-axis diameters (SAD) were recorded, the ablation zone volume was estimated, assuming ellipsoid morphology by . Papers with only one diameter of the ablation zone were excluded for quantitative comparison. Sphericity index of ablation zones with reported SAD and LAD was calculated by SI = SAD2/LAD2. The cumulative applied energy was determined by multiplying the power level (Watt), as set on the MWA generator, and the ablation time (seconds). The relation between applied energy and ablation zone volume is the only way to quantitatively compare the various published reports. To this end, the ratio of ablation zone volume to applied energy R(AZ:E) of each ablation experiment was calculated by dividing the ablation zone volume (mL) by the applied energy (kJ). The correlation coefficient was determined using IBM SPSS Statistics version 23 (IBM Corporation). The mean volume and applied energy of the subgroups were represented in a bubble chart. The sizes of the bubbles are determined by the sample size of the subgroup.

Results

Figure 1 shows a flow diagram of study identification and the exclusion process, resulting in 34 eligible studies [8–12, 14–42] (Tables 2, 3, and 4). Cross-referencing of the identified studies did not reveal any additional papers. In three studies, both ex vivo and in vivo data were published [8-10], and these were counted as six separate studies in Table 1. In one study, ex vivo, in vivo, and clinical data were presented [11], and these were counted as three separate studies. Also, in one paper, the ex vivo data of four MWA devices were presented, and these were counted as four separate studies [12]. Assessment of publication bias (overreporting of significant positive results) is not applicable for this review, because the data is not presented as negative or positive. We performed a partial correction for the heterogeneity (variability in the study characteristics) in the analyzed studies by reporting stratified results (for instance ex vivo vs. in vivo and animal vs. human).
Fig. 1

Flow chart of article selection process

Table 2

Included ex vivo studies for qualitative analysis

Author + yearSubjectsMWA systemAblation zone measured byAblation protocol (energy and time)Ablation diameter (cm)Ablation volume (mL)Sphericity index
Hines-Peralta et al 2006 [9]Bovine 120 ablationsAGP50–150 W 4–20 min3.0 × 3.5–7.6 × 12.316.5–372.0a0.33–0.75
Lopresto et al 2012 [35]3 bovine liverBGP30 W 10 min3.7 ± 0.4 × 3.2 ± 0.419.8a0.75
Sommer et al 2012 [39]8 porcine 15 ablationsBGP20–105 W 5 min2.3 × 4.0–3.5 × 6.711.1–42.30.27–0.33
Lubner et al 2012 [10]Bovine 18 ablationsCGP135 W 4–16 min3.5 ± 0.2–4.8 ± 0.2
Hoffmann et al 2013 [12]13 bovine 108 ablationsABDEGPManufacturer recommendations

A: 4.34

B: 4.55

D: 4.09

E: 2.68

A: 57.5

B: 72.3

D: 56.0

E: 17.1

A: 0.75

B: 0.68

D: 0.58

E: 0.64

Collettini et al 2013 [41]50 cuboid bovineEMR45 W 7 min

7.3 ± 2.1 (seq1)

4.7 ± 1.6 (seq2)

Dodd et al 2013 [37]

15 blood-perfused bovine livers

60 ablations

GGP60 W 10 min4.73 ± 0.21 × 2.93 ± 0.10–5.22 ± 0.17 × 2.82 ± 0.1221.30 ± 0.95–22.6 ± 1.530.29–0.38
Liu et al 2014 [17]6 bovineCGP100 W 10 min6.45 ± 0.36 × 3.88 ± 0.250.8a0.36
Niemeyer et al 2015 [20]BovineAGP60–180 W for 2, 4, and 6 minPlots with diametersPlots with volumes
Cavagnaro et al 2015 [21]Bovine 32 ablationsBGP

5 W 10 min;

10 W 10 min;

15 W 10 min;

20 W 10 min;

40 W 10 min

1.40 ± 0.09 × 1.18 ± 0.08

2.18 ± 0.13 × 1.84 ± 0.18

2.51 ± 0.14 × 2.20 ± 0.10

3.17 ± 0.29 × 2.73 ± 0.46

4.33 ± 0.18 × 3.63 ± 0.08

1.02a

3.86a

6.36a

12.37a

25.60a

0.71

0.71

0.77

0.74

0.70

Cavagnaro et al 2015 [22]BovineBGP60 W 10 min5.34 ± 1.7 × 4.29 ± 2.051.46a0.65
Paul et al 2015 [23]3 porcineBCT100 W 4.5 min4.1 ± 0.2 × 5.6 ± 0.249.29a0.54
Kim et al 2015 [24]BovineDGP

Single 24 W 675 s;

multi 14-24 W 401 s;

single 28 W 339 s;

multi 16-28 W 306 s

1.72 × 1.84;

1.56 × 1.69

2.21 × 2.38

1.92 × 2.14

4.63 ± 0.5;

