Comparison of volumetric-modulated arc therapy and dynamic conformal arc treatment planning for cranial stereotactic radiosurgery.

Jessica Molinier1, Christine Kerr, Sebastien Simeon, Norbert Ailleres, Marie Charissoux, David Azria, Pascal Fenoglietto.   

Abstract

The aim was to analyze arc therapy techniques according to the number and position of the brain lesions reported by comparing dynamic noncoplanar conformal arcs (DCA), two coplanar full arcs (RAC) with volumetric-modulated arc therapy (VMAT), multiple noncoplanar partial arcs with VMAT (RANC), and two full arcs with VMAT and 10° table rotation (RAT). Patients with a single lesion (n= 10), multiple lesions (n = 10) or a single lesion close to organs at risk (n = 5) and previously treated with DCA were selected. For each patient, the DCA treatment was replanned with all VMAT techniques. All DCA plans were compared with VMAT plans and evaluated in regard to the different quality indices and dosimetric parameters. For single lesion, homogeneity index (HI) better results were found for the RANC technique (0.17 ± 0.05) compared with DCA procedure (0.27± 0.05). Concerning conformity index (CI), the RAT technique gave higher and better values (0.85 ± 0.04) compared with those obtained with the DCA technique (0.77 ± 0.05). DCA improved healthy brain protection (8.35 ± 5.61 cc vs. 10.52 ± 6.40 cc for RANC) and reduced monitor unit numbers (3046 ± 374 MU vs. 4651 ± 736 for RANC), even if global room occupation was higher. For multiple lesions, VMAT techniques provided better HI (0.16) than DCA (0.24 ± 0.07). The CI was improved with RAT (0.8 ± 0.08 for RAT vs. 0.71 ± 0.08 for DCA). The V10Gy healthy brain was better protected with DCA (9.27 ± 4.57 cc). Regarding the MU numbers: RANC < RAT< RAC < DCA. For a single lesion close to OAR, RAT achieved high degrees of homogeneity (0.27 ± 0.03 vs. 0.53 ± 0.2 for DCA) and conformity (0.72± 0.06vs. 0.56 ± 0.13 for DCA) while sparing organs at risk (Dmax = 12.36 ± 1.05Gyvs. 14.12 ± 0.59 Gy for DCA, and Dmean = 3.96 ± 3.57Gyvs. 4.72 ± 3.28Gy for DCA). On the other hand, MU numbers were lower with DCA (2254 ± 190 MUvs. 3438 ± 457 MU for RANC) even if overall time was inferior with RAC. For a single lesion, DCA provide better plan considering low doses to healthy brain even if quality indexes are better for the others techniques. For multiple lesions, RANC seems to be the best compromise, due to the ability to deliver a good conformity and homogeneity plan while sparing healthy brain tissue. For a single lesion close to organs at risk, RAT is the most appropriate technique.

Entities:  

Mesh:

Year:  2016        PMID: 26894335      PMCID: PMC5690199          DOI: 10.1120/jacmp.v17i1.5677

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

Stereotactic radiosurgery consists in delivering a single high dose on a small target while sparing the surrounding healthy tissue. Consequently, this treatment requires a rigorous precision in immobilization and patient positioning. Cranial stereotactic radiosurgery was initially performed with a stereotactic invasive head frame, such as the Leksell frame. Several relocatable stereotactic frames, like the Gill‐Thomas‐Cosman frame, were proposed to replace the frame‐based system, but the setup accuracy was not convincing. , , The recent advances in imaging development allowed assessing the feasibility of frameless radiosurgery. Two image‐guidance systems are now available: the ExacTrac (ET) system from Brainlab and On‐Board Imager (OBI) system from Varian. Coupled with a robotic couch, ET allows a six‐dimensional patient positioning before and during irradiation from two floor‐mounted kV X‐ray tubes. The OBI system generates either two orthogonal planar images or a three‐dimensional image with the use of cone‐beam computed tomography (CBCT). Several studies compared these two systems and concluded they had similar accuracy. , These advances considerably increased the setup precision, and allowed the use of different types of frameless masks with a comparable accuracy to that of the invasive frame approach. , , , , , Besides a high level of precision for positioning, stereotactic radiosurgery requires a sharp dose falloff to spare the neighboring tissue. Many modalities are now capable of performing these radiosurgery treatments. With sources, GammaKnife created collimated convergent beams to the target. , CyberKnife, a linear accelerator coupled with a 6D robotic arm, is shown to deliver multiple nonisocentric pencil beams. The linear accelerator (linac) also belongs to this category. The main techniques used with linac are the dynamic conformal arcs, , the fixed‐field, intensity‐modulated radiation therapy (IMRT) , and, more recently, the volumetric‐modulated arc therapy (VMAT). , , In this study, four rotational techniques for cranial stereotactic radiosurgery were compared: the use of two coplanar full arcs with VMAT () and of multiple noncoplanar partial arcs with VMAT (), and the two full arcs with VMAT and the 10° table rotation () techniques. The aim of the study was to analyze the dosimetric parameters of different arc therapy techniques according to the number and the position of the brain lesions reported.

