Patient specific selection of lateral wall cochlear implant electrodes based on anatomical indication ranges.

Max Eike Timm1, Omid Majdani1, Tobias Weller1, Mayra Windeler1, Thomas Lenarz1, Andreas Büchner1, Rolf Benedikt Salcher1.   

Abstract

OBJECTIVES: The aim of this study was to identify anatomical indication ranges for different lateral wall cochlear implant electrodes to support surgeons in the preoperative preparation.
METHODS: 272 patients who were implanted with a FLEX20, FLEX24, FLEX28, or a custom-made device (CMD) were included in this study. The cochlear duct length (CDL) and basal cochlear diameter (length A) were measured within preoperative imaging data. The parameter A was then employed to additionally compute CDL estimates using literature approaches. Moreover, the inserted electrode length (IEL) and insertion angle (IA) were measured in postoperative CT data. By combining the preoperative measurements with the IA data, the covered cochlea length (CCL) and relative cochlear coverage (CC) were determined for each cochlea.
RESULTS: The measurements of the CDL show comparable results to previous studies. While CDL measurements and estimations cover similar ranges overall, severe deviations occur in individual cases. The electrode specific IEL and CCL are fairly consistent and increase with longer electrodes, but relatively wide ranges of electrode specific CC values were found due to the additional dependence on the respective CDL. Using the correlation of IEL and CCL across electrode arrays, CDL ranges for selected arrays were developed (FLEX24: 31.3-34.4, FLEX28: 36.2-40.1, FLEXSoft: 40.6-44.9).
CONCLUSIONS: Our analysis shows that electrode specific CC varies due to the CDL variation. Preoperative measurement of the CDL allows for an individualized implant length selection yielding optimized stimulation and a reduced risk of intraoperative trauma. The CDL, as derived from preoperative CT imaging studies, can help the implant surgeon select the appropriate electrode array to maximize the patient's outcomes.

Entities:  

Mesh:

Year:  2018        PMID: 30365565      PMCID: PMC6203394          DOI: 10.1371/journal.pone.0206435

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Cochlear implantation is a technology for patients with total, severe or frequency specific hearing loss which can restore the patient’s ability to understand speech [1,2]. The cochlear implant (CI) works by directly stimulating the auditory nerve. This is accomplished by inserting a cochlear implant electrode array into the patient’s cochlea. An electric field stimulus is then applied by a number of contacts distributed along the electrode array, targeting the spiral ganglion cells and auditory nerve fibers. For cochlear implantation, various types of electrode arrays from different manufacturers are available. These electrode arrays differ in size, length, number of electrode contacts and material characteristics [3]. Prior to surgery, a decision must be made by the patient and physician on which electrode to implant. To do so, multiple factors must be taken into account including the residual hearing and medical history of the patient, which may include otosclerosis, patient preference as well as the length and shape of the cochlea. The cochlear length is known to have large variations [4-22]. Previous studies have shown that for patients who only hear with their CI, improved outcomes after CI surgery can be expected with longer electrode arrays and accordingly deeper insertion angles [23-25]. Other studies have shown that the insertion angle not only depends on the electrode array type but also on the length of the cochlea [26,27]. Furthermore, dysplasia and other syndromes exist which have an effect on cochlear geometry, especially regarding the length, shape and number of turns (e.g. Mondini dysplasia) [28]. Different methods are available for the evaluation of the cochlea duct length (CDL) and the depth of insertion within clinical imaging data of cochlea without malformation: one option is to manually trace the contour of the cochlea or electrode array and subsequently use spline interpolation to determine the corresponding length [15,29]. Other methods, which are based on mathematical correlations, use the basal diameter A (within a logarithmic equation) to estimate the cochlea or array length [30-32]. The benefit of the latter is that these types of estimates do not require special software tools but can be employed using common DICOM viewers. However, if these estimations are used for patient specific considerations on which CI array to use, the corresponding CDL values must be accurate and reliable. In order to address both the impact of cochlear length variations onto cochlear implant surgery as well as the reliability of popular literature approaches to assess this variability, the proposed study was conducted. Based on a large dataset of imaging data of CI patients, evaluations were performed on the distribution of CDL values, the correlation of electrode array length and the length of the cochlea covered by the respective array and suitability of specific CI arrays for certain ranges of CDL values. Following up on the study of Rivas et al. [33] who addressed the impact of A-value assessment deviations onto the electrode choice, it was further evaluated to which extent inaccuracies of the A-value method itself [30] would have led to a different choice of CI array than the respective contour tracings.