3.75 ± 0.8

15.33 ± 3.4

10.98 ± 2.5

0.87

0.85

0.86

0.80

Pillai et al 2015 [25]Perfused bovineDGP25-28 W; temperature 110°; 17 min

4.4 ± 0.31 × 5.8 ± 0.4 (− heat sink)

3.6 ± 0.4 × 4.8 ± 0.3 (+ heatsink)

232 ± 28 (− heat sink)

181 ± 21 (+ heatsink)

0.58

0.56

Dodd et al 2015 [37]Blood-perfused bovineCGP140 W 5 and 10 min5.61 ± 0.20 × 3.2 ± 0.08; 6.51 ± 0.20 × 3.81 ± 0.07

30.22 ± 1.85;

49.30 ± 1.85

0.33

0.4

Ringe et al 2015 [30]

Perfused porcine

108 ablations

EGP45 W 10 min2.1 × 1.52.47a0.51
Weiss et al 2015 [33]16 bovine liversBGP

40 W 1 min

2 min

3 min

6 min

10 min

0.76

1.08

1.30

1.79

2.14

Harari et al 2016 [8]BovineCGPSimultaneous 2 and 3 antennas 50 W 5 min

3.56 ± 0.39 × 4.51 ± 0.63;

3.97 ± 0.29 × 4.97 ± 0.32;

32.1 ± 5.5;

45.8 ± 8.8;

0.62

0.64

Amabile et al 2016 [11]

20 bovine livers

108 ablations

BGP

3–30 min

20–130 W

1.6 ± 0.1 × 2.4 ± 0.3–7.2 ± 0.4 × 10.1 ± 0.73.22–274.15a0.66–0.70

A Acculis, B Amica, C NeuWave 140, D AveCure, E Evident, F Emprint, G MicroThermX, CT computed tomography, MRI magnetic resonance imaging, GP gross pathology

aEstimated volume

Table 3

Included in vivo studies for qualitative analysis

Author + yearSubjectsMWA systemAblation zone measured byAblation protocol (energy and time)Ablation diameter (cm)Ablation volume (mL)Sphericity index
Hines-Peralta et al 2006 [9]

porcine

14 pigs, 45 ablations

AGP

50–150 W

4–20 min

2.8 × 4.1–5.8 × 5.516.8–110.6a0.35–0.90
Awad et al 2007 [14]

3 porcine

9 ablations

AGP

100 W

2–8 min

3.7 × 4.5–5.3 × 6.433.5 ± 17.3–92.0 ± 6.50.60–0.68
Garrean et al 2009 [15]4 porcine 16 ablationsAGP

70–100 W

4 min

3.0–6.54
Meloni et al 2011 [26]4 porcine 16 ablationsBGP

40–60 W

15 min

3.3 ± 0.6 × 2.9 ± 0.5–4.2 ± 1.1 × 3.1 ± 1.114.5–21.1a0.54–0.77
Lubner et al 2012 [10]Porcine 48 ablationsCGP140 W 2–10 min2.0 ± 0.2 × 3.2 ± 1.2–3.4 ± 0.6 × 4.1 ± 0.90.39–0.69
Correa-Gallego et al 2014 [18]Porcine 6 ablationsEGP45 W 10 min7.05 cm2
Gockner et al 2015 [19]

3 sheep

9 ablations

ACT80 W 2 min4.15 ± 0.46 × 2.37 ± 0.3716.5 ± 5.10.33
Bedoya et al 2014 [27]6 porcineCGP5 delivery methods of 30 kJ of energy

2.3 ± 0.7 × 1.4 ± 0.5;

3.5 ± 0.6 × 2.1 ± 0.4;

3.8 ± 0.9 × 2.4 ± 0.7;

4.6 ± 0.6 × 3.0 ± 0.4;

5.2 ± 0.8 × 3.3 ± 0.9

23.6 ± 26.5;

67.6 ± 34.5;

105.4 ± 78.3;

176.7 ± 45.9;

265.7 ± 208.1

0.37

0.36

0.40

0.43

0.40

Moreland et al 2015 [28]

5 porcine

28 ablations

CGP

65 W

5 min

3.3 ± 0.9
Harari et al 2016 [8]PorcineCGP2 and 3 antennas 65 W 5 min

3.33 ± 0.80 × 3.72 ± 0.93;

4.02 ± 0.51 × 4.68 ± 0.44

21.3 ± 13;

47.8 ± 13

0.80

0.74

Amabile et al 2016 [11]

12 porcine

28 ablations

BGP

5–10 min

60, 80, 100 W

2.5 ± 0.2 × 4.3 ± 0.3–4.9 ± 0.1 × 8.5 ± 0.414.07–106.86a0.56–0.66
Wu et al 2016 [34]4 porcine 15 ablationsCGP5 min 100 W

