Literature DB >> 27553935

MR-guided microwave ablation in hepatic tumours: initial results in clinical routine.

Rüdiger Hoffmann1, Hansjörg Rempp2, David-Emanuel Keßler2, Jakob Weiß2, Philippe L Pereira3, Konstantin Nikolaou2, Stephan Clasen2.   

Abstract

OBJECTIVES: Evaluation of the technical success, patient safety and technical effectiveness of magnetic resonance (MR)-guided microwave ablation of hepatic malignancies.
METHODS: Institutional review board approval and informed patient consent were obtained. Fifteen patients (59.8 years ± 9.5) with 18 hepatic malignancies (7 hepatocellular carcinomas, 11 metastases) underwent MR-guided microwave ablation using a 1.5-T MR system. Mean tumour size was 15.4 mm ± 7.7 (7-37 mm). Technical success and ablation zone diameters were assessed by post-ablative MR imaging. Technique effectiveness was assessed after 1 month. Complications were classified according to the Common Terminology Criteria for Adverse Events (CTCAE). Mean follow-up was 5.8 months ± 2.6 (1-10 months).
RESULTS: Technical success and technique effectiveness were achieved in all lesions. Lesions were treated using 2.5 ± 1.2 applicator positions. Mean energy and ablation duration per tumour were 37.6 kJ ± 21.7 (9-87 kJ) and 24.7 min ± 11.1 (7-49 min), respectively. Coagulation zone short- and long-axis diameters were 31.5 mm ± 10.5 (16-65 mm) and 52.7 mm ± 15.4 (27-94 mm), respectively. Two CTCAE-2-complications occurred (pneumothorax, pleural effusion). Seven patients developed new tumour manifestations in the untreated liver. Local tumour progression was not observed.
CONCLUSIONS: Microwave ablation is feasible under near real-time MR guidance and provides effective treatment of hepatic malignancies in one session. KEY POINTS: • Planning, applicator placement and therapy monitoring are possible without using contrast enhancement • Energy transmission from the generator to the scanner room is safely possible • MR-guided microwave ablation provides effective treatment of hepatic malignancies in one session • Therapy monitoring is possible without applicator retraction from the ablation site.

Entities:  

Keywords:  Hepatocellular carcinoma; Liver metastases; MR guidance; Magnetic resonance fluoroscopy; Microwave ablation

Mesh:

Year:  2016        PMID: 27553935     DOI: 10.1007/s00330-016-4517-x

Source DB:  PubMed          Journal:  Eur Radiol        ISSN: 0938-7994            Impact factor:   5.315


  13 in total

1.  Simulated accuracy assessment of small footprint body-mounted probe alignment device for MRI-guided cryotherapy of abdominal lesions.

Authors:  Naoyuki Shono; Brian Ninni; Franklin King; Takahisa Kato; Junichi Tokuda; Takahiro Fujimoto; Kemal Tuncali; Nobuhiko Hata
Journal:  Med Phys       Date:  2020-03-28       Impact factor: 4.071

2.  Adaptation of Dictionary Learning for Electrode Displacement Elastography.

Authors:  Robert M Pohlman; Tomy Varghese
Journal:  Annu Int Conf IEEE Eng Med Biol Soc       Date:  2020-07

3.  Optimized Echo Decorrelation Imaging Feedback for Bulk Ultrasound Ablation Control.

Authors:  Mohamed A Abbass; Allison-Joy Garbo; Neeraja Mahalingam; Jakob K Killin; T Douglas Mast
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2018-06-14       Impact factor: 2.725

4.  Differential Imaging of Liver Tumors before and after Microwave Ablation with Electrode Displacement Elastography.

Authors:  Robert M Pohlman; James L Hinshaw; Timothy J Ziemlewicz; Meghan G Lubner; Shane A Wells; Fred T Lee; Marci L Alexander; Kelly L Wergin; Tomy Varghese
Journal:  Ultrasound Med Biol       Date:  2021-05-16       Impact factor: 3.694

5.  In vivo ultrasound thermal ablation control using echo decorrelation imaging in rabbit liver and VX2 tumor.

Authors:  Mohamed A Abbass; Syed A Ahmad; Neeraja Mahalingam; K Sameer Krothapalli; Jack A Masterson; Marepalli B Rao; Peter G Barthe; T Douglas Mast
Journal:  PLoS One       Date:  2019-12-05       Impact factor: 3.240

6.  MR-Guided Microwave Ablation in T1 Renal Cell Carcinoma: Initial Results in Clinical Routine.

Authors:  Zhaonan Li; De-Chao Jiao; Chaoyan Wang; Wenguang Zhang; Jing Li; Xinwei Han
Journal:  Biomed Res Int       Date:  2021-05-13       Impact factor: 3.411

Review 7.  Challenges Facing Percutaneous Ablation in the Treatment of Hepatocellular Carcinoma: Extension of Ablation Criteria.

Authors:  Yanzhao Zhou; Yi Yang; Bingyan Zhou; Zhengzheng Wang; Ruili Zhu; Xun Chen; Jingzhong Ouyang; Qingjun Li; Jinxue Zhou
Journal:  J Hepatocell Carcinoma       Date:  2021-06-21

8.  Artefact and ablation performance of an MR-conditional high-power microwave system in bovine livers: an ex vivo study.

Authors:  Antonia Grimm; Moritz Winkelmann; Jakob Weiß; Georg Gohla; Gunnar Blumenstock; Konstantin Nikolaou; Stephan Clasen; Rüdiger Hoffmann
Journal:  Eur Radiol Exp       Date:  2019-09-23

9.  MR-Guided High-Power Microwave Ablation in Hepatic Malignancies: Initial Results in Clinical Routine.

Authors:  Moritz T Winkelmann; Georg Gohla; Jens Kübler; Jakob Weiß; Stephan Clasen; Konstantin Nikolaou; Rüdiger Hoffmann
Journal:  Cardiovasc Intervent Radiol       Date:  2020-07-22       Impact factor: 2.740

Review 10.  Advanced Techniques in the Percutaneous Ablation of Liver Tumours.

Authors:  Terrence Ch Hui; Justin Kwan; Uei Pua
Journal:  Diagnostics (Basel)       Date:  2021-03-24
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.