Materials and methods

Ethics statement

The ethics committee of the Hannover Medical School, Germany, approved this retrospective study. Due to the retrospective design, no written information was given to the patients of the study group. All patient data were anonymized and de-identified prior the retrospective analysis.

Subjects

At the Hannover Medical School, pre- and postoperative imaging of all CI patients is obtained by either Cone Beam CT (CBCT) or standard CT scans. We performed a retrospective study of 272 preoperative imaging datasets of patients who were implanted with a MED-EL FLEX20, FLEX24 or FLEX28 electrode between 2006 and 2017. Postoperative scans were available in 259 of these cases. Furthermore, patients who received custom-made devices (CMD) of 16mm length as well as 16 mm partial insertions of FLEX24 and 20 mm insertions with the FLEX28 were evaluated. Partial insertions were performed in an attempt to preserve residual hearing which were all considered CMD for the purpose of data analysis. Within the overall study group, most patients were implanted with a MED-EL FLEX28 (165 patients), 46 patients with a FLEX24, 52 patients with a FLEX20 and 12 patients with a MED-EL Flex CMD (see Fig 1). In our practice, pre- and postoperative imaging are a routine part of clinical care analyses.
Fig 1

Study case overview.

Overview of implanted electrode arrays which were included into this study. Note that the CMD group consists of FLEX16 electrode arrays and partial insertions of FLEX24 and FLEX28 electrodes).

Study case overview.

Overview of implanted electrode arrays which were included into this study. Note that the CMD group consists of FLEX16 electrode arrays and partial insertions of FLEX24 and FLEX28 electrodes).

Imaging data analysis

All datasets were evaluated using the DICOM-Viewer OsiriX MD (version 2.5.1 64bit, Pixmeo SARL, Switzerland). The corresponding analysis included the following steps: Tracings of the cochlear lateral wall from the center of the round window to the apex yield the corresponding CDL (see Figs 2 and 3)[15]. This is an automated feature within OsiriX MD. Performing one of these measurements takes approximately 2 minutes. An example is given in the supplementary material of this manuscript (see S1 Video).
Fig 2

Spline measurement with OsiriX MD.

Visualization of lateral wall tracing in OsiriX MD in top (A) and side views (B, C).

Fig 3

Complete segmentation of the lateral wall.

Visualization of a completed lateral wall tracing in OsiriX MD (A, B) with the additional display of the enrolled path (C) for which the length is computed automatically.

Measurements of the basal turn diameter A as well as the cochlear angle (CA) (see Fig 4).
Fig 4

Measurement of global cochlea dimensions.

Visualization of measurements of A (defined as the distance from the round window through the modiolus to the opposite wall of the cochlea) and B (maximal distance orthogonal to A).

Measurement of the insertion angle (IA), defined as the angle from the center of the round window to the most apical contact (see Fig 5).
Fig 5

Inclusion of postoperative imaging analysis.

(A) Postoperatively, marker points were placed in the center of the round window and along the inserted electrode array in the middle of the respective contact artefacts, and the corresponding insertion angle was measured. (B) The latter value was then used within the corresponding preoperative imaging data to determine the length along the lateral wall up to the insertion angle.

Tracings of the cochlear lateral wall in the preoperative scan (in order to avoid inaccuracies due to artifacts of the implanted array, see Fig 5) from the center of the round window to the insertion angle, yielding the covered cochlea length (CCL). Measurement of the inserted electrode length (IEL) by placing marker points in the centers of all 12 electrode contacts as well as the entrance point of the electrode array in the round window (see Fig 5). Computation of the individual cochlear coverage (CC) in percent by dividing the corresponding CCL by the respective CDL.

Spline measurement with OsiriX MD.

Visualization of lateral wall tracing in OsiriX MD in top (A) and side views (B, C).

Complete segmentation of the lateral wall.

