Tiago B Ferreira1, Filomena J Ribeiro1.
Abstract
PURPOSE: To assess the comparability and repeatability of keratometric and astigmatism values measured by four techniques: Orbscan IIz® (Bausch and Lomb), Lenstar LS 900® (Haag-Streit), Cassini® (i-Optics), and Total Cassini (anterior + posterior surface), in healthy volunteers. PATIENTS AND METHODS: Fifteen healthy volunteers (30 eyes) were assessed by the four techniques. In each eye, three consecutive measures were performed by the same operator. Keratometric and astigmatism values were recorded. The intraclass correlation coefficient (ICC) was used to assess comparability and repeatability. Agreement between measurement techniques was evaluated with Bland-Altman plots.Entities:
Keywords: astigmatism; comparability; keratometry; repeatability; topography
Year: 2017 PMID: 29339918 PMCID: PMC5745151 DOI: 10.2147/OPTH.S146730
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Descriptive statistics
| Parameter | Total Cassini | Cassini® | Orbscan IIz® | Lenstar LS 900® |
|---|---|---|---|---|
| K1 (D) | 42.77 (36.93–46.06) | 42.90 | 42.82 | |
| K2 (D) | 44.31 (41.24–47.06) | 44.17 | 44.06 | |
| Astigmatism (D) | 0.88 (0.44–2.04) | 0.89 (0.25–5.80) | 0.80 (0.27–2.17) | 0.87 (0.34–3.04) |
| Axis (°) | 89.33 (11.33–147.50) | 91.92 (10.67–167.00) | 93.00 (16.00–167.00) | 94.33 (12.67–163.00) |
| J0 (D) | 0.23 | 0.30 | 0.33 (−0.78 to 2.82) | 0.32 (−0.82 to 1.48) |
| J45 (D) | −0.01 (−0.76 to 0.36) | 0.03 | 0.04 (−0.54 to 1.63) | 0.05 |
Notes: All values are presented as median (minimum–maximum). Orbscan showed higher values of K1 and K2 compared to Lenstar. Cassini showed a higher value of J0 compared to Total Cassini and a lower value of J45 compared to Lenstar.
P<0.005.
Abbreviations: K1, minimum keratometry; K2, maximum keratometry.
Figure 1Bland–Altman plots for astigmatism axis, astigmatism magnitude, and Km. The limits of agreement are shown by the red line. For astigmatism axis, the best agreement was between Total Cassini and Lenstar LS 900® and the worst was between Total Cassini and Cassini®. For astigmatism magnitude, the best agreement was between Total Cassini and Cassini, and the worst was between Cassini and Lenstar. Regarding Km, the best agreement was between Cassini and Lenstar, and the worst was between Orbscan IIz® and Lenstar.
Abbreviation: Km, mean keratometry.
Repeatability and comparability
| Parameter | Intra-keratometer (ICC, 95% CI)
| Inter-keratometer (ICC, 95% CI)
| ||||
|---|---|---|---|---|---|---|
| Cassini® | Orbscan IIz® | Lenstar LS 900® | Cassini–Orbscan | Cassini–Lenstar | Orbscan–Lenstar | |
| K1 | 0.954 (0.893–0.980) | 0.982 (0.959–0.992) | 0.997 (0.993–0.999) | 0.979 (0.954–0.990) | 0.947 (0.882–0.976) | 0.969 (0.932–0.986) |
| K2 | 0.814 (0.609–0.918) | 0.979 (0.953–0.991) | 0.997 (0.993–0.999) | 0.954 (0.901–0.979) | 0.949 (0.886–0.977) | 0.968 (0.930–0.985) |
| Astigmatism | 0.913 (0.806–0.963) | 0.955 (0.901–0.980) | 0.942 (0.859–0.977) | 0.964 (0.922–0.983) | 0.798 (0.550–0.910) | 0.810 (0.583–0.913) |
| Axis | 0.984 (0.962–0.993) | 0.970 (0.933–0.987) | 0.978 (0.945–0.991) | 0.976 (0.948–0.989) | 0.960 (0.911–0.982) | 0.964 (0.921–0.984) |
| J0 | 0.918 (0.825–0.963) | 0.984 (0.963–0.993) | 0.984 (0.960–0.994) | 0.989 (0.976–0.995) | 0.891 (0.761–0.950) | 0.893 (0.766–0.951) |
| J45 | 0.621 (0.313–0.812) | 0.920 (0.827–0.965) | 0.950 (0.877–0.980) | 0.522 (−0.018 to 0.776) | 0.815 (0.593–0.915) | 0.690 (0.320–0.859) |
Notes: All values are presented as median (minimum–maximum). All ICCs, both intra-keratometer and inter-keratometer, have P<0.001. Repeatability was high for all measurement techniques and assessed parameters, with ICC >0.900, except for K2 and J45 measured by Cassini (ICC <0.900). Comparability was also high for most parameters (ICC >0.900), but several inter-keratometer comparisons showed ICC <0.900.
Abbreviations: ICC, intraclass correlation coefficient; K1, minimum keratometry; K2, maximum keratometry.