MATERIALS AND METHODS

Patients

Twenty‐five patients previously treated with DCA were selected for this study leading to 100 dosimetric plans. Ten patients were irradiated for a single lesion, 10 for multiple lesions, and 5 for single lesion close to organs at risk. The diagnoses were mainly lung and breast metastasis (respectively, 32% and 16% of patients). The median PTV was 2.4 cm3 (range 0.68–13.7 cm3). The mean prescribed dose was 21 Gy (range 14–25 Gy). The mean normalization isodose (isodose covering 99% of the PTV) was 81.2%. In the case of multiple lesions, patients with the same prescribed dose to each target were selected. Patient and lesion characteristics are summarized in Table 1.
Table 1

Patient characteristics.

Patient Diagnosis PTV Volume (cc) Prescribed Dose (Gy) Normalization Isodose (%) No. of Arcs
Single Lesion
1glioma3.8621844
21 ovary metastasis3.5122.588.14
3glioma2.152581.44
41 breast metastasis13.72079.24
51 ovary metastasis13.322.584.44
61 melanoma metastasis1.642582.93
71 breast metastasis2.622.587.53
81 lung metastasis2.422.583.53
91 lung metastasis1.522.582.54
101 lung metastasis7.320824
Multiple Lesions
112 pancreas metastasis 0.69+2.68 22.570 3+4
124 breast metastasis 0.5+3.17+2.78+0.46 22.592.4 4+3+4+4
132 lung metastasis 1.91+0.57 25 5+4
142 melanoma metastasis 0.79+0.56 2581.8 4+4
152 lung metastasis 2.78+1.79 22.586.3 4+4
162 melanoma metastasis 1.2+2.6 22.582 3+3
172 breast metastasis 0.9+1.5 22.582.4 4+4
182 lung metastasis 1.69+5 22.583.6 4+3
192 lung metastasis 4.3+5.1 22.584.6 4+4
202 lung metastasis 5.3+2.4 2081.5 4+3
Single Lesion Close to OAR
21Neurinoma4.811470.53
22AVM2.871461.73
23Meningeoma2.171572.93
24Neurinoma7.21480.53
25Schwannoma1.11574.73
Patient characteristics.

Cranial stereotactic radiosurgery procedure

Each patient was imaged by computed tomography (CT) (GE Healthcare, Waukesha, WI) using 1.25 mm spacing and by a T1‐weighted magnetic resonance (MR) (Siemens Heathcare, Erlangen, Germany) scan using 1 mm spacing. The images were imported into iPlan RT Image (v.4.1; Brainlab AG Feldkirchen, Germany). A neurosurgeon outlined the gross tumor volume (GTV) and the organs at risk on the registered scans. A 2 mm margin was added to GTV to create the planning target volume (PTV), and the healthy brain was defined as the total brain volume excluding PTV. Each treatment plan was produced using iPlan RT Dose (v.4.5; Brainlab) and several dynamic noncoplanar conformal arcs (3–4 per lesion). Couch rotations varied according to the location of the lesion and were generally spaced out 40°. The gantry rotations were ranged from 330° to 210° and from 30° to 150°. Dose distributions were calculated with the pencil beam convolution (PBC) algorithm. The treatment plan was validated by physicians with regard to different criteria: the PTV coverage by 80% isodose, the healthy brain tissue volume receiving 10 Gy ( in our department per the Minniti et al. study ), and doses to OAR in some cases. Finally, treatments were delivered on a Novalis TrueBeam STx linear accelerator (Varian Medical Systems, Helsinki, Finland) equipped with the Varian High Definition 120 MLC, beam energy of 6 MV, and 600 MU/min dose rate.