Visualization of a completed lateral wall tracing in OsiriX MD (A, B) with the additional display of the enrolled path (C) for which the length is computed automatically.

Measurement of global cochlea dimensions.

Visualization of measurements of A (defined as the distance from the round window through the modiolus to the opposite wall of the cochlea) and B (maximal distance orthogonal to A).

Inclusion of postoperative imaging analysis.

(A) Postoperatively, marker points were placed in the center of the round window and along the inserted electrode array in the middle of the respective contact artefacts, and the corresponding insertion angle was measured. (B) The latter value was then used within the corresponding preoperative imaging data to determine the length along the lateral wall up to the insertion angle.

Measurement data analysis

After measurement data was acquired according to the methodology stated above, data analysis regarding the CDL and its estimation was performed in the following manner: Estimation of the cochlea length using the estimation method of Escudé et al. [30] by using A and the average CA value of 900 deg (or 2.5 turns): Comparison of the measured CDL values derived by the lateral wall tracings and the ones estimated using the above equation. Results regarding the coverage of the cochlea with specific electrode arrays, corresponding anatomical indication ranges for each electrode array and clinically relevant evaluation errors were derived as follows: Correlation of IEL and CCL in order to derive the relation between the length of the inserted electrode and the covered length of the cochlea. Determining the CDL indication ranges for the different electrode arrays was based on (a) the previous correlation and (b) the manufacturer’s recommendation of 80% cochlear coverage (CC) [34]. Evaluating the accuracy of Escude’s method in terms of estimated CDL values suggesting the same CI array indication as the corresponding lateral wall measurements.

Statistical analysis

The data was statistically analyzed using IBM SPSS Statistics (Version 24.0.0.0). To test whether a normal distribution exists we used the Kolmogorov-Smirnov test.

Results

All measurement data can be found in S1 Table. Fig 6A and 6B show histograms of both measured and estimated CDLs of the 272 cochleae: In both cases a normal distribution was found with a mean length of 37.9mm and standard deviations of 2.4mm and 2.3mm respectively. The actual deviations of measurements and estimations are shown in Fig 6C: while the general trend of estimations and measurements is in agreement (R2 = 0.37), the mean absolute deviation was found to be at 1.4mm +/-1.1mm with a maximal deviation of 6.8mm.
Fig 6

CDL distribution.

Histograms of (A) measured and (B) estimated CDL values. (C) comparison of individual CDL measurements and estimations.

CDL distribution.

Histograms of (A) measured and (B) estimated CDL values. (C) comparison of individual CDL measurements and estimations. In order to evaluate the clinical relevance of these deviations, the postoperative measurement data was included into the analysis. Note that no tip fold over could be observed in any of the reviewed cases. Fig 7A shows the derived IEL and CCL values in a boxplot, grouped by the respective electrode array type. Means and the standard deviations of the IEL for the different arrays are 15.4mm +/-2.7mm for the CMD group, 18.7mm +/- 1.1 mm for the FLEX20, 23.5mm +/- 0.9 mm for the FLEX24 and 26.6mm +/- 1.1mm for the FLEX28. Mean values and standard deviations of the CCL are slightly larger (about 2 mm on average) than the respective IEL values with 20.8mm +/- 1.1mm for the FLEX20, 25.2mm +/- 1.2mm for the FLEX24 and 29.2mm +/-1.4mm for the FLEX28. The mean CCL of the CMD group was found to be 17mm +/- 2.9mm. Taking the preoperatively measured CDL into account, the individual cochlear coverage (CC) of the lateral wall can be calculated as the ratio of CCL and CDL, which is shown in Fig 7B. The achieved mean CC is 56% +/-3.5% for the FLEX20, 67.9% +/-6.1% for the FLEX24 and 76.4% +/-5.3% for the FLEX28. For the CMD devices a coverage of 46% +/- 7.6% was achieved. The CC variation for the longest electrode (FLEX28) ranged between 57.8% and 95.2% while it ranged from 58–85.3% for the FLEX24 and from 32.6–43.2% for the FLEX20.
Fig 7

Distribution of electrode specific CI outcomes.

(A) IEL and CCL distribution as well as (B) CC ranges for the different electrode array groups.

