| Literature DB >> 31186023 |
Jan Hofmaier1, Raphael Bodensohn2, Sylvia Garny2, Indrawati Hadi2, Daniel F Fleischmann2,3,4, Michael Eder2, Yavuz Dinc2, Michael Reiner2, Stefanie Corradini2, Katia Parodi5, Claus Belka2,3, Maximilian Niyazi2,3.
Abstract
BACKGROUND: In this dosimetric study, a dedicated planning tool for single isocenter stereotactic radiosurgery for multiple brain metastases using dynamic conformal arc therapy (DCAT) was compared to standard volumetric modulated arc therapy (VMAT).Entities:
Keywords: Brain metastases; Dynamic conformal arc therapy; Single Isocenter; Stereotactic radiosurgery; VMAT
Mesh:
Year: 2019 PMID: 31186023 PMCID: PMC6560766 DOI: 10.1186/s13014-019-1315-z
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1Arc setup and beams-eye-view for one of the arcs for a patient with 6 lesions. 3 of the lesions are treated simultaneously by the selected arc (blue)
Patient characteristics
| Number of patients | 20 |
|---|---|
| Sex | Male: 10, Female: 10 |
| Median age at RT [years] | 62, range: 51–77 |
| Total number of treated lesions | 66 |
| Median number of treated lesions per patient | 3, range: 2–6 |
| Median target volume [cm3] | 0.8, range: 0.1–11.9 |
| Median prescribed isodose [Gy] | 19, range: 15–20 |
Results for all investigated metrics. While some metrics were investigated on a plan-wise basis (cumulative MU, and the brain receiving 4, 5 and 8 Gy) others were examined on a lesion-wise basis (V10Gy, V12Gy, conformity index, gradient index)
| MBSRS | Monaco® | ||||
|---|---|---|---|---|---|
Plan data |
|
| |||
| Cumulative MU | 4569 (3543–7790) | 17 | 5840 (4089–10,769) | 3 | < 0.05 |
| V4Gy (brain) [cm3] | 45.6 (18.5–215.9) | 19 | 123.1 (37.5–418.6) | 1 | < 0.05 |
| V5Gy (brain) [cm3] | 33.7 (13.0–120.5) | 19 | 69.1 (26.0–273.6) | 1 | < 0.05 |
| V8Gy (brain) [cm3] | 17.4 (6.3–52.6) | 19 | 26.6 (10.6–86.4) | 1 | < 0.05 |
PTV data |
|
| |||
| V10Gy (healthy brain) [cm3] | 3.2 (0.4–19.3) | 55 | 4.9 (1.0–19.9) | 4 | < 0.05 |
| V12Gy (healthy brain) [cm3] | 2.1 (0.1–13.1) | 53 | 3.1 (0.5–13.9) | 6 | < 0.05 |
PTV data without PTVs overlapping in half prescribed isodose |
|
| |||
| Paddick CI | 0.75 (0.58–0.89) | 30 | 0.73 (0.38–0.88) | 22 | < 0.05 |
| GIPTV | 5.99 (3.50–15.73) | 48 | 7.17 (3.35–33.00) | 5 | < 0.05 |
Fig. 2Dose distributions for a patient with 5 metastases (only 4 of them are visible here). The prescribed isodose was 20 Gy for all lesions. The upper three panels show transverse, coronal and sagittal (left to right) slices for the DCAT plan, the lower panels show the respective slices for the VMAT plan. Color wash scales from 6.7 to 21 Gy
Fig. 3The relative difference in V10Gy and V12Gy between DCAT and VMAT scattered against the sphericity of the corresponding PTV. The reported volumes are normalized to the value obtained with the MBSRS tool. A sphericity value of 1 corresponds to a perfect sphere, values < 1 to shapes deviating from a sphere. Markers above the red dotted line correspond to lesions where DCAT was superior, markers below to lesions where VMAT was superior