No contrast 23.9 ± 1.2

contrast 22.3 ± 1.8

non-perfused 39.3 ± 1.7

A Acculis, B Amica, C NeuWave 140, E Evident, CT computed tomography, GP gross pathology

aEstimated volume

Table 4

Included clinical studies for qualitative analysis

Author + yearSubjectsMWA systemAblation zone measured byAblation protocol (energy and time)Ablation diameter (cm)Ablation volume (mL)Sphericity index
Ratanaprasatporn et al 2013 [40]10 ablations and resections (3 HCC; 7 metastatic)DGP10–32 W; 110°–120° 10 min4.18.7
Di Vece et al 2014 [42]20 patients with primary (9) and secondary (11) liver tumorsBUS60–70 W 10 min4.85 ± 0.67 × 3.85 ± 0.460.63
Winokur et al 2014 [16]36 ablations with Amica (20) and NeuWave (16)BCCT

B: 43.8 ± 27.4 kJ

C: 21.4 ± 12.6 kJ

B: 5.1 ± 1.5 × 3.0 ± 0.9

C: 3.9 ± 0.7 × 2.7 ± 0.5

B: 33.0 ± 18.9

C: 15.5 ± 6.7

B: 0.49

C: 0.39

Berber et al 2015 [29]5 patients; 9 malignant tumorsFCTPatient specificDetails per ablation
Berber et al 2016 [31]18 patients; 54 malignant liver tumorsFCTPatient specificAblation zone and time of 100 W0.9
Zaidi et al 2016 [32]53 laparoscopic ablationsFCTPatient specificAblation time/size plot for 100 W0.9
Amabile et al 2016 [11]

46 patients; 32 HCC;

19 metastasis

BCT

5 min 60 W

10 min 60 W

HCC 3.3 ± 0.5 × 4.8 ± 0.7

metastasis 4.0 ± 0.9 × 5.5 ± 1.7

HCC 3.7 ± 0.3 × 5.2 ± 0.6

metastasis 4.1 ± 0.6 × 6.5 ± 0.9

HCC 0.64–0.74

metastasis 0.70–0.71

Shyn et al 2016 [36]52 patients 93 ablationsBMRI/CTPatient specificCorrelation with energy

B Amica, C NeuWave 140, D AveCure, Evident, F Emprint, CT computed tomography, MRI magnetic resonance imaging, US ultrasound, GP gross pathology, HCC hepatocellular carcinoma

Flow chart of article selection process Included ex vivo studies for qualitative analysis A: 4.34 B: 4.55 D: 4.09 E: 2.68 A: 57.5 B: 72.3 D: 56.0 E: 17.1 A: 0.75 B: 0.68 D: 0.58 E: 0.64 7.3 ± 2.1 (seq1) 4.7 ± 1.6 (seq2) 15 blood-perfused bovine livers 60 ablations 5 W 10 min; 10 W 10 min; 15 W 10 min; 20 W 10 min; 40 W 10 min 1.40 ± 0.09 × 1.18 ± 0.08 2.18 ± 0.13 × 1.84 ± 0.18 2.51 ± 0.14 × 2.20 ± 0.10 3.17 ± 0.29 × 2.73 ± 0.46 4.33 ± 0.18 × 3.63 ± 0.08 1.02a 3.86a 6.36a 12.37a 25.60a 0.71 0.71 0.77 0.74 0.70 Single 24 W 675 s; multi 14-24 W 401 s; single 28 W 339 s; multi 16-28 W 306 s 1.72 × 1.84; 1.56 × 1.69 2.21 × 2.38 1.92 × 2.14 4.63 ± 0.5; 3.75 ± 0.8 15.33 ± 3.4 10.98 ± 2.5 0.87 0.85 0.86 0.80 4.4 ± 0.31 × 5.8 ± 0.4 (− heat sink) 3.6 ± 0.4 × 4.8 ± 0.3 (+ heatsink) 232 ± 28 (− heat sink) 181 ± 21 (+ heatsink) 0.58 0.56 30.22 ± 1.85; 49.30 ± 1.85 0.33 0.4 Perfused porcine 108 ablations 40 W 1 min 2 min 3 min 6 min 10 min 0.76 1.08 1.30 1.79 2.14 3.56 ± 0.39 × 4.51 ± 0.63; 3.97 ± 0.29 × 4.97 ± 0.32; 32.1 ± 5.5; 45.8 ± 8.8; 0.62 0.64 20 bovine livers 108 ablations 3–30 min 20–130 W A Acculis, B Amica, C NeuWave 140, D AveCure, E Evident, F Emprint, G MicroThermX, CT computed tomography, MRI magnetic resonance imaging, GP gross pathology aEstimated volume Included in vivo studies for qualitative analysis porcine 14 pigs, 45 ablations 50–150 W 4–20 min 3 porcine 9 ablations 100 W 2–8 min 70–100 W 4 min 40–60 W 15 min 3 sheep 9 ablations 2.3 ± 0.7 × 1.4 ± 0.5; 3.5 ± 0.6 × 2.1 ± 0.4; 3.8 ± 0.9 × 2.4 ± 0.7; 4.6 ± 0.6 × 3.0 ± 0.4; 5.2 ± 0.8 × 3.3 ± 0.9 23.6 ± 26.5; 67.6 ± 34.5; 105.4 ± 78.3; 176.7 ± 45.9; 265.7 ± 208.1 0.37 0.36 0.40 0.43 0.40 5 porcine 28 ablations 65 W 5 min 3.33 ± 0.80 × 3.72 ± 0.93; 4.02 ± 0.51 × 4.68 ± 0.44 21.3 ± 13; 47.8 ± 13 0.80 0.74 12 porcine 28 ablations 5–10 min 60, 80, 100 W No contrast 23.9 ± 1.2 contrast 22.3 ± 1.8 non-perfused 39.3 ± 1.7 A Acculis, B Amica, C NeuWave 140, E Evident, CT computed tomography, GP gross pathology aEstimated volume Included clinical studies for qualitative analysis B: 43.8 ± 27.4 kJ C: 21.4 ± 12.6 kJ B: 5.1 ± 1.5 × 3.0 ± 0.9 C: 3.9 ± 0.7 × 2.7 ± 0.5 B: 33.0 ± 18.9 C: 15.5 ± 6.7 B: 0.49 C: 0.39 46 patients; 32 HCC; 19 metastasis 5 min 60 W 10 min 60 W HCC 3.3 ± 0.5 × 4.8 ± 0.7 metastasis 4.0 ± 0.9 × 5.5 ± 1.7 HCC 3.7 ± 0.3 × 5.2 ± 0.6 metastasis 4.1 ± 0.6 × 6.5 ± 0.9 HCC 0.64–0.74 metastasis 0.70–0.71 B Amica, C NeuWave 140, D AveCure, Evident, F Emprint, CT computed tomography, MRI magnetic resonance imaging, US ultrasound, GP gross pathology, HCC hepatocellular carcinoma