Distribution of electrode specific CI outcomes.

(A) IEL and CCL distribution as well as (B) CC ranges for the different electrode array groups. In order to correlate IEL and CCL, a scatter plot of the two is given in Fig 8A. Linear regression of these data points yielded the following correlation function:
Fig 8

Anatomical indication ranges.

(A) Depiction of the CCL range computed for a fully inserted (+/-5%) FLEX28 array, (B) visualization of the corresponding CDL indication ranges for FLEX24, FLEX28 and FLEXSoft electrode arrays.

The red area within the graph indicates (based on the example of a FLEX28 electrode array) the CCL range which can be expected for a successfully implanted array, if successful insertion is assumed to lie within +/- 5% of the array length that is supposed to be implanted (i.e. 28mm for a FLEX28 electrode array), the correlation function above can be employed to derive the corresponding CCL range. Table 1 shows the corresponding IEL and CCL ranges for different MED-EL electrode arrays suitable for patients without residual hearing. As mentioned before, the manufacturer recommends 80% CC (Mistrík & Jolly, 2016). Thus, translating the CCL to a clinical indicated range can be accomplished by dividing the derived CCL ranges by 0.8. The corresponding indicated ranges for the different arrays are listed in Table 1 and depicted in Fig 8B. Note that the indicated range for the FLEX28 array matches the peak of the derived CDL distribution.
Table 1

Indication range for the FLEX electrodes.

ArrayFLEX 24FLEX 28FLEX Soft
Array-length in mm242831,5
CCL in mm26.3(25–27.5)30.5 (29–32)34.2(32.5–35.9)
Recommended CDL (lateral wall) in mm32.938.242.7
Indicated CDL (lateral wall) ranges (mm)31.3–34.436.2–40.140.6–44.9
Recommended organ of Corti length (mm)26.530.534
Indicated organ of Corti range (mm)25.3–27.729.1–3232.4–35.7

Anatomical indication ranges.

(A) Depiction of the CCL range computed for a fully inserted (+/-5%) FLEX28 array, (B) visualization of the corresponding CDL indication ranges for FLEX24, FLEX28 and FLEXSoft electrode arrays. Several publications refer to the CDL as the length of the organ of Corti (OC) and not the lateral wall. That is why the data of Hardy and Lee [5,31,35] was used to project the lateral wall indication ranges onto the organ of Corti (see Fig 9A). Linearly relating the CDL mean values +/- one standard deviation for lateral wall (37.9 mm +/- 2.4 mm) and organ of Corti (31.5 mm +/-2.3 mm) yielded the following equation:
Fig 9

Organ of Corti length.

(A) Histogram of the organ of Corti (OC) derived by Hardy and Lee (red) and MHH data (black), (B) Histogram of calculated OC length for MHH data with electrode lengths for FLEX24, FLEX28 and FLEXSoft (+/- 5%).

This equation was used to translate the lateral wall indication ranges into the ones for organ of Corti, which are displayed in Fig 9B and stated within the last two rows of Table 1.

Organ of Corti length.

(A) Histogram of the organ of Corti (OC) derived by Hardy and Lee (red) and MHH data (black), (B) Histogram of calculated OC length for MHH data with electrode lengths for FLEX24, FLEX28 and FLEXSoft (+/- 5%). Finally, we evaluated if CDL estimations using the A value and CA [30] could be used instead of actual measurements to derive an accurate anatomical indication for a specific array. Note that only the length of the cochlea and not the length of the implanted array were taken into account for this analysis. Using the FLEX28 implanted group we then analyzed to what extent the CDL estimation method would predict cochlear anatomies as too short, applicable or too long for this array. Within Fig 10A, the x-axis represents the measured CDL whereas the y-axis shows the corresponding estimations. The shaded areas within this graph represent the FLEX28 indication ranges for measured and estimated CDL values respectively (3rd column of Table 1). Correct identification, i.e. a match of measured and estimated indication (highlighted in the table), of short cochleae could be achieved in 67.8%, of long cochleae in 17.1% and of matching cochleae in 77.2% of the cases (see Table 2 and Fig 10B).
Fig 10

CDL identification of FLEX28 candidates.

