A retrospective analysis of setup and intrafraction positional variation in stereotactic radiotherapy treatments.

Micah Barnes1,2, Adam Yeo1,3, Kenton Thompson4, Claire Phillips5, Tomas Kron1,6,7, Nicholas Hardcastle1,6,7.   

Abstract

PURPOSE: The aim of this study was to provide a comprehensive assessment of patient intrafraction motion in linac-based frameless stereotactic radiosurgery (SRS) and radiotherapy (SRT).
METHODS: A retrospective review was performed on 101 intracranial SRS/SRT patients immobilized with the Klarity stereotactic thermoplastic mask (compatible with the Brainlab frameless stereotactic system) and aligned on a 6 Degree of Freedom (DoF) couch with the Brainlab ExacTrac image guidance system. Both pretreatment and intrafraction correction data are provided as observed by the ExacTrac system. The effects of couch angle and treatment duration on positioning outcomes are also explored.
RESULTS: Initial setup data for patients is shown to vary by up to ±4.18 mm, ±2.97°, but when corrected with a single x-ray image set with ExacTrac, patient positions are corrected to within ±2.11 mm, ±2.27°. Intrafraction patient motion is shown to be uniformly random and independent of both time and couch angle. Patient motion was also limited to within approximately 3 mm, 3° by the thermoplastic mask.
CONCLUSIONS: Our results indicate that since patient intrafraction motion is unrelated to couch rotation and treatment duration, intrafraction patient monitoring in 6 DoF is required to minimize intracranial SRS/SRT margins.
© 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  ExacTrac; IGRT; SRS; SRT; image guidance; intrafraction motion; patient position; thermoplastic mask

Mesh:

Year:  2020        PMID: 33140915      PMCID: PMC7769413          DOI: 10.1002/acm2.13076

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


  21 in total

1.  Image-guided stereotactic radiosurgery for cranial lesions: large margins compensate for reduced image guidance frequency.

Authors:  Harun Badakhshi; David Kaul; Peter Wust; Edzard Wiener; Volker Budach; Volker Buadch; Reinhold Graf
Journal:  Anticancer Res       Date:  2013-10       Impact factor: 2.480

2.  Assessment of patient setup errors in IGRT in combination with a six degrees of freedom couch.

Authors:  Daniel Schmidhalter; Marco Malthaner; Ernst J Born; Alessia Pica; Michael Schmuecking; Daniel M Aebersold; Michael K Fix; Peter Manser
Journal:  Z Med Phys       Date:  2014-01-10       Impact factor: 4.820

3.  Fast intra-fractional image-guidance with 6D positioning correction reduces delivery uncertainty for stereotactic radiosurgery and radiotherapy.

Authors:  Nikolai Tarnavski; Svend Aage Engelholm; Per Munck Af Rosenschold
Journal:  J Radiosurg SBRT       Date:  2016

4.  A clinical comparison of patient setup and intra-fraction motion using frame-based radiosurgery versus a frameless image-guided radiosurgery system for intracranial lesions.

Authors:  Naren Ramakrishna; Florin Rosca; Scott Friesen; Evrim Tezcanli; Piotr Zygmanszki; Fred Hacker
Journal:  Radiother Oncol       Date:  2010-01-28       Impact factor: 6.280

5.  Setup uncertainties in linear accelerator based stereotactic radiosurgery and a derivation of the corresponding setup margin for treatment planning.

Authors:  Mutian Zhang; Qinghui Zhang; Hua Gan; Sicong Li; Su-min Zhou
Journal:  Phys Med       Date:  2016-02-16       Impact factor: 2.685

Review 6.  Stereotactic radiosurgery alone for multiple brain metastases? A review of clinical and technical issues.

Authors:  Arjun Sahgal; Mark Ruschin; Lijun Ma; Wilko Verbakel; David Larson; Paul D Brown
Journal:  Neuro Oncol       Date:  2017-04-01       Impact factor: 12.300

7.  Time dependence of intrafraction patient motion assessed by repeat stereoscopic imaging.

Authors:  Mischa S Hoogeman; Joost J Nuyttens; Peter C Levendag; Ben J M Heijmen
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-11-08       Impact factor: 7.038

8.  Reliability of the bony anatomy in image-guided stereotactic radiotherapy of brain metastases.

Authors:  Matthias Guckenberger; Kurt Baier; Iris Guenther; Anne Richter; Juergen Wilbert; Otto Sauer; Dirk Vordermark; Michael Flentje
Journal:  Int J Radiat Oncol Biol Phys       Date:  2007-09-01       Impact factor: 7.038

9.  Intra-fraction setup variability: IR optical localization vs. X-ray imaging in a hypofractionated patient population.

Authors:  Maria Francesca Spadea; Barbara Tagaste; Marco Riboldi; Eleonora Preve; Daniela Alterio; Gaia Piperno; Cristina Garibaldi; Roberto Orecchia; Antonio Pedotti; Guido Baroni
Journal:  Radiat Oncol       Date:  2011-04-15       Impact factor: 3.481

10.  Frameless linac-based stereotactic radiosurgery (SRS) for brain metastases: analysis of patient repositioning using a mask fixation system and clinical outcomes.

Authors:  Giuseppe Minniti; Claudia Scaringi; Enrico Clarke; Maurizio Valeriani; Mattia Osti; Riccardo Maurizi Enrici
Journal:  Radiat Oncol       Date:  2011-11-16       Impact factor: 3.481

View more
  3 in total

1.  Combined effect of dose gradient and rotational error on prescribed dose coverage for single isocenter multiple brain metastases in frameless stereotactic radiotherapy.

Authors:  Jai-Woong Yoon; Soah Park; Kwang-Ho Cheong; Sei-Kwon Kang; Tae Jin Han
Journal:  Radiat Oncol       Date:  2021-08-31       Impact factor: 3.481

2.  Intrafractional monitoring of patients using four different immobilization mask systems for cranial radiotherapy.

Authors:  Daniel Reitz; Johannes Muecke; Vanessa da Silva Mendes; Guillaume Landry; Michael Reiner; Maximilian Niyazi; Claus Belka; Philipp Freislederer; Stefanie Corradini
Journal:  Phys Imaging Radiat Oncol       Date:  2022-07-25

3.  Intra-fractional motion error during HyperArc stereotactic radiosurgery on patients with brain metastases: Comparison of open and full-face clamshell-style immobilization devices.

Authors:  Shingo Ohira; Riho Komiyama; Naoyuki Kanayama; Yoshihiro Ueda; Shoki Inui; Masayoshi Miyazaki; Masahiko Koizumi; Koji Konishi
Journal:  J Appl Clin Med Phys       Date:  2022-01-20       Impact factor: 2.102