Ex vivo animal studies

In total, 18 studies published ex vivo animal results, three in porcine liver [23, 30, 39] and 15 in bovine liver (Table 1) [8–12, 17, 20, 22, 24, 25, 33, 35, 37, 38, 41]. Four studies were performed in perfused liver [25, 30, 37, 38]. Dodd et al used 15 blood-perfused (37 °C) bovine livers for 60 MW ablations with system G (MicroThermX) [37]. Ablation zone volumes, measured by gross pathology, were unaffected by changes in portal venous blood flow (range of 60–100 mL/min per 100 g tissue). These authors also repeated the blood-perfused study with 60 ablations in ten livers with system C (NeuWave) [38]. Again, a change in blood flow rate did not affect the size and shape of the ablation zone, as evaluated by pathology. Pillai et al perfused ex vivo bovine livers with 37 °C Ringer solution and compared volumes and diameters of ablation zones in relation to the distance to the major hepatic vein in three ablation experiments [25]. For system D (AveCure), ablation zones within 8 mm to the major hepatic veins were 22% smaller than ablation zones more than 30 mm away from major hepatic veins [25]. Ringe et al used perfused glass tubes in porcine liver to simulate the hepatic veins [30]. They analyzed 108 ablation zones generated by system E (Evident) and found that ablation zones were influenced by flow rate (0, 700, and 1400 mL/min) at a maximum distance of 10 mm to the glass tube. Hoffmann et al compared the four systems A, B, D, and E (Acculis, Amica, AveCure, and Evident) in bovine liver [12]. They found that system C (NeuWave, 3 antennas) created the largest and most spherical zones. Most systems note in their manual that ablation algorithms are based on “internal ex-vivo experiments.” However, no peer-reviewed publications were found for ex vivo testing for system F (Emprint).

In vivo animal studies

In 12 animal studies, the effects of microwave ablation in in vivo liver parenchyma were analyzed [8–11, 14, 15, 18, 19, 26–28, 34]. Gockner et al compared MWA using system A (Acculis) with and without transarterial embolization before ablation in sheep [19]. Extent and shape of the ablation zones were determined by CT. Ablation zone diameters increased by 22.8% by using transarterial embolization before ablation (p < 0.01). Hines-Peralta et al compared ex vivo bovine and in vivo porcine MWA, using system A (Acculis) [9]. Unexpectedly, 8-mm larger (57 mm vs. 49 mm) diameter of ablation zones (p < 0.01) were obtained in in vivo (57 ± 2 mm) experiments, compared to ex vivo (49 ± 2 mm). However, for ablation times longer than 8 min, ex vivo diameters still increased while in vivo diameters remained constant. Also, Lubner et al compared ex vivo bovine and in vivo porcine MWA [10]. Ablation zones were similar during the first 2 min, but in vivo ablations did not continue to grow as much as ex vivo [10]. No in vivo studies were performed with system D (AveCure) and system F (Emprint).

