Multipolar RFA of the liver: Influence of intrahepatic vessels on ablation zones and appropriateness of CECT in detecting ablation dimensions - Results of an in-vivo porcine liver model.

Janis L Vahldiek1, Christoph Erxleben1, Keno Kyrill Bressem1, Ole Gemeinhardt2, Franz Poch2, Bernhard Hiebl3, Kai S Lehmann2, B Hamm1, Stefan M Niehues1.   

Abstract

BACKGROUND: Radiofrequency ablation (RFA) is an important treatment option for hepatic tumors and metastases. Post-ablation recurrence rates are reported up to 36.5 percent and seem to depend on tumor size, intrahepatic localization of tumors and adjacent hepatic vessels. Multipolar RFA has the potential to overcome/reduce these limitations. Experimental and standardized data on achievable lesion sizes, influence of hepatic vessels and non-invasive evaluation of complete ablation is still insufficient.
OBJECTIVES: The aim of this study was to evaluate the influence of intrahepatic vessels on shape and size of multipolar RF-ablation zones in healthy porcine liver and to evaluate the appropriateness of immediate post-ablation contrast-enhanced computed tomography (CECT) in detecting RF-ablation dimensions.
MATERIAL AND METHODS: We conducted multipolar RFAs in each of the livers of 10 healthy, narcotized and laparotomized domestic pigs by inserting three parallel probes with a constant probe distance and a constant energy supply. In 4 ablations we interrupted hepatic blood flow using Pringle's maneuver. Immediate post-ablation CECT scans were acquired. After euthanasia the livers were sliced perpendicularly to the probes at the probes' active centers. CECT scans were reconstructed equivalently in order to compare RF-lesion size and shape to the macroscopic sections.
RESULTS: In total, 19 RF-lesions were analyzed. Every RF-lesion that was ablated during physiological liver perfusion showed an irregular and cloverleaf-like shape (n = 15). Interrupting the hepatic blood flow during RFA led to well-defined, round and homogeneous ablation zones which were 3.8 times larger compared to RF-lesions ablated during continuous hepatic perfusion (n = 4). We found an excellent correlation between immediate post-ablation CECT slices and macroscopic sections when comparing RF-lesion diameters and area, although CECT tended to overestimate ablation dimensions.
CONCLUSIONS: The interruption of hepatic blood flow using Pringle's maneuver during multipolar RFA with three applicators significantly reduces heat sink effects of hepatic vessels and generates large and coherent ablation zones. This approach should be considered in each case of ablation planning adjacent to larger hepatic vessels or when ablating larger tumor volumes. Immediate post-ablation CECT has limited value in detecting incomplete RFA periprocedurally.

Entities:  

Keywords:  *contrast media; Interventional/methods; Liver perfusion; animals; catheter ablation; electrodes; imaging; multidetector computed tomography; radiography; three-dimensional/*methods

Mesh:

Year:  2018        PMID: 30347610     DOI: 10.3233/CH-189313

Source DB:  PubMed          Journal:  Clin Hemorheol Microcirc        ISSN: 1386-0291            Impact factor:   2.375


  5 in total

Review 1.  Complications Risk Assessment and Imaging Findings of Thermal Ablation Treatment in Liver Cancers: What the Radiologist Should Expect.

Authors:  Vincenza Granata; Roberta Fusco; Federica De Muzio; Carmen Cutolo; Sergio Venanzio Setola; Igino Simonetti; Federica Dell'Aversana; Francesca Grassi; Federico Bruno; Andrea Belli; Renato Patrone; Vincenzo Pilone; Antonella Petrillo; Francesco Izzo
Journal:  J Clin Med       Date:  2022-05-13       Impact factor: 4.964

2.  Evaluation of potential tissue heating during percutaneous drill-assisted bone sampling in an in vivo porcine study.

Authors:  Stefan M Niehues; Sefer Elezkurtaj; Keno K Bresssem; Bernd Hamm; Christoph Erxleben; Janis Vahldiek; Lisa C Adams
Journal:  Skeletal Radiol       Date:  2021-08-30       Impact factor: 2.199

3.  Perivascular vital cells in the ablation center after multibipolar radiofrequency ablation in an in vivo porcine model.

Authors:  F G M Poch; C A Neizert; B Geyer; O Gemeinhardt; S M Niehues; J L Vahldiek; K K Bressem; K S Lehmann
Journal:  Sci Rep       Date:  2021-07-06       Impact factor: 4.379

4.  Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an ex vivo study.

Authors:  Antonia Grimm; Moritz Winkelmann; Jakob Weiß; Georg Gohla; Gunnar Blumenstock; Konstantin Nikolaou; Stephan Clasen; Rüdiger Hoffmann
Journal:  Eur Radiol Exp       Date:  2019-09-23

5.  Influence of interapplicator distance on multibipolar radiofrequency ablation during physiological and interrupted liver perfusion in an in vivo porcine model.

Authors:  F G M Poch; C A Neizert; B Geyer; O Gemeinhardt; L Bruder; S M Niehues; J L Vahldiek; K K Bressem; M E Kreis; K S Lehmann
Journal:  Sci Rep       Date:  2020-10-01       Impact factor: 4.379