VMAT treatment planning

For each patient, three additional VMAT plans were created: two coplanar full arcs (), multiple noncoplanar partial arcs (), and two noncoplanar full arcs with 10° table rotations (). These modalities are shown in Fig. 1.
Figure 1

Examples of two coplanar full arcs (left), noncoplanar partial arcs (center), and two full arcs with table rotations (right).

Examples of two coplanar full arcs (left), noncoplanar partial arcs (center), and two full arcs with table rotations (right).

The technique consisted in two full arcs with a gantry rotation from a starting angle of 179° and stopped at 181° in the counterclockwise direction and reverse for the second arc. The collimator rotations were 45° and 315°, and the couch angle was set at 0°. The isocenter coordinates were similar to the DCA in the case of a single lesion, and the Eclipse automated isocenter was used for multiple lesions. For single lesions, the beam arrangement was the same as for the DCA (3–4 arcs per lesion). In case of multiple lesions, four noncoplanar arcs were used with a single isocenter, compared with a multiple isocenter treatment for DCA. They were similar for every case of multiple lesions. The two first arcs started at 30° and stopped at 150° with 45° and 315° collimator angles and 70° and 30° couch rotations, respectively. The two others were ranged from 330° to 210°, with 0° and 90° collimator angles and 330° and 290° couch rotations, respectively. This beam arrangement allowed covering the whole brain with a single isocenter. The isocenter was placed using the Eclipse automated isocenter tool. The beam configuration (fields and isocenter) was identical to that of the , but 10° and 350° couch rotations were added for the two arcs. The VMAT plans were generated using Eclipse treatment planning (v10, Varian Medical Systems, Helsinki, Finland) and planned on the same CT scans. They were designed using a progressive resolution algorithm (PRO, v10.0.28, Varian Medical Systems), and calculated using the anisotropic analytical algorithm (AAA, v10.0.28; Varian Medical Systems). All VMAT plans were optimized from results obtained with DCA plans. Thus, of healthy brain and and for OAR (if necessary) of the DCA plan were used as optimization objectives for VMAT plans. For the study, treatment plans of each patient were compared with the same PTV coverage. Consequently, for each patient, VMAT plans were optimized to deliver the same peripheral dose as for the DCA.

Plan evaluation

DCA treatment plans are the benchmark in our department. They were compared with each VMAT plan. All of them were evaluated using quality indexes for PTVs and the dose volume histograms (DVH) for healthy brain and OAR. The target coverage was compared with conformity and homogeneity indexes (CI and HI). The conformity index chosen was the Paddick CI defined as: The normalization isodose was defined as the isodose covering 99% of the PTV. This index estimated both the PTV coverage and the surrounding tissue irradiation. CI was 1 when the PTV was covered by the normalization isodose while sparing the whole healthy tissue. The HI was defined as the ratio of the difference between the maximum and minimum doses to the mean dose for the PTV: The HI assessed the dose distribution homogeneity within the PTV. The more the dose distribution was homogenous in the target the more the index reached 0. Further, for healthy brain tissue was recorded for all cases except for the case of a single lesion close to OAR. In this particular case, prescribed dose was 14 Gy and 10 Gy isodose was generally target coverage isodose. Consequently, 10 Gy isodose was not representative of the dose received by normal brain. In addition, for single lesion close to OAR, maximum and mean doses( and ) of the OAR were compared. Finally, the number of monitor units (MUs) was assessed for each technique.

RESULTS

The quality indexes and the MUs results for the three cases (single lesion, multiple lesions, and single lesion close to OAR) are summarized in 2, 3, 4, respectively. Doses to healthy brain (cases 1 and 2) and OAR (case 3) are presented in 2, 3, 4.
Table 2

Comparison of CI, HI, and MU for DCA, , and in single lesion case.