Clinical studies

Nine studies reported clinical results of the ablation zone [11, 16, 18, 29, 31, 32, 36, 40, 42]. Ratanaprasatporn et al performed a prospective study in ten patients with liver tumors, treated with system D (AveCure) [40]. After liver resection, the ablation zone volumes were measured on gross pathology. Six of the ten treatments showed ablation with complete necrosis on pathological examination [40]. Di Vece et al reported a mean long-axis diameter of 4.85 cm in 20 patients treated with system B (Amica) [42]. Winokur et al analyzed the ablation properties of system B (Amica, 25 ablations in 20 patients) and system C (NeuWave, 11 ablations in 8 patients) in order to analyze if the manufacturer published reference values are useful [16]. The study indicated that in vivo (clinical) ablation zone volumes are significantly smaller than stated by reference values from the manufacturers (0.69 cm3 vs. 1.29 cm3; p = 0.003) [16]. Berber et al compared predicted ablation diameters (system F, Emprint) with the ablation zone at the 2-week post-ablation CT scan of nine patients [29]. The maximum diameter of the ablation zone was 1.12 ± 0.11 times larger than predicted. No residual tumors were seen at the 2-week scan. In another study, Berber reported a scatterplot showing the CT diameter of ablation zones obtained after 100 W of power during laparoscopic MW ablation with system F (Emprint) in 15 patients [31]. In a study of 149 laparoscopic ablations, Zaidi et al reported system F to be satisfactory in achieving the predicted ablation sizes [32]. Amabile et al compared ablation zone dimensions in liver tumors in patients with both in vivo (porcine) and ex vivo (bovine) experiments (only liver parenchyma). They concluded that the ex vivo animal data reliably predicted the dimensions in (in vivo) human liver tumors. Shyn et al compared ablation zone diameters and volumes with applied and net energy (after correcting for reflectivity), and with manufacturer chart predictions [36]. Applied energy (r = 0.52) and net energy (r = 0.53) did not correlate better than manufacturer chart prediction (r = 0.60). Also, no differences were seen between cirrhotic and non-cirrhotic livers. None of the clinical studies were performed with system A (Acculis), E (Evident), and G (MicroThermX).

Comparison of ablation volume–applied energy ratio

In total, 14 animal studies reported data on ablation zone volume and applied energy and could be used to compare R(AZ:E) [8, 9, 11, 12, 14, 17, 19, 22–24, 26, 30, 35, 39]. These studies were categorized in 22 subgroups based on device and tissue (ex vivo and in vivo). A significant correlation between volume and energy was found for the ex vivo experiments (r = 0.85, p < 0.001) in contrast to the in vivo experiments (r = 0.54, p = 0.27) (Fig. 2). Results of SI calculation are shown in Tables 2, 3, and 4.
Fig. 2

Bubble chart of the ratio of ablation zone volume (mL) to applied energy (kJ) R(AZ:E) for 22 subgroups in animal studies of all devices with adequate volume/energy representation. The sizes of the bubbles are determined by the sample size of the subgroup

Bubble chart of the ratio of ablation zone volume (mL) to applied energy (kJ) R(AZ:E) for 22 subgroups in animal studies of all devices with adequate volume/energy representation. The sizes of the bubbles are determined by the sample size of the subgroup

