Song Duan1, Yurong Zhu2, Feng Liu3, Sherman Xuegang Xin4. 1. Department of Radiation Oncology, Sun Yat-sen Memorial Hospital, Sun Yat-sen University. 2. Department of Biomedical Engineering, Southern Medical University. 3. School of Information Technology and Electrical Engineering, University of Queensland. 4. School of Medicine, South China University of Technology, Guangzhou Higher Education Mega Centre.
Abstract
PURPOSE: Magnetic resonance electrical property tomography (MR EPT) is a technique for non-invasively obtaining the electric property (EP) distribution of biological tissues, with a promising potential for application in the early detection of tumors. However, the contrast capability (CC) of this technique has not been fully studied. This work aims to theoretically explore the CC for detecting the variation of EP values and the size of the imaging region. METHODS: A simulation scheme was specifically designed to evaluate the CC of MR EPT. The simulation study has the advantage that the magnetic field can be accurately obtained. EP maps of the designed phantom embedded with target regions of designated various sizes and EPs were reconstructed using the homogeneous Helmholtz equation based on B1+ with different signal-to-noise ratios (SNRs). The CC was estimated by determining the smallest detectable EP contrast when the target region size was as large as the Laplacian kernel and the smallest detectable target region size when the EP contrast was the same as the difference between healthy and malignant tissues in the brain, based on the reconstructed EP maps. RESULTS: Using noise free B1+, the smallest detectable contrastσ and contrastεr were 1% and 3%, respectively, and the smallest detectable target region size was 1 mesh unit (the base unit size used in the simulation) for conductivity and relative permittivity. The smallest detectable EP contrast and target region size were decreased as the B1+ SNR increased. CONCLUSION: The CC of MR EPT was related with the SNR of the magnetic field. A small EP contrast and size were necessary for detection at a high-SNR magnetic field. Obtaining a high-SNR magnetic field is important for improving the CC of MR EPT.
PURPOSE: Magnetic resonance electrical property tomography (MR EPT) is a technique for non-invasively obtaining the electric property (EP) distribution of biological tissues, with a promising potential for application in the early detection of tumors. However, the contrast capability (CC) of this technique has not been fully studied. This work aims to theoretically explore the CC for detecting the variation of EP values and the size of the imaging region. METHODS: A simulation scheme was specifically designed to evaluate the CC of MR EPT. The simulation study has the advantage that the magnetic field can be accurately obtained. EP maps of the designed phantom embedded with target regions of designated various sizes and EPs were reconstructed using the homogeneous Helmholtz equation based on B1+ with different signal-to-noise ratios (SNRs). The CC was estimated by determining the smallest detectable EP contrast when the target region size was as large as the Laplacian kernel and the smallest detectable target region size when the EP contrast was the same as the difference between healthy and malignant tissues in the brain, based on the reconstructed EP maps. RESULTS: Using noise free B1+, the smallest detectable contrastσ and contrastεr were 1% and 3%, respectively, and the smallest detectable target region size was 1 mesh unit (the base unit size used in the simulation) for conductivity and relative permittivity. The smallest detectable EP contrast and target region size were decreased as the B1+ SNR increased. CONCLUSION: The CC of MR EPT was related with the SNR of the magnetic field. A small EP contrast and size were necessary for detection at a high-SNR magnetic field. Obtaining a high-SNR magnetic field is important for improving the CC of MR EPT.