CI HI MU
Patient DCA RANC RAT DCA RAC RANC RAT DCA RAC RANC RAT
10.740.820.870.870.250.210.180.192876552146945804
20.670.840.860.850.170.170.160.203173384851316362
30.720.730.760.780.300.260.170.193693619352725636
40.840.900.910.910.360.240.140.232645394739164517
50.820.850.850.870.270.250.230.273071335741823930
60.800.820.800.820.280.220.270.242851746858127518
70.790.860.840.870.210.160.150.132686599439115371
80.800.870.850.870.270.160.130.162673515335605236
90.730.770.810.810.280.210.140.203599634152836396
100.790.890.860.890.290.220.120.173193519747465204
Mean0.770.840.840.850.270.210.170.203046530246515597
SD0.050.050.040.040.050.040.050.0437412857361014
Table 3

Comparison of CI, HI, and MU for DCA, , and in multiple lesions case.

CI HI MU
Patient DCA RAC RANC RAT DCA RAC RANC RAT DCA RAC RANC RAT
110.650.730.690.760.350.200.170.166166411035664074
120.600.700.740.740.130.230.190.2112090457342514617
130.650.650.600.660.140.180.180.188104568551295762
140.640.650.590.680.260.140.150.136982433742034134
150.710.730.750.820.220.160.150.166199550041025425
160.730.860.840.860.290.140.160.135697608648956297
170.700.800.740.810.270.140.140.126715619458426189
180.800.850.800.850.180.140.150.146477616850656103
190.850.880.820.880.260.140.190.175597653743716511
200.800.900.820.900.300.160.190.165266607860435848
Mean0.710.770.740.800.240.160.170.166929522747465496
SD0.080.100.090.080.070.030.020.031984873792905
Table 4

Comparison of CI, HI, and MU for DCA, , and in the case of single lesion close to an OAR.

CI HI MU
Patient DCA RAC RANC RAT DCA RAC RANC RAT DCA RAC RANC RAT
210.640.620.700.710.470.350.290.252121372028413392
220.420.500.540.720.770.310.300.302080430537263585
230.400.720.520.790.700.300.290.252354385236894670
240.680.730.590.740.350.300.340.292180348138753212
250.640.660.650.620.340.210.230.252539482730613823
Mean0.560.650.600.720.530.290.290.272255403734383736
SD0.130.090.070.060.200.050.040.03190534457569
Figure 2

Comparison of healthy brain for DCA, , and in single lesion case.

Figure 3

Comparison of healthy brain for DCA, , and in multiple lesion case.

Figure 4

Comparison of doses to OAR ( and ) for DCA, , and in the case of single lesion close to an OAR.

Comparison of CI, HI, and MU for DCA, , and in single lesion case. Comparison of CI, HI, and MU for DCA, , and in multiple lesions case. Comparison of CI, HI, and MU for DCA, , and in the case of single lesion close to an OAR. Comparison of healthy brain for DCA, , and in single lesion case. Comparison of healthy brain for DCA, , and in multiple lesion case. Comparison of doses to OAR ( and ) for DCA, , and in the case of single lesion close to an OAR.

Case 1: Single lesion

The VMAT techniques improved the conformity index with for and and for , whereas the DCA reached The HI results were for the technique compared with for the DCA. However, DCA enhanced healthy brain protection with a of , while the best VMAT technique was and a of . For Patients 4 and 5, treatment plans were accepted despite an out of tolerance for normal brain (13.37 and 20.89 cc, respectively) because of the large target volume (13.7 and 13.3 cc, respectively). The DCA technique reduced the MU number with For comparison, the MU number was , , and for , and , respectively.

Case 2: Multiple lesions

provided higher conformity degrees with compared with DCA with HI results were equivalent for VMAT techniques with For the DCA, HI was The healthy brain protection was improved with DCA: was of whereas reached . Concerning the MU numbers, was the technique delivering the least MUs with a decrease of 31%, 14%, and 13% compared with the DCA, , and techniques.

Case 3: Single lesion close to an OAR

The increased the conformity and the target coverage homogeneity ( and ) compared with the DCA ( and ). On the other hand, the MU numbers were lower with the DCA with a MU decrease of 44%, 34%, and 40% compared with the , and techniques, respectively. Concerning doses to OAR, was reduced with the with . For the DCA, the was . The was slightly similar for all techniques.

