Trueness and precision of digital impressions obtained using an intraoral scanner with different head size in the partially edentulous mandible.

Hironari Hayama1, Kenji Fueki2, Juro Wadachi1, Noriyuki Wakabayashi1.   

Abstract

PURPOSE: It remains unclear whether digital impressions obtained using an intraoral scanner are sufficiently accurate for use in fabrication of removable partial dentures. We therefore compared the trueness and precision between conventional and digital impressions in the partially edentulous mandible.
METHODS: Mandibular Kennedy Class I and III models with soft silicone simulated-mucosa placed on the residual edentulous ridge were used. The reference models were converted to standard triangulated language (STL) file format using an extraoral scanner. Digital impressions were obtained using an intraoral scanner with a large or small scanning head, and converted to STL files. For conventional impressions, pressure impressions of the reference models were made and working casts fabricated using modified dental stone; these were converted to STL file format using an extraoral scanner. Conversion to STL file format was performed 5 times for each method. Trueness and precision were evaluated by deviation analysis using three-dimensional image processing software.
RESULTS: Digital impressions had superior trueness (54-108μm), but inferior precision (100-121μm) compared to conventional impressions (trueness 122-157μm, precision 52-119μm). The larger intraoral scanning head showed better trueness and precision than the smaller head, and on average required fewer scanned images of digital impressions than the smaller head (p<0.05). On the color map, the deviation distribution tended to differ between the conventional and digital impressions.
CONCLUSIONS: Digital impressions are partially comparable to conventional impressions in terms of accuracy; the use of a larger scanning head may improve the accuracy for removable partial denture fabrication.
Copyright © 2018 Japan Prosthodontic Society. Published by Elsevier Ltd. All rights reserved.

Entities:  

Keywords:  Digital impression; Intraoral scanner; Partially edentulous mandible; Precision; Trueness

Mesh:

Year:  2018        PMID: 29502933     DOI: 10.1016/j.jpor.2018.01.003

Source DB:  PubMed          Journal:  J Prosthodont Res        ISSN: 1883-1958            Impact factor:   4.642


  15 in total

1.  The transfer accuracy of digital and conventional full-arch impressions influenced by fixed orthodontic appliances: a reference aid-based in vitro study.

Authors:  Maximiliane Amelie Schlenz; Katharina Klaus; Alexander Schmidt; Bernd Wöstmann; Marco Mersmann; Sabine Ruf; Niko Christian Bock
Journal:  Clin Oral Investig       Date:  2022-09-15       Impact factor: 3.606

2.  The effect of the improperly scanned scan body images on the accuracy of virtual implant positioning in computer-aided design software.

Authors:  Se-Won Park; Yong-Do Choi; Du-Hyeong Lee
Journal:  J Adv Prosthodont       Date:  2020-06-18       Impact factor: 1.904

3.  Intraoral scanning to fabricate complete dentures with functional borders: a proof-of-concept case report.

Authors:  Alexey Unkovskiy; Eugen Wahl; Anne Teresa Zander; Fabian Huettig; Sebastian Spintzyk
Journal:  BMC Oral Health       Date:  2019-03-13       Impact factor: 2.757

4.  Repeatability of Intraoral Scanners for Complete Arch Scan of Partially Edentulous Dentitions: An In Vitro Study.

Authors:  Jae-Hyun Lee; Je-Hyeon Yun; Jung-Suk Han; In-Sung Luke Yeo; Hyung-In Yoon
Journal:  J Clin Med       Date:  2019-08-08       Impact factor: 4.241

5.  Influence of Applied Liquid-Type Scanning-Aid Material on the Accuracy of the Scanned Image: An In Vitro Experiment.

Authors:  Hyun-Su Oh; Young-Jun Lim; Bongju Kim; Won Hyeon Kim; Myung-Joo Kim; Ho-Beom Kwon
Journal:  Materials (Basel)       Date:  2020-04-27       Impact factor: 3.623

6.  CAD/CAM Customized Glass Fiber Post and Core With Digital Intraoral Impression: A Case Report.

Authors:  Antonio Libonati; Virginia Di Taranto; Gianni Gallusi; Edoardo Montemurro; Vincenzo Campanella
Journal:  Clin Cosmet Investig Dent       Date:  2020-02-10

7.  Can Dental Office Lighting Intensity Conditions Influence the Accuracy of Intraoral Scanning?

Authors:  Anca Jivanescu; Andrei-Bogdan Faur; Raul Nicolae Rotar
Journal:  Scanning       Date:  2021-05-27       Impact factor: 1.932

8.  Digital evaluation of axial displacement by implant-abutment connection type: An in vitro study.

Authors:  Sung-Jun Kim; KeunBaDa Son; Kyu-Bok Lee
Journal:  J Adv Prosthodont       Date:  2018-10-22       Impact factor: 1.904

9.  Accuracy of Digital and Conventional Full-Arch Impressions in Patients: An Update.

Authors:  Alexander Schmidt; Leona Klussmann; Bernd Wöstmann; Maximiliane Amelie Schlenz
Journal:  J Clin Med       Date:  2020-03-04       Impact factor: 4.241

10.  Accuracy on Scanned Images of Full Arch Models with Orthodontic Brackets by Various Intraoral Scanners in the Presence of Artificial Saliva.

Authors:  Jihu Song; Minji Kim
Journal:  Biomed Res Int       Date:  2020-02-27       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.