(A) the comparison of measurement and estimation-based identifications of FLEX28 candidates resulted in (B) correct identifications in 66% of the analyzed anatomies using Escudé’s method.

Table 2

Correct identification of cochlea length.

Reference smallReference matchReference largeCorrect Identification
Estimation large01177
Estimation match1913233132
Estimation small4028140
total number5917141271
correct identification in percentage67.877.217.166.1

CDL identification of FLEX28 candidates.

(A) the comparison of measurement and estimation-based identifications of FLEX28 candidates resulted in (B) correct identifications in 66% of the analyzed anatomies using Escudé’s method.

Discussion

In our study, we analyzed the anatomy of the cochlea in 272 clinical imaging datasets. The postoperative location of different lateral wall electrode arrays could be assessed in 259 of these 272 cases. The derived variability of the cochlea length is in agreement with previous studies [4,5,15-17] and highlights again the importance of considering the patient specific anatomy in the field of cochlear implantation (see Fig 6A). The projection of the wide portfolio of available electrode arrays [3] onto this range of cochlear length values then allows for optimal coverage of the intracochlear neural structures: achieving 80% CC with lateral wall electrodes in case of an average or long cochlea, for instance, would currently only be achievable with MED-EL devices who also offer electrode array lengths of more than 26mm. However, it was not yet clearly proven if electrode arrays stimulate the neural fiber endings at the organ of Corti or the spiral ganglion cells directly, i.e. what cochlear coverage achieves the best possible speech perception. Nevertheless, recent publications show superior speech understanding for the FLEX28 array (which typically achieves insertion angles of 540–720 degrees) in comparison to shorter electrode arrays [24]. The derived indication ranges yield a possible explanation for this finding, i.e. the sufficient coverage of neural structures for most cochleae with the FLEX28 array. The comparison of measured and estimated [30] CDLs showed normal distributions in either case with no significant differences, but quite severe deviations were found for individual cases (see Fig 6C) with deviations of up to 6.8mm. Hence, the CDL estimation method described by Escudé et al. [30] may be applicable for retrospective analysis but is not suitable for clinical use where accuracy matters for each and every individual case. This was further highlighted in Fig 10B where the estimation deviations were evaluated in terms of anatomical indications for specific CI electrode array lengths: the results show that in total, correct indications could only be derived in 66% of the analyzed cochleae. Especially concerning are the 11% of the cochleae whose CDL values were overestimated: in a clinical setting, overestimations could result in implantations of electrode arrays too long for a specific cochlea such that the array cannot be fully inserted and/or possibly cause intracochlear damage. Underestimations, on the other hand, were found to occur in 22% of the cases and could entail insufficient coverage of the cochlear neurons and hence result in poor speech understanding, since studies show that longer electrodes result in a better speech understanding for electric stimulation only [23,24]. The IEL and corresponding CCL values for the analyzed electrode arrays are shown in Fig 7A. CCL values are larger than the according IEL values for each electrode array due to the latter being located in closer proximity to the modiolus than the lateral wall and thus representing a shorter distance. The projection of the array from the modiolus out onto the lateral wall hence creates a larger spiral profile and an accordingly larger CCL value. The group of CMDs shows the highest deviation regarding the IEL to CCL correlation. This fact is owed to the CMD group containing FLEX24 and FLEX28 arrays which were only partially inserted (covering only the individual frequency region with severe to profound hearing loss) as well as custom made devices of 16mm length. While CC increases for longer electrodes overall (see Fig 7B), it can be seen that the CC distributions overlap especially regarding the FLEX24 and FLEX28 arrays. This can be explained by the fact that the CC is not only dependent on the length of the electrode array but also on the length of the cochlea [27]. In some cases, a FLEX24 in a short cochlea may therefore cover a larger fraction of the cochlea than a FLEX28 in a large cochlea. It should further be noted that the smallest CC values can be found within the FLEX20 group because this electrode was developed to not cover the full frequency range of the cochlea in order for patients to benefit from acoustic stimulation in the low frequencies using an EAS system. Variations in CC values generated with the same array can also be seen in Fig 8A. Although the overall cochlea length does not directly influence CCL values, it can be assumed that the same kind of array will be situated differently in a small cochlea than in a large one. Variations in cochlear anatomy are therefore likely to also influence the length of the cochlea covered by an electrode array. Regarding the anatomical indication ranges it was found based on the measured CDL that most cochleae are covered by the FLEX28 with a target CDL range of 36.2mm to 40.1mm (11+132+28 of the 271 cases corresponding to 63% of the analyzed cases). Without knowing the anatomy, a FLEX28 electrode also seems to be the best choice since it could be implanted in 76.8% (see Fig 11) of the cases with a reduced risk of harming any intracochlear structures.
Fig 11

Extended indication range for MED-EL FLEX28.