DISCUSSION

In this study, the algorithms used for the DCA (PBC with iPlan RT Dose) and VMAT (AAA with Eclipse) were different. However, they were specifically commissioned to perform cranial stereotactic treatment. Because the same characteristics (dose grid step, heterogeneity correction) were used, the variances were not significant between the two algorithms for cranial treatment. All VMAT plans, in particular the were clearly superior in terms of target conformity and homogeneity compared with the DCA. This was probably due to the use of an inverse optimization algorithm which adjusted the dose to the target. Lagerwaard et al. compared the VMAT plan with a single dynamic conformal arc with five noncoplanar arcs for the treatment of vestibular schwannomas. They also concluded that VMAT plans consistently achieved a higher CI and decreased in areas of low‐dose irradiation. According to Wolff et al., VMAT even provided a new alternative for single‐fraction SRS irradiation. Also, Gevaert et al. investigated the dosimetric performances of Novalis‐Tx (with dynamic conformal arcs), CyberKnife and GammaKnife for 15 patients with arterious malformation and acoustic neuromas. For DCA, our results were consistent with the Gevaert study: Paddick conformity and homogeneity indexes were, respectively, and (0.68 and 0.34 in our global study). For GammaKnife and CyberKnife, CI were similar to our VMAT techniques (CI = 0.77). Floriano et al. reported a mean result slightly better () with 40 patients treated with CyberKnife. Massager et al. also investigated this parameter with 203 patients treated for a vestibular schwannomas by GammaKnife and obtained , too. Nevertheless, the DCA enhanced healthy brain protection in all cases. Several studies exposed the link between irradiated healthy brain volume and radionecrosis development. Minniti et al. assessed survival and toxicity in 206 patients treated with SRS and showed that, for healthy brain whose and , the risk of radionecrosis was 47%. They concluded for lesions , hypofractionated stereotactic radiotherapy should be considered. For Blonigen et al., this proposition was also considered in case of patients with or to minimize the radionecrosis risk. For patients with multiple lesions, Clark et al. evaluated the feasibility of single‐isocenter versus multi‐isocenter VMAT for multiple intracranial targets. They compared single arc/single isocenter (SASI) with triple arc/single isocenter (TASI) and with triple arc/triple isocenter (TATI), using dosimetric parameters. Single‐isocenter VMAT delivered a conformity equivalent to that of the multiple isocenter VMAT. The Paddick conformity index was, respectively, 0.761 (SASI), 0.699 (TASI), and 0.713 (TATI). Our results are consistent with the Clark study because the mean CI reported was of 0.77 for single‐isocenter VMAT techniques. Lee et al., who compared single isocenter VMAT with the dynamic arc and IMRT plans for multiple cranial tumors, provided slightly better results for single‐isocenter VMAT, with a Paddick CI of 0.83. On the other hand, for healthy brain for VMAT techniques was higher than that of the DCA and generally out of tolerance (12.5 cc). In the literature, the tolerances for healthy brain were relative to a single lesion , and were not appropriate in this case. Instead of comparing global , a per‐lesion average should have been calculated, as was done in the Clark study. Concerning cases with a single lesion close to OAR, better spared the surrounding organ at risk while conserving a high degree of conformity. Similar results were reported by Wolff et al. for intracranial targets and by Lagerwaard et al. in schwannoma vestibular cases. These two studies compared VMAT treatment with conformal arc therapy and concluded that there was a higher conformity with similar sparing of the OAR. An analogous conclusion was drawn by Fogliata et al. in a dosimetric study comparing IMRT, VMAT, and helical tomotherapy for benign intracranial tumors. Abacioglu et al. enhanced conformity index with VMAT and FFF beams and GammaKnife for vestibular schwannoma and cavernous sinus meningioma with bigger volume (respectively, 4.2 and 7.9 cc). They obtained a Paddick CI of 0.84 for GammaKnife and of 0.86 for VMAT. Treatment time is also a parameter to be seriously taken into account because radiosurgery remains a time‐consuming technique. In the case of a single lesion (close or not to OAR), DCA delivered less MU compared with the VMAT irradiations. However, the mean overall time is much longer than or because of the number of couch rotations (3–4 for DCA vs. 0 and 2 for and , respectively). Indeed, therapists had to go into the room for each arc to rotate the couch and the gantry. Consequently, the treatment time was multiplied tenfold. In addition, the more the treatment time, the more the risk intrafractional motion increases, and can become critical in the case of lesion close to OAR. On the issue of the shorter treatment delivery times, Lagerwaard et al. replaced conventional five‐arc radiosurgery by for vestibular schwannomas. This reasoning also applies for the case of multiple lesions. delivered least MU but had more couch rotations than or .