Entities:
Keywords:
contrast capability; electromagnetic simulation; magnetic resonance electrical property tomography; signal-to-noise ratio
Electrical properties (EP) of biological tissues, including conductivity () and relative permittivity (ε), have been the subject of research for several years.[1-7] Numerous studies have been conducted on the measurement of the EP values of various malignant and healthy tissue types at radiofrequency (RF) and microwave frequency.[1-5,8,9] These investigations have consistently demonstrated that the EP values of malignant tissues have definite levels of difference compared with those of healthy tissues. Numerous factors, including relative intracellular and extracellular fluid volumes, ionic concentrations, and cellular membrane permeability, account for the difference mechanism of EP values between malignant and healthy tissues.[1,2,10] Considerable differences in EP values of even up to 577% may exist between malignant and healthy tissues.[7] Considering the biological continuous process of the alteration in EP values of tissues, we can assume that only a small change of EP values, for example, only 1%, occurs at the early stage of the progressive process of tissues developed from healthy to malignant status.[11] Given the large contrast gap of EP values between healthy and malignant tissues, people believe that detecting the variations in EP values of tissues at the beginning of the entire progressive process from healthy to malignant status may be beneficial for the early detection of tumors.[12-14]Although the knowledge on EP values of tissues can be easily obtained in vitro, for example, using the open-ended coaxial probe method,[10,15] non-invasively obtaining the tomographic EP data in vivo is difficult until the technique magnetic resonance electrical property tomography (MR EPT) was proposed. On the basis of the measurement of the transmitted field of the MRI system at the Larmor frequency,[16-20] MR EPT can image the EP distribution inside the human body without requiring a current injection, as implemented in electric impedance tomography (EIT)[21] and magnetic resonance EIT.[22,23] The measurement mechanism is in the field being electromagnetically distorted by the loaded human body. Therefore, the quantitative assessment of the distorted allows the inverse reconstruction of the EP values of tissues. The concept of EP imaging was first introduced by Haacke et al.[24] and was applied practically by Wen et al.[25]. Wen et al.[25] indicated that the distribution of the RF field in high-field MRI directly correlates with the EP distributions in the sample and that the relationship between the RF field and EP can be explained by a simplified EPT reconstruction equation, that is, the homogeneous Helmholtz equation. Since its proposal by Wen et al.,[25] this simplified EPT reconstruction equation has become the most popular algorithm applied in the majority of EPT studies.[14,26-33] In recent years, numerous studies have focused on the capability of the MR EPT technique. For example, in prior work, the validity of homogeneous Helmholtz equation was investigated by Seo et al.,[33] demonstrating that a large reconstruction error occurred in inhomogeneous objects in which the EP homogenous assumption in the homogeneous Helmholtz equation fails. An alternative study conducted by Lee et al.[34] discussed the limit of signal-to-noise ratio (SNR) using a homogenous phantom, and it showed that the SNR of reconstructed EP is proportional to the square of the linear dimension of the ROI over which the EPs are determined and to the square root of the number of voxels in the ROI. Duan et al.[14] quantitatively analyzed the reconstruction errors of Helmholtz-based MR EPT at the interfaces of adjacent tissues, showing the limitation of this homogeneous Helmholtz equation when applied to the human body. Mandija et al.[35] indicated that numerical error is a major cause of limited accuracy in Helmholtz-based MR-EPT reconstructions. These studies have shown the various limitations of Helmholtz-based MR EPT in EP reconstruction, especially for inhomogeneous objects. Such limitations can lead to problems when the MR EPT algorithm is applied in vivo. For example, when the MR EPT is applied for the early detection of tumors that have different EP values with surrounding tissues, the detection of lesions may be affected due to the limitations of this technique. This question highlights the importance of the research on the contrast capability (CC) of MR EPT, i.e., the detection capability on the variation in EP values of tissues and the size of the image region. Wang et al.[36] evaluated the sensitivity for structures of various EP values and geometrical sizes at 7T for a boundary-informed electrical property tomography technique that can reliably retrieve EP distributions inside an object using an external EP reference area. However, to our knowledge, the CC of MR EPT remains unexplored.This study aims to investigate the CC of MR EPT using the homogeneous Helmholtz equation. Hence, the reconstruction of MR EPT is required, and obtaining the data is the first step. However, in the practical MR scan, the data inevitably involves engineering artifacts, such as Gibbs ringing that occurs when the spatial resolution is insufficient to represent small structures in the object[37] and the distortion or blurring caused by resonance offsets and hardware limitations[38]; in addition, the presently available B1 mapping methods in engineering cannot obtain the phase map of the field.[39-41] In practice, a rough approximation of transmit phase was obtained by the transceive phase assumption using a quadrature volume coil.[27,39,40,42] On the contrary, accurately obtaining the full set of data using numerical electromagnetic simulation is easy. Therefore, in this work, the CC of MR EPT was theoretically investigated via numerical simulation. A carefully designed simulation scheme was utilized to obtain the data of a designed dielectric model embedded with a series of target regions with various sizes and EP values. The EP distributions were reconstructed using the homogeneous Helmholtz equation based on the obtained with different SNR. The smallest detectable EP contrast (when the target region size was as large as the Laplacian kernel) and smallest detectable target region size (when the EP contrast was the same as the difference between healthy and malignant tissues in the brain) were determined following the reconstructed EP maps and used to describe the CC of MR EPT.