Structure preservation indication range for the FLEX28 array which could have been fully inserted in 76% of the investigated cases.

Extended indication range for MED-EL FLEX28.

Structure preservation indication range for the FLEX28 array which could have been fully inserted in 76% of the investigated cases.

Conclusion

Following previous investigations, we showed that due to the variation of cochlear size, preoperative cochlea length assessment should be included into the routine preoperative care. Furthermore, individual implant selection based on the patient specific cochlea length is feasible and likely to improve implantation outcomes. However, length estimations exclusively based on mathematical correlations may result in false recommendations and could potentially injure the cochlea and lead to poorer outcomes. Furthermore, the presented results demonstrate that most cochleae are sufficiently covered by a FLEX28 electrode array while the FLEX24 and FLEXSoft should be employed for short and long cochleae respectively.

Video of CDL measurement in preoperative data.

Preoperative measurement of a cochlear duct length (CDL) in a Cone Beam CT scan (CBCT). (MOV) Click here for additional data file.

Measurement data of the study.

(XLSX) Click here for additional data file.
  31 in total

Review 1.  How cochlear implants encode speech.

Authors:  Jay T Rubinstein
Journal:  Curr Opin Otolaryngol Head Neck Surg       Date:  2004-10       Impact factor: 2.064

2.  Diversity in cochlear morphology and its influence on cochlear implant electrode position.

Authors:  Kim S van der Marel; Jeroen J Briaire; Ron Wolterbeek; Jorien Snel-Bongers; Berit M Verbist; Johan H M Frijns
Journal:  Ear Hear       Date:  2014 Jan-Feb       Impact factor: 3.570

3.  The internal dimensions of the cochlear scalae with special reference to cochlear electrode insertion trauma.

Authors:  Slavomir Biedron; Andreas Prescher; Justus Ilgner; Martin Westhofen
Journal:  Otol Neurotol       Date:  2010-07       Impact factor: 2.311

4.  Visualization, measurement and modelling of the cochlea using rotating midmodiolar slice planes.

Authors:  G Jakob Lexow; Daniel Schurzig; Nils-Claudius Gellrich; Thomas Lenarz; Omid Majdani; Thomas S Rau
Journal:  Int J Comput Assist Radiol Surg       Date:  2016-03-19       Impact factor: 2.924

5.  Insertion depth impacts speech perception and hearing preservation for lateral wall electrodes.

Authors:  Brendan P O'Connell; Jacob B Hunter; David S Haynes; Jourdan T Holder; Matt M Dedmon; Jack H Noble; Benoit M Dawant; George B Wanna
Journal:  Laryngoscope       Date:  2017-03-17       Impact factor: 3.325

6.  Evaluation of Rigid Cochlear Models for Measuring Cochlear Implant Electrode Position.

Authors:  Ahmet Cakir; Robert F Labadie; M Geraldine Zuniga; Benoit M Dawant; Jack H Noble
Journal:  Otol Neurotol       Date:  2016-12       Impact factor: 2.311

7.  In vivo measures of cochlear length and insertion depth of nucleus cochlear implant electrode arrays.

Authors:  D R Ketten; M W Skinner; G Wang; M W Vannier; G A Gates; J G Neely
Journal:  Ann Otol Rhinol Laryngol Suppl       Date:  1998-11

8.  Cochlea size variability and implications in clinical practice.

Authors:  P Pelliccia; F Venail; A Bonafé; M Makeieff; G Iannetti; M Bartolomeo; M Mondain
Journal:  Acta Otorhinolaryngol Ital       Date:  2014-02       Impact factor: 2.124

9.  Investigation of the effect of cochlear implant electrode length on speech comprehension in quiet and noise compared with the results with users of electro-acoustic-stimulation, a retrospective analysis.