CONCLUSIONS

This study was based on dosimetric indices and did not include the other many aspects of stereotactic treatment, such as the imaging process. For a single lesion, DCA provided better plans considering low doses to healthy brain even if quality indexes were better for the VMAT techniques. In the case of multiple lesions, VMAT plans were more appropriate than the DCA due to the conformity and homogeneity of the dose distribution and the reduction of treatment time (few couch rotation and a single isocenter). Among VMAT techniques, the were the best compromise between treatment time and healthy brain protection. Finally, the performed the highest degree of quality treatment in case of the single lesion close to an organ at risk.
  32 in total

1.  Dynamic arc radiosurgery field shaping: a comparison with static field conformal and noncoplanar circular arcs.

Authors:  T D Solberg; K L Boedeker; R Fogg; M T Selch; A A DeSalles
Journal:  Int J Radiat Oncol Biol Phys       Date:  2001-04-01       Impact factor: 7.038

2.  Relocatable frame for stereotactic external beam radiotherapy.

Authors:  S S Gill; D G Thomas; A P Warrington; M Brada
Journal:  Int J Radiat Oncol Biol Phys       Date:  1991-03       Impact factor: 7.038

3.  A clinical comparison of patient setup and intra-fraction motion using frame-based radiosurgery versus a frameless image-guided radiosurgery system for intracranial lesions.

Authors:  Naren Ramakrishna; Florin Rosca; Scott Friesen; Evrim Tezcanli; Piotr Zygmanszki; Fred Hacker
Journal:  Radiother Oncol       Date:  2010-01-28       Impact factor: 6.280

Review 4.  Physics and dosimetry of the gamma knife.

Authors:  A Wu
Journal:  Neurosurg Clin N Am       Date:  1992-01       Impact factor: 2.509

5.  Treatment accuracy of fractionated stereotactic radiotherapy.

Authors:  Shaleen Kumar; Kevin Burke; Colin Nalder; Paula Jarrett; Cephas Mubata; Roger A'hern; Mandy Humphreys; Margaret Bidmead; Michael Brada
Journal:  Radiother Oncol       Date:  2005-01       Impact factor: 6.280

6.  Setup accuracy of the Novalis ExacTrac 6DOF system for frameless radiosurgery.

Authors:  Thierry Gevaert; Dirk Verellen; Koen Tournel; Nadine Linthout; Samuel Bral; Benedikt Engels; Christine Collen; Tom Depuydt; Michael Duchateau; Truus Reynders; Guy Storme; Mark De Ridder
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-04-07       Impact factor: 7.038

7.  Accuracy of a relocatable stereotactic radiotherapy head frame evaluated by use of a depth helmet.

Authors:  K E Burton; S J Thomas; D Whitney; D S Routsis; R J Benson; N G Burnet
Journal:  Clin Oncol (R Coll Radiol)       Date:  2002-02       Impact factor: 4.126

8.  Intensity modulation with photons for benign intracranial tumours: a planning comparison of volumetric single arc, helical arc and fixed gantry techniques.

Authors:  Antonella Fogliata; Alessandro Clivio; Giorgia Nicolini; Eugenio Vanetti; Luca Cozzi
Journal:  Radiother Oncol       Date:  2008-08-28       Impact factor: 6.280

9.  Irradiated volume as a predictor of brain radionecrosis after linear accelerator stereotactic radiosurgery.

Authors:  Brian J Blonigen; Ryan D Steinmetz; Linda Levin; Michael A Lamba; Ronald E Warnick; John C Breneman
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-09-23       Impact factor: 7.038

10.  Simulation of intrafraction motion and overall geometrical accuracy of a frameless intracranial radiosurgery process.