Materials and Methods
Modeling
In this study, a numerical simulation was carefully designed to obtain field distribution as the base for analyzing the CC of MR EPT. Simulations were performed on a commercial simulation software (SEMCAD X. 14.6, Schmid & Partner Engineering AG, Zurich, Switzerland), which considered a transmit coil loaded with a dielectric model, as shown in Fig. 1.
Fig. 1
Photographs of the high-pass birdcage coil loaded with the dielectric cube model that embedded a cubic target region in its center. The background cube (grey), target region (red), current source (green), and capacitor (blue) are shown using black arrows. The co-ordinate system at the bottom left shows the x-, y-, and z-directions.
The transmit coil was a 16-leg, 3T high-pass birdcage coil. In the coil setup, the coil leg length was 40 cm and the diameter of the end-ring was 30 cm. A total of 16 capacitors were distributed on each end-ring with equal space having a uniform capacitance of 16.22 pF. Each leg had a unit amplitude current source at the center, and all sources had the same current direction. From leg to leg, the phases of the current sources were assigned with a step increment of 2π/16.A dielectric cube (referred to as the background cube) was constructed in the SEMCAD, which was considered for the side length of 25 mesh units (1 mesh unit was equal to the side length of a given isotropic voxel size). A small cubic target region was set in the center of the background cube, as shown in Fig. 1. The EP of background cube was the same as that of healthy white matter (σ = 0.3421 S/m and ε = 52.53).[1-3] To investigate the detection capability of MR EPT on the variation in EP values of tissues and the size of the image region, different EPs and sizes were assigned to the target region. The details of the simulation settings of the target region are as follows.In the investigation of the detection capability on the variation in EP values of tissues, various EP values were assigned to the target region, thereby forming the following pairs of (contrast, contrast): (0.01:0.02:1.01, 0) and (0, 0.01:0.02:0.27), where contrast and contrast are the contrasts of conductivity and relative permittivity, respectively, and are calculated as Eqs. [1] and [2]:
where σTR and ε represent the conductivity and relative permittivity of the target region and σB and ε denote the conductivity and relative permittivity of the background cube, respectively. The detection capability on the variation of conductivity and relative permittivity was studied separately. When the detection capability on the variation of conductivity was investigated, contrast was increased from 0.01 to 1.01, and contrast was fixed at 0, i.e., the target region has the same ε as the background cube. Similarly, when the detection capability on the variation of relative permittivity was investigated, contrast was increased from 0.01 to 0.27, and contrast was fixed at 0. The smallest investigated contrastEP was set as 1%, which is small compared with the large difference (even up to 577%) in EP values between the normal and malignant tissues.[7,43] The contrastEP smaller than 1% was not studied here. The maximum contrastEP investigated was equal to the contrast between the humanglioma and healthy white matter.[1-3,43] The side length of the target region was set as 7 mesh units; thus, the target region was as large as the Laplacian kernel used in this study, which is the smallest size of image region meeting the assumption “locally constant EP” in the homogenous Helmholtz equation.In the investigation of the detection capability on the size of the image region, the target region was constructed with a side length of 1–10 mesh units and EP values equal to humanglioma (σ = 0.6836 S/m, ε
= 66.18).[43]
Generation of magnetic field
The built model was positioned at the center of the birdcage coil. The voxel size in the region of the phantom was set as 1 × 1 × 1 mm3, which was a recommended voxel size for MR image acquisition, considering scan time and image quality.[44] After the SEMCAD simulation, data of the model were exported into MATLAB (MathWorks, Natick, MA, USA) for further processing. To observe the changes in magnitude and phase in the target region, the magnitude and phase maps were normalized by the magnitude and phase maps generated using a homogenous model in which the EP of target region was equal to background cube, i.e., the EP values of the background cube and target region were equal to the EP of the healthy white matter. The normalized magnitude and phase maps across the center of the target region were presented. In practical measurement, the data of magnetic fields are consistently polluted by noise. In this manuscript, the measurement noise was included by adding complex Gaussian noise with zero mean to the field, thereby forming several SNR levels of magnitude in the range of 60 to +∞. The SNR was defined as , where is the simulated data and N denotes the Gaussian noise added in the simulated .
