Investigation of accuracy and reproducibility of abutment position by intraoral scanners.

Shota Fukazawa1, Chikayuki Odaira1, Hisatomo Kondo2.   

Abstract

PURPOSE: We examined the accuracy, including trueness and precision, of the intraoral scanners comparing with laboratory scanner to reveal the error level of intraoral scanners.
METHODS: Measurements were performed using a computer numerical control coordinate measuring machine (CNCCMM) of the reference models as a control. Subsequently, intraoral scanners and a laboratory scanner were used for measurements of the reference trueness and precision of the distance were evaluated by image analyzing software.
RESULTS: With regard to reference model, there was a significant difference between in the trueness measured by C.O.S. (COS) and that measured by the other scanners. The trueness measured by the second-generation 3M™ true definition scanner (TDS2) and third-generation 3M™ true definition scanner (TDS3) was bigger than the one by TRIOS (TR) and KaVo (KA). With regard to reference model "B," error of the trueness measured by COS was significantly bigger, compared with the one measured by the other scanners. However, error range of intraoral scanners, except for COS, was considerably small and it should be covered with cement space.
CONCLUSIONS: The results of this study indicated that an optical impression method with an intraoral scanner could be applied to the implant therapy for multiple teeth missing.
Copyright © 2017 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  CAD/CAM; Intraoral scanner; Optical impression; Precision; Trueness

Mesh:

Year:  2017        PMID: 28216020     DOI: 10.1016/j.jpor.2017.01.005

Source DB:  PubMed          Journal:  J Prosthodont Res        ISSN: 1883-1958            Impact factor:   4.642


  19 in total

1.  A new 3D-method to assess the inter implant dimensions in patients - A pilot study.

Authors:  Alexander Schmidt; Jan-Wilhelm Billig; Maximiliane A Schlenz; Bernd Wöstmann
Journal:  J Clin Exp Dent       Date:  2020-02-01

2.  In Vitro Comparison of Three Intraoral Scanners for Implant-Supported Dental Prostheses.

Authors:  Vitória Costa; António Sérgio Silva; Rosana Costa; Pedro Barreiros; Joana Mendes; José Manuel Mendes
Journal:  Dent J (Basel)       Date:  2022-06-15

3.  Effect of Scanner Type and Scan Body Location on the Accuracy of Mandibular Complete-Arch Digital Implant Scans: An In Vitro Study.

Authors:  Gülce Çakmak; Hakan Yilmaz; Alejandro Treviño Santos; Ali Murat Kökat; Burak Yilmaz
Journal:  J Prosthodont       Date:  2021-09-21       Impact factor: 3.485

4.  Accuracy of commercial intraoral scanners.

Authors:  Mattia Sacher; Georg Schulz; Hans Deyhle; Kurt Jäger; Bert Müller
Journal:  J Med Imaging (Bellingham)       Date:  2021-05-24

Review 5.  Intraoral scanners in dentistry: a review of the current literature.

Authors:  Francesco Mangano; Andrea Gandolfi; Giuseppe Luongo; Silvia Logozzo
Journal:  BMC Oral Health       Date:  2017-12-12       Impact factor: 2.757

Review 6.  Trueness and precision of digital implant impressions by intraoral scanners: a literature review.

Authors:  Minoru Sanda; Keita Miyoshi; Kazuyoshi Baba
Journal:  Int J Implant Dent       Date:  2021-07-27

7.  Bias Evaluation of the Accuracy of Two Extraoral Scanners and an Intraoral Scanner Based on ADA Standards.

Authors:  Naiyu Cui; Jiayin Wang; Xingyu Hou; Shixun Sun; Qixuan Huang; Ho-Kyung Lim; HongXin Cai; Qi Jia; Eui-Seok Lee; Heng Bo Jiang
Journal:  Scanning       Date:  2021-06-10       Impact factor: 1.932

8.  Marginal and internal fit and intaglio surface trueness of interim crowns fabricated from tooth preparation of four finish line locations.

Authors:  Keunbada Son; Young-Tak Son; Ji-Min Lee; Kyu-Bok Lee
Journal:  Sci Rep       Date:  2021-07-06       Impact factor: 4.379

9.  Effect of Tooth Types on the Accuracy of Dental 3D Scanners: An In Vitro Study.

Authors:  Keunbada Son; Kyu-Bok Lee
Journal:  Materials (Basel)       Date:  2020-04-09       Impact factor: 3.623

10.  Applying intraoral scanner to residual ridge in edentulous regions: in vitro evaluation of inter-operator validity to confirm trueness.

Authors:  Akinori Tasaka; Yuuki Uekubo; Tomoharu Mitsui; Takao Kasahara; Takuya Takanashi; Shinya Homma; Satoru Matsunaga; Shinichi Abe; Masao Yoshinari; Yasutomo Yajima; Kaoru Sakurai; Shuichiro Yamashita
Journal:  BMC Oral Health       Date:  2019-12-02       Impact factor: 2.757

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.