| Literature DB >> 31467506 |
Anandhan Dhanasingh1, Claude Jolly1.
Abstract
OBJECTIVE: Determine the occurrence rate of cochlear implant (CI) electrode tip fold-over and electrode scalar deviation as reported in patient cases with different commercial electrode types. DATA-SOURCES: PubMed search for identifying peer-reviewed articles published till 2018 on CI electrode tip fold-over and scalar deviation. Key-words for searching were "Cochlear electrode tip fold-over", "Cochlear electrode scalar position" and "Cochlear electrode scalar location". ARTICLES-SELECTION: Only if electrode related issues were investigated in patient cases. 38 articles met the inclusion-criteria.Entities:
Keywords: Electrode tip fold-over; Pre-curved electrode; Scalar deviation; Straight electrode
Year: 2019 PMID: 31467506 PMCID: PMC6712287 DOI: 10.1016/j.joto.2019.01.002
Source DB: PubMed Journal: J Otol ISSN: 1672-2930
Fig. 1Literature review process.
Summarizing the 13 studies on tip fold-over.
| Study | No. of cases taken for analysis/method | No. of electrode per type/brand | No. of cases reported tip fold-over | ||
|---|---|---|---|---|---|
| A | B | C | |||
| 275 (CT) | – | – | 275 (F28, F24, F20, F16) | 0 | |
| 23 (CT) | – | MS (23) | – | 2 | |
| 1722 (CT) | 7 (CA), 6 (SM), 2(SS) | ||||
| 65 (Intra-op CT) | CA (12), SM (1), SS (31) | 1J (2), MS (3) | F28 (13) | 1 (SM) | |
| 2 (x-ray & CT) | – | MS | – | 2 | |
| 117 (intra-op x-ray) | SM | – | – | 9 | |
| 1 (Intra-op fluoroscopy) | SM | – | – | 1 (100%) | |
| 45 (intra-op x-ray/fluoroscopy) | SM | – | – | 2 cases. 1st case corrected in the same surgery. 2nd case underwent revision surgery | |
| 303 (CT) | CA, SS | MS, 1J | – | 3(CA), 1(MS), 1 (SS) and 1(1J) | |
| 63 (CT) | – | – | F24,F28, Std | 1 | |
| 215 (Post-op x-ray) | CA, SS | – | Std, M, S, FL, F28 | 2 (F) | |
| 277 (Intra-op x-ray) | CA | – | – | 5 | |
| 72 (Intra-op rotational x-ray) | CA | – | – | 4 | |
CA: Contour Advance; SM: Slim-Modiolar; MS: Mid-Scala; SS: Slim Straight; Std: Standard; M: Medium; S: Compressed; FL: FLEX SOFT, F28: FLEX28; F24: FLEX24; F20: FLEX20; F16: FLEX16.
26 studies reported on electrode scalar deviation.
| Study | No. of cases taken for analysis | Analyzing method | No. of electrode from type/brand | No. of cases reported with scalar deviation | |||||
|---|---|---|---|---|---|---|---|---|---|
| A | B | C | MH | MS | LW | ||||
| 110 | Imaging | CA (92), SM (18) | – | – | 18 (19.5%) | – | – | ||
| 23 | Imaging | – | MS (23) | – | – | 5 (22%) | – | ||
| 32 | Imaging/EcochG | – | MS (32) | – | – | 7 (22%) | – | ||
| 65 | Imaging | CA (12), SM(1), SS (31) | 1J (2), MS (3) | F28 (16), | 1 (8%) | 1 (20%) | – | ||
| 23 | Imaging | SM (23) | – | – | 6 ( | – | – | ||
| 368 | Imaging | CA (368) | – | – | 118 ( | – | – | ||
| 26 | Imaging | SS (5) | – | F28 (21) | – | – | 1 ( | ||
| 45 | Imaging | SM (45) | – | – | 0 ( | – | – | ||
| 48 | Imaging | – | – | F24, F28, Std (48) | – | – | 0 ( | ||
| 18 | EcochG | – | MS (18) | – | – | 6( | – | ||
| 50 | NRT | SS (50) | – | – | – | – | 2 ( | ||
| 24 | Imaging | CA (24), | – | – | 3 (13%) | – | – | ||
| 56 | Imaging | CA (36), SS (20) | – | – | 19 ( | – | 2 ( | ||
| 220 | Imaging | CA (115), SS (19), | 1J(21), MS (14) | F28 (28), Std (17), F24 (4) & M (2) | 59 ( | 8( | 4 ( | ||
| 45 | Imaging | CA (15) | MS (3) | 4 ( | 1( | 2 ( | |||
| SS, 1J & F collectively (27) | |||||||||
| 39 | Imaging | – | – | F28 (18), FL (17), F24 (4) | – | – | 0 ( | ||
| 23 | NRT | CA (23) | – | – | 6 ( | – | – | ||
| 85 | NRT | CA (85) | – | – | 16 ( | – | – | ||
| 61 | Imaging | CA (31), | – | FL, F28, F24, Std (30) | 8 ( | – | 0 (0%) FLEX, | ||
| 63 | Imaging | – | – | F28 (40), F24 (2), FL (7), Std (14) | – | – | 5 ( | ||
| 116 | Imaging | CA, MS (69) | (47) LW from all 3 CI brands | 29 ( | – | 5 ( | |||
| 215 | imaging | 107 | 108 | – | – | 1(0.92%) F | |||
| 13 | Imaging | CA (7) | 1J (3) | Std (2), F24 (1) | 3 ( | – | 1 ( | ||
| 223 | Imaging | C (21), CA (202) | – | – | 19 ( | – | – | ||
| 32 | Imaging | 20 | 10 | 2 | 7 ( | 4( | 0 ( | ||
| 23 | Imaging | C/CA (13) | H (1) | – | 6 ( | – | 7 ( | ||
| LW electrodes from brand A & B (9) | |||||||||
NRT: neural response telemetry, EcochG: electrocochleography.
One study reporting on electrode scalar deviation from older generation of electrodes which are currently not available in the market.
| Study | No. of cases taken for analysis | Analyzing method | No. of electrode from type/brand | No. of cases reported with scalar deviation | |||||
|---|---|---|---|---|---|---|---|---|---|
| A | B | C | D | MH | MS | LW | |||
| 38 | Histology | Nucleus 22 (16) (LW), 24R (1) (MH), 24 M (4) (double array LW), 24 RST (1) (LW) | Clarion (1) (LW) | MED-EL 40+ (1) (LW) | Ineraid (10) (LW) | 0 | – | Nucleus 22 (8) | |
Fig. 2Histogram chart showing the % of electrode related issues with two general types of CI electrodes.
Fig. 3Electrode tip geometry from all three major CI brands.