Reconstruction of EP
After the of the model was obtained, the EP distributions of the dielectric model were reconstructed using the following homogeneous Helmholtz equation [3][25]:
where ω is angular frequency and represents the complex permittivity and is assumed to be isotropic. μ denotes the magnetic permeability. In the human body, μ was assumed to be constant and equal to the magnetic permeability of the free space μ0 = 4π × 10−7. By separating the real and imaginary components of Eq. [3], conductivity σ and relative permittivity ε can be computed as
where ε0 = 8.8542 × 10−12 F/M refers to the free space permittivity. The numerical Laplacian calculation in Eqs. [4] and [5]
was realized using the Savitzky–Golay kernel (KSG).[45,46] The applied KSG was as following Eqs. [6] and [7]:
where the index i indicates the third dimension of the small (5 × 3 × 3, Kernelsmall) and large (7 × 3 × 3, Kernellarge) kernels. These two kernels were used to calculate the second-order derivatives of the target voxel in the x, y, and z-directions; summing the results of the three derivatives results in the Laplacian kernel. The 7 × 3 × 3 and 5 × 3 × 3 kernels were used for in- and inter-plane computations, respectively. In the inter-plane direction, a smaller kernel was selected to reduce the number of slices required for reconstruction, thereby decreasing the measurement time.
Evaluation of CC
The CC evaluation was based on the reconstructed EP maps. The designated target region is detectable if the mean EP of the target region (), the mean EP of the background (), and the standard deviation of the EP of the background (STDB,EP) satisfied the following relation [8]:The smallest detectable contrastEP (when the target region was as large as the Laplacian kernel) and smallest detectable target region size (when the EP of target region was equal to glioma) were determined and used to describe the CC.
Results
Detection capability on the variation of EP
The investigation of the detection capability on the variation of EP was conducted using a target region with fixed size and varied EP. The size of the target region was approximately the size of the Laplacian kernel used for the EP reconstruction. The data of the built model were generated using the above-described numerical simulation procedure. Fig. 2 shows the transverse distributions of normalized magnitude and phase across the central area of the target region, in which Fig. 2a illustrates the normalized magnitude maps with respect to the contrast 1–27% (corresponding to the first to the last column, respectively). Fig. 2b presents the normalized phase maps with respect to the contrast 1 to 101%, The normalized magnitude and phase in the target region were increased with contrast and contrast.
Fig. 2
Transverse distributions of normalized magnitude and phase across the center of the target region for different contrastEP values. The size of the target region was approximately the size of the Laplacian kernel used for the electric property (EP) reconstruction. (a) Normalized magnitude maps in which contrast was increased from 1% to 27%, corresponding to the first to the last column. (b) Normalized phase maps in which contrast was increased from 1 to 101%.
Figure 3 displays the reconstructed EP distributions of the built model across the center of the target region for the contrast from 1 to 101%, and the contrast from 1 to 27%, when the SNR was 70, 85, and +∞ (noise free). The conductivity and relative permittivity over the target region were increased with contrast and contrast, respectively. With the decrease in SNR, the target region disappeared in the reconstructed EP maps, especially for the target regions with small contrastEP.