Discussion

In this review on studies performed with seven FDA-approved MWA devices, we found that only a minority (9 out of 43 studies) was based on clinical studies. On top of that, a major limitation in these clinical studies is the lack of data on the ratio ablation zone volume: applied energy R(AZ:E). In only 14 preclinical studies, the R(AZ:E) was calculated. To our knowledge, this is the first study which compares the ratio ablation zone volume and applied energy of FDA-approved MWA devices between 2005 and 2017. MWA systems are categorized as class II medical devices by the FDA. This classification requires that a new device must be proven to be substantially equivalent to a device that was legally marketed (predicate device) prior to May 28, 1976. Since microwave technology has been around for decades, ex vivo bench testing is sufficient for the regulatory approvals and no additional submission of clinical data is required [7]. Also, reporting results on ablation zone volume and applied energy in clinical setting are limited. This might be the reason for the lack of clinical data of MWA systems. An analysis of the studies in our review suggests a number of possible explanations for the discrepancy between manufacturers’ provided ablation algorithms and the actual clinical ablation zone sizes. This discrepancy was demonstrated by Winokur et al who indicated that in vivo (clinical) ablation zone volumes are significantly smaller than stated by reference values from the manufacturers of the Amica and NeuWave system [16]. Depending on the type of study (ex vivo vs. in vivo), ablation time was found to be a determinant for ablation zone diameter. For ablation time longer than 8 min, ex vivo diameter still increased while in vivo diameters remained constant, suggesting plateau formation which probably is caused by the antagonizing effect of perfusion [9, 16]. Obviously, preclinical studies using animal livers are only performed in normal liver parenchyma with absence of tumor tissue (tumor characteristics) and underlying liver disease (liver characteristics). Deshazer et al simulated these properties in a two-compartmental computer model and showed that ablation zone volume could increase with 36% in patients with cirrhotic liver as compared to healthy liver tissue [6]. Tumor tissue revealed a 20% higher thermal conductivity (the property of tissue to conduct heat) than healthy liver tissue. Hyperperfused and hypoperfused tumors within normal liver parenchyma showed minimal variation in ablation zone volume. Furthermore, steatotic parenchyma had 50% lower thermal conductivity than healthy liver tissue [6]. Hepatic steatosis is of importance in patients with CRLM treated with neo-adjuvant chemotherapy and should be taken into account when planning MWA treatment for these patients [43]. Perfusion in ex vivo experiments also determines to a great extent the effect of heat distribution by convection in MWA; convection of heat is mainly influenced by vascularization of the tumor and adjacent large blood vessels which can cause heat sink. In most ex vivo studies, livers were not perfused and thus not subject to heat sink effects. Perfusion of cirrhotic liver tissue is reported to be 36% lower than healthy liver tissue [44]. This is of importance because approximately 90% of the patients with HCC suffer from cirrhosis [45]. Additionally, the majority of HCCs have a predominant arterial perfusion, which makes prediction of the obtained ablation zone in relation to perfusion phenomena even more imprecise. Compared to perfusion in liver parenchyma, HCC tumor tissue has a significantly higher arterial perfusion and lower portal venous hepatic blood flow [46]. Therefore, the results of Amabile et al are difficult to interpret, because they found a better correlation between ablation zone dimensions in liver tumors (HCC) in patients and non-perfused (ex vivo) bovine liver than in perfused (in vivo) porcine liver. A possible explanation for this seemingly contradictory finding is that less than 20% of the ablation zone volume encompasses tumor and more than 80% liver parenchyma [47]. This suggests that liver parenchyma might be of more importance for the ablation zone volume than the tumor tissue. Clinical studies were limited in number and additionally in general no distinction in tumor type (primary or secondary) is made. None of the studies in this review reported the underlying liver diseases, like hepatic steatosis, fibrosis, or cirrhosis. In a retrospective study, MWA volumes of HCC in cirrhotic liver and CRLM in healthy liver were compared. R(AZ:E) for HCC with system A (Acculis) is twofold higher for HCC (R(AZ:E) = 0.61) compared to CRLM (R(AZ:E) = 0.35). Also, ablation treatment for HCC with system B (Amica) resulted in a 50% increase of ablation zone volume, compared to CRLM [47]. For treatment of HCC up to 4 cm, no significant differences were found in local tumor progression between radiofrequency ablation (RFA) and MWA [48]. During liver ablation, the power of heating (Watt) is an important factor. When using low power, relatively more heat will be dispersed into the surrounding tissue, and temperature of the ablation zone will be low. In contrast, higher power for short time will lead to high temperatures around the antenna and contraction of the target tissue. Bedoya et al compared different ways to deliver 30 kJ of energy (25 W 20 min, 50 W 10 min, and 100 W 5 min) in in vivo porcine livers [27]. Significantly larger ablation zone volumes were achieved with high power ablations (23.6 ± 26.5 mL; 105.4 ± 78.3 mL; 265.7 ± 208.1, respectively; p < 0.03). Interestingly, they also investigated the effect of pulsed energy delivery (25 kJ) to limit the effects of heat sink and to provide larger ablation zone volumes than continuous energy delivery (67.4 ± 34.5 cm3 vs. 23.6 ± 26.5 cm3, p = 0.43). Another important factor is the reflection of energy by the antenna cable. The majority of the microwave systems report the output power of the generator as the applied energy at the antenna. However, 15–30% of the output energy per meter length will be lost by the antenna cable [49]. So, for comparison of ablation devices by R(AZ:E), the most relevant parameter is the amount of energy deposited into the liver tissue, which is known as the net energy [5, 49, 50]. Only systems B (Amica) and D (AveCure) display the reflection of energy on the generator. A comparable disagreement is found for the various antenna designs: if an antenna is built to have stronger fields in certain areas (Watts/area), a different amount of tissue reaching 60 °C might be expected. So, differences in R(AZ:E) between devices might be due to antenna design and cable length [49]. Despite the limitation of using R(AZ:E), it is the only parameter to compare the currently available data quantitatively. Most liver tumors treated by MWA are spherical which requires also a spherical ablation zone to achieve a sufficient ablation margin. However, most ablation systems create ellipsoidal ablation zones with poor sphericity values. Reflection of energy by the antenna shaft results in heating of the shaft which contributes to bigger LAD. This may also increase the risk of thermal damage of adjacent liver tissue. Data about the sphericity of ablation zones in patients are scarce. In this study, we calculated the SI for all studies in which the LAD and SAD were reported. However, data are heterogeneous and depending on power and time of the ablation. A recently published study compared system F (Emprint) with systems B (Amica) and E (Evident) [51]. Significantly more spherical ablation zones in patients were achieved with system F than with systems B and E. Complete ablation was possible with a single antenna placement regardless of the angle approach because of the almost spherical ablation zone. There might be several strategies to decrease ASR after microwave ablation. First of all, more in vivo and clinical studies with MWA should be conducted with cirrhotic and other diseased parenchyma, like steatotic liver, to confirm computer model studies. Secondly, presentation of data in studies should be reported adequately by means of applied energy (without reflection), LAD, SAD, SI, and volume of each ablation zone (Table 5). Additionally, for clinical studies, the status of the underlying liver parenchyma should be reported.
Table 5