Authors:  Vladimir Feygelman; Luke Walker; Prakash Chinnaiyan; Kenneth Forster
Journal:  J Appl Clin Med Phys       Date:  2008-10-24       Impact factor: 2.102

View more
  13 in total

1.  Modifying the planning target volume to optimize the dose distribution in dynamic conformal arc therapy for large metastatic brain tumors.

Authors:  Kengo Ogura; Yasuhiro Kosaka; Toshiyuki Imagumbai; Kazuhito Ueki; Ryo Narukami; Takayuki Hattori; Masaki Kokubo
Journal:  Jpn J Radiol       Date:  2017-04-06       Impact factor: 2.374

2.  A patient-specific QA comparison between 2D and 3D diode arrays for single-lesion SRS and SBRT treatments.

Authors:  Yongsook C Lee; Yongbok Kim
Journal:  J Radiosurg SBRT       Date:  2021

3.  Dosimetric comparison of dynamic conformal arc integrated with segment shape optimization and variable dose rate versus volumetric modulated arc therapy for liver SBRT.

Authors:  Deepak Thaper; Rose Kamal; Gaganpreet Singh; Arun S Oinam; Hanuman P Yadav; Rishabh Kumar; Vivek Kumar
Journal:  Rep Pract Oncol Radiother       Date:  2020-05-20

4.  Simultaneous stereotactic radiosurgery of multiple brain metastases using single-isocenter dynamic conformal arc therapy: a prospective monocentric registry trial.

Authors:  Raphael Bodensohn; Anna-Lena Kaempfel; Daniel Felix Fleischmann; Indrawati Hadi; Jan Hofmaier; Sylvia Garny; Michael Reiner; Robert Forbrig; Stefanie Corradini; Niklas Thon; Claus Belka; Maximilian Niyazi
Journal:  Strahlenther Onkol       Date:  2021-04-21       Impact factor: 3.621

5.  Evaluations of the setup discrepancy between BrainLAB 6D ExacTrac and cone-beam computed tomography used with the imaging guidance system Novalis-Tx for intracranial stereotactic radiosurgery.

Authors:  Se An Oh; Jae Won Park; Ji Woon Yea; Sung Kyu Kim
Journal:  PLoS One       Date:  2017-05-19       Impact factor: 3.240

6.  Single isocenter stereotactic radiosurgery for patients with multiple brain metastases: dosimetric comparison of VMAT and a dedicated DCAT planning tool.

Authors:  Jan Hofmaier; Raphael Bodensohn; Sylvia Garny; Indrawati Hadi; Daniel F Fleischmann; Michael Eder; Yavuz Dinc; Michael Reiner; Stefanie Corradini; Katia Parodi; Claus Belka; Maximilian Niyazi
Journal:  Radiat Oncol       Date:  2019-06-11       Impact factor: 3.481

7.  Dosimetric impact of rotational errors on the quality of VMAT-SRS for multiple brain metastases: Comparison between single- and two-isocenter treatment planning techniques.

Authors:  Georgia Prentou; Eleftherios P Pappas; Andreas Logothetis; Efi Koutsouveli; Evaggelos Pantelis; Panagiotis Papagiannis; Pantelis Karaiskos
Journal:  J Appl Clin Med Phys       Date:  2020-02-05       Impact factor: 2.102

8.  A simple knowledge-based tool for stereotactic radiosurgery pre-planning.

Authors:  Daniel S Goldbaum; Justin D Hurley; Russell J Hamilton
Journal:  J Appl Clin Med Phys       Date:  2019-11-19       Impact factor: 2.102

9.  Retrospective quality metrics review of stereotactic radiosurgery plans treating multiple targets using single-isocenter volumetric modulated arc therapy.

Authors:  Yunfeng Cui; Hao Gao; Jiahan Zhang; John P Kirkpatrick; Fang-Fang Yin
Journal:  J Appl Clin Med Phys       Date:  2020-04-02       Impact factor: 2.102

10.  Combination of IFITM1 knockdown and radiotherapy inhibits the growth of oral cancer.

Authors:  Jie Yang; Lei Li; Yan Xi; Ruimei Sun; Hu Wang; Yanxin Ren; Liufang Zhao; Xiaoli Wang; Xiaojiang Li
Journal:  Cancer Sci       Date:  2018-09-21       Impact factor: 6.716

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.