Fig. 3
Reconstructed electric property (EP) distributions of the constructed model across the center of the target region. (a) Reconstructed relative permittivity maps. From the first to the last columns, contrast was from 1% to 27%. From the first to the third rows, the SNRs were 70, 85, and +∞ (noise free), respectively. (b) Reconstructed conductivity maps. From the first to the last columns, the contrast values were 1–101%. From the first to the third rows, the SNRs were 70, 85, and +∞, respectively. SNR, signal-to-noise ratio.
Figure 4 plots the smallest detectable contrastEP for different SNRs. When the noise free data were used, the smallest detectable contrast and contrast were as small as 1% and 3%, respectively. The smallest detectable contrastEP was decreased as the SNR of data was increased. The target regions cannot be detected in the investigated range of EP contrast when the SNR smaller than 65 and 75, respectively.
Fig. 4
Smallest detectable contrast (triangle) and contrast (circle) for different SNRs. SNR = +∞ indicates that no gauss noise was added to the simulated data. The smallest detectable contrastEP was decreased as the SNR of data was increased. The target regions cannot be detected in the investigated range of electric property (EP) contrast on the basis of the reconstructed conductivity and relative permittivity maps, when the SNRs were smaller than 65 and 75, respectively. SNR, signal-to-noise ratio.
Detection capability on the size of the image region
The detection capability on the size of the image region was investigated using a target region with fixed EP and varied size. The assigned EP values of target region were the same as the EP of humanglioma. Figure 5 presents the transverse distributions of normalized magnitude and phase across the central area of the target region for different target region sizes (side lengths from 1 to 10 mesh units). The normalized magnitude and phase in the target region were increased with the target region size.
Fig. 5
Transverse distributions of normalized magnitude (a) and phase (b) across the central area of the target region for different target region sizes. The electric property (EP) values of the target region and background were the same as those for human glioma and healthy grey matter, respectively. From the first to the last columns, the side length of the target region was increased from 1 to 10 mesh units.
Figure 6 presents the reconstructed EP distributions of the built model across the center of the target region for a size of 1–10 mesh units, when the SNR was 70, 85, and +∞ (noise free). The reconstructed EP over the target region is decreased as the target region size decreases. With the decrease in SNR, the target region disappeared in the reconstructed EP maps, especially for the target regions with small size.
Fig. 6
Reconstructed relative permittivity (a) and conductivity (b) distributions of the constructed model across the center of the target region. From the first to the last columns, the side lengths of the target region were 1–10 mesh units, respectively. From the first to the third rows in (a) and (b), the SNRs were 70, 85, and +∞ (noise free), respectively. SNR, signal-to-noise ratio.
Figure 7 plots the smallest detectable target region size for different SNRs. When the noise free data were used, the smallest detectable target region size was 1 mesh unit for conductivity and relative permittivity. The smallest detectable target region size decreased as the SNR of was increased. The target regions cannot be detected in the investigated range of target region size when the SNRs were smaller than 65 and 70, respectively.
Fig. 7
Smallest detectable target region size determined on the basis of reconstructed relative permittivity (circle) and conductivity (triangle) maps, for different SNRs. The smallest detectable target region size is decreased as the SNR of data was increased. The target regions cannot be detected in the investigated range of target region size on the basis of the reconstructed conductivity and relative permittivity maps when the SNRs were smaller than 65 and 70, respectively. SNR, signal-to-noise ratio.