Guidelines for reporting on future studies describing ablation experiments

VariableValue
Type of experimentsEx vivo (perfused, non-perfused), in vivo, clinical
Subjectsanimal (porcine, bovine), human
Type of liver parenchymanormal, cirrhosis, fibrosis, steatotic
Device
 Applied energy (kJ)Energy = ablation time (seconds) × power (Watt) / 1000
 Ablation diameters (cm)Long-axis diameter (LAD) and short-axis diameter (SAD)
 Ablation zone volume (mL) \documentclass[12pt]{minimal} \usepackage{amsmath} \usepackage{wasysym} \usepackage{amsfonts} \usepackage{amssymb} \usepackage{amsbsy} \usepackage{mathrsfs} \usepackage{upgreek} \setlength{\oddsidemargin}{-69pt} \begin{document}$$ V=\frac{4}{3}\pi \left( LAD/2\right){\left( SAD/2\right)}^2 $$\end{document}V=43πLAD/2SAD/22
 Sphericity indexSI = SAD2/LAD2

LAD long-axis diameter, SAD short-axis diameter

Guidelines for reporting on future studies describing ablation experiments LAD long-axis diameter, SAD short-axis diameter There are some limitations to this study. Ablation outcomes were compared by R(AZ:E). Ablation zone volume as reported in the selected papers was recorded. If only one or two diameters were reported, the ablation zone volume was estimated assuming ellipsoid or spherical morphology. Also, the diameters or volumes of the ablation zones were assessed by different measurement modalities (gross pathology, CT, MRI, or ultrasound) which induce bias. Finally, a systematic review formally requires a control group, but it is clear that this is not available for the current review. In conclusion, manufacturers’ algorithms on microwave ablation zone sizes are based on preclinical animal experiments with normal liver parenchyma. Clinical data reporting on ablation zone volume in relation to applied energy and sphericity index during MWA are scarce which requires more adequate reporting on MWA data.
  51 in total

1.  Hepatic perfusion parameters in chronic liver disease: dynamic CT measurements correlated with disease severity.

Authors:  B E Van Beers; I Leconte; R Materne; A M Smith; J Jamart; Y Horsmans
Journal:  AJR Am J Roentgenol       Date:  2001-03       Impact factor: 3.959

Review 2.  Local recurrence after hepatic radiofrequency coagulation: multivariate meta-analysis and review of contributing factors.

Authors:  Stefaan Mulier; Yicheng Ni; Jacques Jamart; Theo Ruers; Guy Marchal; Luc Michel
Journal:  Ann Surg       Date:  2005-08       Impact factor: 12.969

3.  Microwave ablation in a hepatic porcine model: correlation of CT and histopathologic findings.

Authors:  Michael M Awad; Lara Devgan; Ihab R Kamel; Michael Torbensen; Michael A Choti
Journal:  HPB (Oxford)       Date:  2007       Impact factor: 3.647

4.  Ultrasound monitoring of a novel microwave ablation (MWA) device in porcine liver: lessons learned and phenomena observed on ablative effects near major intrahepatic vessels.

Authors:  S Garrean; J Hering; A Saied; P J Hoopes; W S Helton; T P Ryan; N J Espat
Journal:  J Gastrointest Surg       Date:  2008-10-21       Impact factor: 3.452

5.  Preferred reporting items for systematic reviews and meta-analyses: the PRISMA statement.

Authors:  David Moher; Alessandro Liberati; Jennifer Tetzlaff; Douglas G Altman
Journal:  Ann Intern Med       Date:  2009-07-20       Impact factor: 25.391

6.  Image-guided tumor ablation: standardization of terminology and reporting criteria.

Authors:  S Nahum Goldberg; Clement J Grassi; John F Cardella; J William Charboneau; Gerald D Dodd; Damian E Dupuy; Debra A Gervais; Alice R Gillams; Robert A Kane; Fred T Lee; Tito Livraghi; John McGahan; David A Phillips; Hyunchul Rhim; Stuart G Silverman; Luigi Solbiati; Thomas J Vogl; Bradford J Wood; Suresh Vedantham; David Sacks
Journal:  J Vasc Interv Radiol       Date:  2009-07       Impact factor: 3.464

7.  Acute portal venous injury after microwave ablation in an in vivo porcine model: a rare possible complication.

Authors:  Maria Franca Meloni; Anita Andreano; Giorgio Bovo; Bartolomeo Chiarpotto; Claudio Amabile; Sandro Gelsomino; Sergio Lazzaroni; Sandro Sironi
Journal:  J Vasc Interv Radiol       Date:  2011-05-08       Impact factor: 3.464

Review 8.  Microwave ablation technology: what every user should know.

