P Hendriks1, W E M Berkhout2, C I Kaanen2, J H Sluijter2, I J Visser2, J J van den Dobbelsteen3, L F de Geus-Oei2,4, A G Webb5, M C Burgmans2.
Abstract
PURPOSE: To investigate the performance of two microwave ablation (MWA) systems regarding ablation volume, ablation shape and variability.Entities:
Keywords: Ablation volume; Amica; Emprint; Microwave ablation; Sphericity; Variability
Mesh:
Year: 2021 PMID: 33462682 PMCID: PMC8172387 DOI: 10.1007/s00270-020-02742-9
Source DB: PubMed Journal: Cardiovasc Intervent Radiol ISSN: 0174-1551 Impact factor: 2.740
Figure 1Experimental ablation set-up. The liver lobe is fixated within a plastic container with an antenna placed in the horizontal and vertical centre
Figure 2Short-axis diameters (SAD1 and SAD2) and long-axis diameter (LAD) of the ablation zone
Figure 3Sagittal MRI of ex vivo porcine livers after ablation. A Ablation zone after Emprint ablation of 3 minutes at 80 W. B Ablation zone of Amica ablation of 3 minutes at 80 W
Ablation volume for each setting and system
| Settings | Emprint ( | Amica ( | ||
|---|---|---|---|---|
| Median volume (mL) | (Range) | Median volume (mL) | (Range) | |
| 3 min, 60 W | 7.0 | (6.1–7.5) | 15.1 | (9.4–19.3) |
| 3 min, 80 W | 10.3 | (7.6–12.4) | 20.5 | (13.4–34.7) |
| 5 min, 60 W | 13.2 | (10.3–14.7) | 21.7 | (17.2–32.1) |
| 5 min, 80 W | 18.1 | (11.5–21.5) | 31.7 | (24.3–44.7) |
| Total | 11.1 | (6.1–21.5) | 21.1 | (9.4–44.7) |
Figure 4Each individual ablation volume per setting and system
Median long-axis and short-axis diameters for each setting and system
| Settings | Emprint ( | Amica ( | ||||||
|---|---|---|---|---|---|---|---|---|
| Long-axis diameter (mm) | Short-ax-is diameter (mm) | Long-axis diameter (mm) | Short-axis diameter (mm) | |||||
| Median | (Range) | Median | (Range) | Median | (Range) | Median | (Range) | |
| 3 min, 60 W | 25.4 | (21.4–27.3) | 22.2 | (19.4–24.8) | 45.0 | (39.0–51.9) | 25.4 | (19.1–28.4) |
| 3 min, 80 W | 30.6 | (26.7–33.3) | 25.5 | (20.5–27.7) | 53.1 | (49.7–71.0) | 28.1 | (20.2–35.8) |
| 5 min, 60 W | 30.1 | (28.0–31.9) | 28.2 | (24.6–31.0) | 53.8 | (23.4–60.5) | 29.0 | (24.1–47.0) |
| 5 min, 80 W | 33.9 | (30.8–34.7) | 30.1 | (25.7–33.3) | 57.0 | (50.5–62.2) | 32.9 | (25.4–41.4) |
| Total | 30.4 | (21.4–34.7) | 26.7 | (19.4–33.3) | 52.5 | (39.0–71.0) | 28.2 | (19.1–47.0) |
Figure 5Ablation axis sizes for all individual ablations per setting and system: A long-axis diameter (LAD) was measured along the MWA antenna and B short-axis diameters (SAD1 and SAD2) were measured orthogonal to the LAD
Sphericity index for the Emprint and Amica system at different settings
| Settings | Emprint ( | Amica ( | ||
|---|---|---|---|---|
| Mean sphericity index | (Range) | Mean sphericity index | (Range) | |
| 3 min, 60 W | 0.90 | (0.85–1.02) | 0.55 | (0.52–0.60) |
| 3 min, 80 W | 0.83 | (0.75–0.97) | 0.49 | (0.44–0.52) |
| 5 min, 60 W | 0.93 | (0.90–0.98) | 0.71 | (0.44–1.57) |
| 5 min, 80 W | 0.91 | (0.86–0.95) | 0.59 | (0.46–0.74) |
| Total | 0.89 | (0.75–1.02) | 0.59 | (0.44–1.57) |
Figure 6Sphericity index of the ablation zone for the Emprint and Amica system at different settings