Discussion
Given the existence of a large difference in EPs between benign and malignant tissues, the technique of MR EPT may be valuable for the early clinical detection of tumors. The current manuscript focused on the CC of MR EPT, which is meaningful for the application of this technique. The CC of MR EPT was determined at a frequency of 128 MHz on the basis of the homogeneous Helmholtz equation.Two central challenges of EPT are the boundary issue, in which severe reconstruction error of EP can occur at the interfaces of adjacent tissues with different EP,[14,33] and the transmit phase, which is difficult to determine exactly.[40] Certain strategies have been proposed to resolve such challenges.[29,47-50] Some of the newly proposed EPT reconstruction methods involve the dual-excitation algorithm,[48] convection–reaction equation-based MR EPT,[51] gradient-based EPT,[52] contrast source inversion-EPT,[53] and water content-based EPT.[12] These approaches were implemented by rewriting the Helmholtz equation combined with some approximations or prior knowledge. In comparison with the abovementioned approaches, the homogeneous Helmholtz equation is a basic reconstruction formulation derived directly from the Helmholtz equation. Therefore, using the basic reconstruction formulation to indicate the basic characteristics of the CC of MR EPT is reasonable.A carefully designed numerical simulation strategy was used to obtain . The different changes in magnitude and phase of were observed for different target regions. The changes were related to the alterations of the EP and size of the target region. As shown in Figs. 2 and 5, the normalized magnitude and phase over the target regions increased with the increase of the geometric volume of the target region and contrastEP. These results indicated that the influence of the target region on the magnetic field was increased with the increase in the size and EP of the target region. Thus, a large difference in reconstructed EP was observed between the target region and background cube for these cases with lager contrastEP and target region size, as shown in Figs. 3 and 6.Determining the smallest detectable contrastEP at different target region sizes and the smallest detectable target region sizes at different contrastEP values is complicated because of the numerous cases of size and contrastEP in the target region. In this study, the CC of MR EPT was evaluated by determining the smallest detectable contrastEP (when the target region was as large as the Laplacian kernel) and smallest detectable target region size (when the EP of target region was equal to glioma), based on the reconstructed EP maps. When the noise free data were used, the smallest detectable contrastEP was 1% for conductivity and 3% for relative permittivity, and the smallest detectable target region size was 1 mesh unit for conductivity and relative permittivity. The smallest detectable contrastEP and smallest detectable target region size were increased as the SNR declined, as shown in Figs. 4 and 7. This trend can also be observed in Figs. 3 and 6, where the target regions with small contrastEP and size disappeared when the SNR was lower. Therefore, small EP contrast and size were necessary for detecting the imaging region at high SNRs of the magnetic field. Obtaining the high-SNR magnetic field data is important for improving the CC of MR EPT.Image noise is an undesirable component correlated with the desired signal, which is consistently contained in practical measurement.[54] In this study, the influence of image noise on the CC of MR EPT was considered by adding complex Gaussian noise in the simulated data. The target regions with small size and contrastEP are sensitive to image noise. These target regions were undetectable even when small noise was added. In the MR scanning, the image noise is related to numerous parameters,[55] such as physical and instrumental parameters including the strength of the main field B0, the random electrical fluctuations associated with the resistances of the receiver coil and the body, and the imaging sequence parameters, e.g., the excitation tip angles, sequence timing, and data acquisition time. During the implementation of the MR EPT, these parameters will affect the SNR of the data and finally affect the CC of the technique.Previous studies[35,42] indicate that a small Laplacian kernel leads to spatially limited numerical boundary errors, which will be ideal for convoluted tissue structures, such as the human brain. However, in reality, the acquired images are intrinsically affected by local fluctuations, such as those from thermal noise. Small kernels are highly sensitive to local signal noise. To minimize the effect of spatial fluctuations, relatively large Laplacian kernel in combination with Gaussian apodization, such as the van Lier’s kernel[28] and KSG,[45,46] was often adopted. Furthermore, Lee et al.[34] showed that the KSG is optimal for noise reduction. Therefore, the KSG was selected in the current study to reconstruct the EP distribution of the model.This manuscript only focused on the frequency of 128 MHz, which is the most widely available frequency. The CC of MR EPT at other frequencies should be different from the results presented in this study because the SNR of the obtained magnetic field and the local field curvature (positively correlated to tissue EPs) varied with the applied frequency, leading to different noise levels in the reconstructed EP.
Conclusion
The CC of the MR EPT technique was related to the SNR of magnetic fields. Small EP contrast and size were necessary for detection at high SNR of the magnetic field. Obtaining high-SNR magnetic field data is important for improving the CC of MR EPT.