Authors:  Christopher L Brace
Journal:  Curr Probl Diagn Radiol       Date:  2009 Mar-Apr

9.  Microwave ablation: results with a 2.45-GHz applicator in ex vivo bovine and in vivo porcine liver.

Authors:  Andrew U Hines-Peralta; Nadeer Pirani; Peter Clegg; Nigel Cronin; Thomas P Ryan; Zhenjun Liu; S Nahum Goldberg
Journal:  Radiology       Date:  2006-02-16       Impact factor: 11.105

10.  Incidence and predictors of hepatocellular carcinoma in patients with cirrhosis.

Authors:  George N Ioannou; Meaghan F Splan; Noel S Weiss; George B McDonald; Laura Beretta; Sum P Lee
Journal:  Clin Gastroenterol Hepatol       Date:  2007-05-16       Impact factor: 11.382

View more
  15 in total

1.  Interstitial Optical Monitoring of Focal Laser Ablation.

Authors:  Rory Geoghegan; Le Zhang; Alan Priester; Holden H Wu; Leonard Marks; Shyam Natarajan
Journal:  IEEE Trans Biomed Eng       Date:  2022-07-18       Impact factor: 4.756

2.  Microwave Thermoablation of Colorectal Liver Metastases Close to Large Hepatic Vessels Under Pringle Maneuver Minimizes the "Heat Sink Effect".

Authors:  R Rhaiem; R Kianmanesh; M Minon; A Tashkandi; A Aghaei; G Ledoux; Ch Hoeffel; O Bouche; D Sommacale; T Piardi
Journal:  World J Surg       Date:  2020-05       Impact factor: 3.352

3.  Performance of the Emprint and Amica Microwave Ablation Systems in ex vivo Porcine Livers: Sphericity and Reproducibility Versus Size.

Authors:  P Hendriks; W E M Berkhout; C I Kaanen; J H Sluijter; I J Visser; J J van den Dobbelsteen; L F de Geus-Oei; A G Webb; M C Burgmans
Journal:  Cardiovasc Intervent Radiol       Date:  2021-01-18       Impact factor: 2.740

4.  The combination of electroporation and electrolysis (E2) employing different electrode arrays for ablation of large tissue volumes.

Authors:  Nina Klein; Enric Guenther; Florin Botea; Mihail Pautov; Simona Dima; Dana Tomescu; Mihai Popescu; Antoni Ivorra; Michael Stehling; Irinel Popescu
Journal:  PLoS One       Date:  2019-08-22       Impact factor: 3.240

5.  Direction of Tissue Contraction after Microwave Ablation: A Comparative Experimental Study in Ex Vivo Bovine Liver.

Authors:  Junhyok Lee; Hyunchul Rhim; Min Woo Lee; Tae Wook Kang; Kyoung Doo Song; Jeong Kyong Lee
Journal:  Korean J Radiol       Date:  2022-01       Impact factor: 3.500

6.  A multicenter 10-year oncologic outcome of ultrasound-guided percutaneous microwave ablation of clinical T1 renal cell carcinoma: will it stand the test of time?

Authors:  Jie Yu; Hui Wang; Zhi-Gang Cheng; Fang-Yi Liu; Qin-Ying Li; Guang-Zhi He; Yan-Chun Luo; Xiao-Ling Yu; Zhi-Yu Han; Ping Liang
Journal:  Eur Radiol       Date:  2021-06-30       Impact factor: 5.315

Review 7.  Combination of ablation and embolization for intermediate-sized liver metastases from colorectal cancer: what can we learn from treating primary liver cancer?

Authors:  Matthew J Seager; Tobias F Jakobs; Ricky A Sharma; Steve Bandula
Journal:  Diagn Interv Radiol       Date:  2021-09       Impact factor: 2.630

8.  Ultrasound-Guided Percutaneous Laser Ablation of the Thyroid Gland in a Swine Model: Comparison of Ablation Parameters and Ablation Zone Dimensions.

Authors:  Fourat Ridouani; R Michael Tuttle; Mario Ghosn; Duan Li; Richard J Wong; James A Fagin; Sebastien Monette; Stephen B Solomon; Juan C Camacho
Journal:  Cardiovasc Intervent Radiol       Date:  2021-07-12       Impact factor: 2.797

9.  Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an ex vivo study.

Authors:  Antonia Grimm; Moritz Winkelmann; Jakob Weiß; Georg Gohla; Gunnar Blumenstock; Konstantin Nikolaou; Stephan Clasen; Rüdiger Hoffmann
Journal:  Eur Radiol Exp       Date:  2019-09-23

10.  MR-Guided High-Power Microwave Ablation in Hepatic Malignancies: Initial Results in Clinical Routine.

Authors:  Moritz T Winkelmann; Georg Gohla; Jens Kübler; Jakob Weiß; Stephan Clasen; Konstantin Nikolaou; Rüdiger Hoffmann
Journal:  Cardiovasc Intervent Radiol       Date:  2020-07-22       Impact factor: 2.740

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.