| Literature DB >> 31398851 |
Jae-Hyun Lee1, Je-Hyeon Yun2, Jung-Suk Han2, In-Sung Luke Yeo2, Hyung-In Yoon3.
Abstract
Research on whether the number or location of missing teeth affects the accuracy of intraoral scanners in partial edentulous patients is scarce. This study aimed to evaluate the precision of complete-arch scan data of various partial edentulous arches acquired by intraoral scanners. Five different maxillary models were scanned using Carestream CS3600 and Medit i500 scanners. The models employed here were control: Fully dentate; Case 1: Missing a right second premolar and a first molar; Case 2: Missing a right second premolar, a first molar, both left premolars, and a left first molar; Case 3: Missing four incisors and a right canine; and Case 4: Missing four incisors, a left second premolar, and a first molar. Six scans per group were performed and the resulting two datasets were paired to analyze the precision of each group (n = 15). Two-way ANOVA was performed (α = 0.05). The root mean square (RMS) error values in Cases 2, 3, and 4 were significantly higher than those in Case 1 and control. The RMS values of the two intraoral scanners were not significantly different. Scanning precision was significantly lower for both devices when used for scanning dental arches with ≥5 missing teeth.Entities:
Keywords: accuracy; digital dentistry; intraoral scanner; partial edentulism; precision
Year: 2019 PMID: 31398851 PMCID: PMC6722554 DOI: 10.3390/jcm8081187
Source DB: PubMed Journal: J Clin Med ISSN: 2077-0383 Impact factor: 4.241
Figure 1Classification of maxillary models for the scan. The simulated soft tissue area was fabricated from hard acrylic resin material.
Results of two-way ANOVA with the root mean square (RMS) as the dependent variable.
| Source | Type III Sum of Squares |
| Mean Squares |
|
|
|---|---|---|---|---|---|
| Dependent variable: RMS | |||||
| Scanner | 4408.63 | 1 | 4408.63 | 3.68 | 0.057 |
| Case | 94,020.02 | 4 | 23,505.00 | 19.61 *** | <0.001 |
| Scanner × Case | 1262.79 | 4 | 315.70 | 0.26 | 0.901 |
| Error | 167,795.36 | 140 | 1198.54 | ||
*** p < 0.001.
Mean and standard deviation (SD) of root mean square (RMS) values.
| Group | RMS (unit: µm) | ||||
|---|---|---|---|---|---|
| CS3600 | i500 | ||||
| Mean | SD | Mean | SD | ||
|
| 44.37 a | 6.04 | 52.30 a | 14.99 | 0.53 |
| Case 1 | 49.57 a | 9.00 | 58.43 a | 21.45 | 0.48 |
| Case 2 | 96.31 b | 35.29 | 100.22 b | 44.74 | 0.76 |
| Case 3 | 85.59 b | 41.85 | 106.71 b | 48.92 | 0.10 |
| Case 4 | 103.28 b | 30.86 | 115.66 b | 53.28 | 0.33 |
Bonferroni: a < b. Means with the same superscript in each column are not significantly different from each other based on the Bonferroni test (p > 0.05). * p-value between CS3600 and i500.
Figure 2Box plot demonstrating precision measurement. The amount of deviation is represented as root mean square (RMS). * Significantly different from control group scanned with the same intraoral scanner (p ≤ 0.05).
Figure 3Representative images from a color map for assessing precision. The color maps for the CS3600 datasets are displayed to visualize the distribution of deviation. Deviation sites are mainly observed on the side that includes the partially edentulous area.