Cory Knill1, Raminder Sandhu1, Robert Halford1, Michael Snyder1, Zachary Seymour1.
Abstract
PURPOSE: Brainlab's Elements Multiple Brain Mets SRS (MBMS) is a dedicated treatment planning system for single-isocenter multi-target (SIMT) cranial stereotactic radiosurgery (SRS) treatments. The purpose of this study is to present the commissioning experience of MBMS on an Elekta Versa HD.Entities:
Keywords: Elements Multiple Brain Mets SRS; SRS; Versa HD
Year: 2021 PMID: 33756044 PMCID: PMC8035552 DOI: 10.1002/acm2.13223
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1Comparison between commissioning and calculated scatter factors at 100 cm source‐to‐phantom distance and 10 cm depth. All scatter factors were normalized to unity for a 10 cm × 10 cm field.
Measured scatter factors at 100 cm SSD and 10 cm depth. Field size is nominal setting on the VersaHD. Bracketed numbers show the 2*sigma uncertainty in the last digit.
| Field Size (cm) | 6 X | 6 FFF | 10 FFF |
|---|---|---|---|
| 1.0 | 0.677 (3) | 0.696 (3) | 0.680 (4) |
| 2.0 | 0.806 (2) | 0.830 (3) | 0.843 (4) |
| 3.0 | 0.847 (3) | 0.872 (2) | 0.894 (3) |
| 4.0 | 0.879 (1) | 0.901 (4) | 0.923 (1) |
| 6.0 | 0.928 (2) | 0.945 (2) | 0.958 (2) |
| 8.0 | 0.970 (2) | 0.977 (2) | 0.982 (4) |
| 10.0 | 1.000 (0) | 1.000 (0) | 1.000 (0) |
Fig. 2Comparison between commissioning and calculated 6 FFF percent depth doses (PDDs) at 100 cm source‐to‐phantom distance for selected field sizes.
Fig. 3Comparison between commissioning and calculated 6 FFF diagonal profiles at 100 cm source‐to‐phantom distance for select depths.
Fig. 4Difference between the measured and calculated 6 FFF collimator penumbras at 100 cm source‐to‐phantom distance, 10 cm depth for a 3 cm × 3 cm field size. The discrepancy is smaller in the jaw direction (a) compared with the MLC direction (b).
Measured multi‐leaf collimator (MLC) transmission and dynamic leaf shifts.
| 6 X | 6 FFF | 10 FFF | |
|---|---|---|---|
| Transmission | 0.4% | 0.4% | 0.3% |
| Dynamic leaf shift | 0.15 | 0.22 | 0.26 |
Fig. 5Dose metrics for single‐isocenter multi‐target plans of different target sizes created with MBMS.
Fig. 6Axial film results of the end‐to‐end test. (a) Isosdose overlay between Elements‐predicted and film‐measured dose. (b) Line profile between two targets (green line in (a)).
Fig. 7Plan comparisons between: (a) MBMS v1.5 and (b) MBMS v2.0. The updated algorithm removes the 12 Gy dose‐bridging.