Authors:  Andreas Büchner; Angelika Illg; Omid Majdani; Thomas Lenarz
Journal:  PLoS One       Date:  2017-05-15       Impact factor: 3.240

10.  Variations in microanatomy of the human cochlea.

Authors:  Ersin Avci; Tim Nauwelaers; Thomas Lenarz; Volkmar Hamacher; Andrej Kral
Journal:  J Comp Neurol       Date:  2014-04-12       Impact factor: 3.215

View more
  16 in total

1.  Preoperative prediction of angular insertion depth of lateral wall cochlear implant electrode arrays.

Authors:  Mohammad M R Khan; Robert F Labadie; Jack H Noble
Journal:  J Med Imaging (Bellingham)       Date:  2020-06-03

2.  Hybrid active shape and deep learning method for the accurate and robust segmentation of the intracochlear anatomy in clinical head CT and CBCT images.

Authors:  Yubo Fan; Dongqing Zhang; Rueben Banalagay; Jianing Wang; Jack H Noble; Benoit M Dawant
Journal:  J Med Imaging (Bellingham)       Date:  2021-11-24

3.  Cochlear Duct Length Measurements in Computed Tomography and Magnetic Resonance Imaging Using Newly Developed Techniques.

Authors:  Johannes Taeger; Franz Tassilo Müller-Graff; Lukas Ilgen; Phillip Schendzielorz; Rudolf Hagen; Tilman Neun; Kristen Rak
Journal:  OTO Open       Date:  2021-09-24

4.  Is Cochlear Length Related to Congenital Sensorineural Hearing Loss: Preliminary Data.

Authors:  Mehmet Bilgin Eser; Başak Atalay; Mahmut Tayyar Kalcıoğlu
Journal:  J Int Adv Otol       Date:  2021-01       Impact factor: 1.017

5.  Uncoiling the Human Cochlea-Physical Scala Tympani Models to Study Pharmacokinetics Inside the Inner Ear.

Authors:  Daniel Schurzig; Max Fröhlich; Stefan Raggl; Verena Scheper; Thomas Lenarz; Thomas S Rau
Journal:  Life (Basel)       Date:  2021-04-21

6.  Determination of Cochlear Duct Length With 3D Versus Two-dimensional Methods: A Retrospective Clinical Study of Imaging by Computed Tomography and Cone Beam Computed Tomography.

Authors:  Stephan Waldeck; Christian VON Falck; Rene Chapot; Marc Brockmann; Daniel Overhoff
Journal:  In Vivo       Date:  2021 Nov-Dec       Impact factor: 2.155

7.  Initial Hearing Preservation Is Correlated With Cochlear Duct Length in Fully-inserted Long Flexible Lateral Wall Arrays.

Authors:  Emily S Hollis; Michael W Canfarotta; Margaret T Dillon; Meredith A Rooth; Andrea L Bucker; Sarah A Dillon; Allison Young; Kristen Quinones; Harold C Pillsbury; Matthew M Dedmon; Brendan P O'Connell; Kevin D Brown
Journal:  Otol Neurotol       Date:  2021-09-01       Impact factor: 2.619

8.  Incidence of Complete Insertion in Cochlear Implant Recipients of Long Lateral Wall Arrays.

Authors:  Michael W Canfarotta; Margaret T Dillon; Kevin D Brown; Harold C Pillsbury; Matthew M Dedmon; Brendan P O'Connell
Journal:  Otolaryngol Head Neck Surg       Date:  2021-02-16       Impact factor: 5.591

Review 9.  Review on cochlear implant electrode array tip fold-over and scalar deviation.

Authors:  Anandhan Dhanasingh; Claude Jolly
Journal:  J Otol       Date:  2019-01-09

10.  Comment on "The Effect of Cochlear Size on Cochlear Implantation Outcomes".

Authors:  M Bilgin Eser; Başak Atalay; M Tayyar Kalcıoğlu
Journal:  Biomed Res Int       Date:  2020-07-29       Impact factor: 3.411

View more