Jae Hyuck Lee1, Yong Woo Lee1, Jong Soo Lee2, Michael C Knorz3, Gerd U Auffarth4, Chul Young Choi1,4.
Abstract
PURPOSE: To compare measurements of anterior and posterior keratometric values, using a color light-emitting diode corneal topographer and a dual rotating Scheimpflug-Placido topographer.Entities:
Year: 2018 PMID: 30687546 PMCID: PMC6327256 DOI: 10.1155/2018/6764805
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Characteristics and measurement settings for each system.
| Parameter | Dual rotating Scheimpflug–Placido | Color light-emitting diode |
|---|---|---|
| Acquisition time (s) | 0.75 | 2 |
|
| 1.3375 | 1.3375 |
|
| 1–4 (anterior) | 3 |
| 0.5–2 (posterior) |
K, keratometry. Configurable or selectable option.
Demographics and refractive error by group.
| Parameter | Refractive group ( | Normal group ( |
|
|---|---|---|---|
| Mean age (y) ± SD | 32.30 ± 8.05 | 32.78 ± 6.98 | 0.684 |
| Sex, | 0.133† | ||
|
| 24 (40.0) | 32 (53.3) | |
|
| 36 (60.0) | 28 (46.7) | |
| Mean SE ( | −0.42 ± 0.46 (−1.75–0.50) | −2.28 ± 2.25 (−5.50–0.75) | <0.001∗ |
| Surgery, | |||
|
| 48 (80.0) | ─ | |
|
| 12 (20.0) | ─ |
LASEK, laser-assisted subepithelial keratectomy; LASIK, laser in situ keratomileusis; SE, spherical equivalent. Independent sample t-test; †chi-square test.
Comparison of corneal index measurements using dual rotating Scheimpflug–Placido and color light-emitting diode corneal topographers.
| Parameter | DRSP (mean ± SD) | Color LED (mean ± SD) | Difference |
| ICC | |
|---|---|---|---|---|---|---|
| Mean ± SD | 95% CI | |||||
| Postrefractive group | ||||||
|
| ||||||
| sim | 38.78 ± 1.80 | 38.71 ± 1.88 | −0.08 ± 0.44 | −0.19, 0.04 | 0.186 | 0.986 |
| Astigmatism magnitude | 0.83 ± 0.47 | 0.84 ± 0.53 | 0.01 ± 0.37 | −0.09, 0.10 | 0.904 | 0.938 |
| | −0.32 ± 0.28 | −0.29 ± 0.32 | 0.03 ± 0.20 | −0.02, 0.08 | 0.267 | 0.925 |
| | 0.06 ± 0.22 | −0.04 ± 0.25 | −0.10 ± 0.15 | −0.14, −0.06 | 0.000 | 0.905 |
|
| ||||||
| sim | −6.39 ± 0.16 | −5.97 ± 0.32 | 0.42 ± 0.32 | 0.34, 0.51 | 0.000 | 0.327 |
| Astigmatism magnitude | −0.36 ± 0.11 | −0.39 ± 0.21 | −0.04 ± 0.23 | −0.09, 0.02 | 0.216 | 0.183 |
| | 0.17 ± 0.06 | 0.08 ± 0.16 | −0.09 ± 0.16 | −0.13, −0.05 | 0.000 | 0.016 |
| | −0.02 ± 0.15 | −0.01 ± 0.06 | −0.01 ± 0.16 | −0.05, 0.03 | 0.655 | 0.000 |
| Normal group | ||||||
|
| ||||||
| sim | 43.28 ± 1.37 | 43.40 ± 1.38 | 0.12 ± 0.26 | 0.06, 0.19 | 0.001 | 0.991 |
| Astigmatism magnitude | 1.34 ± 0.83 | 1.41 ± 0.88 | 0.07 ± 0.35 | −0.02, 0.15 | 0.133 | 0.956 |
| | −0.61 ± 0.44 | −0.64 ± 0.47 | −0.03 ± 0.17 | −0.08, 0.01 | 0.135 | 0.963 |
| | 0.01 ± 0.25 | 0.00 ± 0.24 | −0.01 ± 0.14 | −0.04, 0.03 | 0.740 | 0.916 |
|
| ||||||
| sim | −6.43 ± 0.27 | −6.26 ± 0.25 | 0.16 ± 0.12 | 0.13, 0.19 | 0.000 | 0.944 |
| Astigmatism magnitude | −0.37 ± 0.11 | −0.41 ± 0.17 | −0.04 ± 0.16 | −0.08, −0.00 | 0.043 | 0.499 |
| | 0.17 ± 0.07 | 0.18 ± 0.09 | 0.01 ± 0.09 | −0.01, 0.04 | 0.274 | 0.469 |
| | 0.00 ± 0.06 | 0.00 ± 0.09 | −0.00 ± 0.10 | −0.03, 0.03 | 0.977 | 0.175 |
CI, confidence interval; color LED, color light-emitting diode corneal topographer; DRSP, dual rotating Scheimpflug–Placido corneal topographer; ICC, intraclass correlation coefficient; K, keratometry. Paired sample t-test.
Comparison of corneal index measurements using dual rotating Scheimpflug–Placido and color light-emitting diode corneal topographers in total group.
| Parameter | Total group | |||||
|---|---|---|---|---|---|---|
| DRSP (mean ± SD) | Color LED (mean ± SD) | Difference |
| ICC | ||
| Mean ± SD | 95% CI | |||||
|
| ||||||
| sim | 41.07 ± 2.76 | 41.09 ± 2.87 | 0.03 ± 0.37 | -0.04, 0.09 | 0.451 | 0.996 |
| Astigmatism magnitude | 1.09 ± 0.72 | 1.13 ± 0.78 | 0.04 ± 0.36 | −0.03, 0.10 | 0.256 | 0.938 |
| | −0.46 ± 0.39 | −0.47 ± 0.44 | −0.00 ± 0.19 | −0.04, 0.03 | 0.874 | 0.947 |
| | 0.04 ± 0.24 | −0.02 ± 0.25 | −0.05 ± 0.15 | −0.08, 0.03 | 0.000 | 0.906 |
|
| ||||||
| sim | −6.41 ± 0.22 | −6.12 ± 0.32 | 0.29 ± 0.27 | 0.24, 0.34 | 0.000 | 0.681 |
| Astigmatism magnitude | −0.36 ± 0.11 | −0.40 ± 0.19 | −0.04 ± 0.20 | −0.07, −0.00 | 0.027 | 0.334 |
| | 0.17 ± 0.06 | 0.13 ± 0.13 | −0.04 ± 0.14 | −0.06, −0.01 | 0.004 | 0.180 |
| | −0.01 ± 0.06 | −0.01 ± 0.12 | −0.00 ± 0.13 | −0.03, 0.02 | 0.695 | 0.000 |
CI, confidence interval; color LED, color light-emitting diode corneal topographer; DRSP, dual rotating Scheimpflug–Placido corneal topographer; ICC, intraclass correlation coefficient; K, keratometry. Paired sample t-test.
Figure 1Bland–Altman plots showing the agreement between the posterior corneal index measurements from the 2 systems. The bold lines represent the mean difference between the 2 methods. The broken lines represent the 95% limits of agreement. (a) Normal group posterior simK, (b) refractive group posterior simK, (c) normal group posterior astigmatism magnitude, (d) refractive group posterior astigmatism magnitude, (e) normal group posterior J0, (f) refractive group posterior J0, (g) normal group posterior J45, and (h) refractive group posterior J45 (color LED, color light-emitting diode corneal topographer; DRSP, dual rotating Scheimpflug–Placido corneal topographer).