Contact force-sensing catheters: performance in an ex vivo porcine heart model.

Eduardo Franco1, Daniel Rodríguez Muñoz2, Roberto Matía2, Antonio Hernández-Madrid2, Inmaculada Sánchez Pérez3, José Luis Zamorano2, Javier Moreno2.   

Abstract

PURPOSE: Contact force (CF) catheters are useful to address proper contact during ablation. However, interactions between the ablation process, or its associated irrigation flow changes, with the CF sensing may translate into unexpected CF value fluctuations. We aimed to test for unintentional CF value variations during radiofrequency applications at a fixed applied force, with two commercially available catheters (TactiCath™ and SmartTouch™), and to evaluate its theoretical clinical significance by correlating CF-derived automatic ablation algorithms (force-time integral and lesion index) and actual lesion size at two standard CF values.
METHODS: Four series of 20 perpendicular epicardial ablations (20 W, 60 s, 17 ml/min) were performed on porcine left ventricle submerged in 37 °C saline. Catheters were mechanically fixed at a constant position and evaluated at 10 and 20 g. CF values were digitally analysed before each application changing irrigation rate (2-17-30 ml/min), and during ablation. Finally, lesions were quantified.
RESULTS: Increasing irrigation before ablation led to a slight but significant CF decrease. During ablation, CF showed a reproducible pattern: fast initial decrease, subsequent increase until higher-than-initial values and final plateau phase (CF variation up to 69% at 10 g). CF variability was significantly higher at 10 g and using TactiCath™. There were no major differences in lesion size between catheters at the same initial CF. CF only correlated mildly to lesion measures, and automatic algorithms globally failed to predict lesion size.
CONCLUSIONS: CF measured values spontaneously vary during ablation following a predictable pattern (initial decrease, subsequent increase and final plateau). This is especially remarkable applying lower CF.

Entities:  

Keywords:  Contact force; Experimental model; Lesion size; Porcine heart; Radiofrequency ablation

Mesh:

Year:  2018        PMID: 30109525     DOI: 10.1007/s10840-018-0435-y

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  18 in total

1.  Electromagnetic Contact-Force Sensing Electrophysiological Catheters: How Accurate is the Technology?

Authors:  Felix Bourier; Gabriele Hessling; Sonia Ammar-Busch; Marc Kottmaier; Alessandra Buiatti; Christian Grebmer; Marta Telishevska; Verena Semmler; Carsten Lennerz; Christine Schneider; Christof Kolb; Isabel Deisenhofer; Tilko Reents
Journal:  J Cardiovasc Electrophysiol       Date:  2016-01-11

2.  THERMOCOOL® SMARTTOUCH® CATHETER - The Evidence So Far for Contact Force Technology and the Role of VISITAG™ MODULE.

Authors:  Tina Lin; Feifan Ouyang; Karl-Heinz Kuck; Roland Tilz
Journal:  Arrhythm Electrophysiol Rev       Date:  2014-05-30

3.  Contact force threshold for permanent lesion formation in atrial fibrillation ablation: A cardiac magnetic resonance-based study to detect ablation gaps.

Authors:  David Andreu; Federico Gomez-Pulido; Mireia Calvo; Alicia Carlosena-Remírez; Felipe Bisbal; Roger Borràs; Eva Benito; Eduard Guasch; Susanna Prat-Gonzalez; Rosario J Perea; Josep Brugada; Antonio Berruezo; Lluís Mont
Journal:  Heart Rhythm       Date:  2015-08-10       Impact factor: 6.343

4.  Paroxysmal AF catheter ablation with a contact force sensing catheter: results of the prospective, multicenter SMART-AF trial.

Authors:  Andrea Natale; Vivek Y Reddy; George Monir; David J Wilber; Bruce D Lindsay; H Thomas McElderry; Charan Kantipudi; Moussa C Mansour; Daniel P Melby; Douglas L Packer; Hiroshi Nakagawa; Baohui Zhang; Robert B Stagg; Lee Ming Boo; Francis E Marchlinski
Journal:  J Am Coll Cardiol       Date:  2014-08-19       Impact factor: 24.094

5.  Contact force and force-time integral in atrial radiofrequency ablation predict transmurality of lesions.

Authors:  Fabien Squara; Decebal Gabriel Latcu; Youssef Massaad; Marouane Mahjoub; Sok-Sithikun Bun; Nadir Saoudi
Journal:  Europace       Date:  2014-05       Impact factor: 5.214

6.  Fiberoptic Contact-Force Sensing Electrophysiological Catheters: How Precise Is the Technology?

Authors:  Felix Bourier; Carola Gianni; Matthew Dare; Isabel Deisenhofer; Gabriele Hessling; Tilko Reents; Sanghamitra Mohanty; Chintan Trivedi; Andrea Natale; Amin Al-Ahmad
Journal:  J Cardiovasc Electrophysiol       Date:  2016-10-24

7.  The relationship between contact force and clinical outcome during radiofrequency catheter ablation of atrial fibrillation in the TOCCATA study.

Authors:  Vivek Y Reddy; Dipen Shah; Josef Kautzner; Boris Schmidt; Nadir Saoudi; Claudia Herrera; Pierre Jaïs; Gerhard Hindricks; Petr Peichl; Aude Yulzari; Hendrik Lambert; Petr Neuzil; Andrea Natale; Karl-Heinz Kuck
Journal:  Heart Rhythm       Date:  2012-07-20       Impact factor: 6.343

8.  Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter.

Authors:  Aravinda Thiagalingam; Andre D'Avila; Lori Foley; J Luis Guerrero; Hendrik Lambert; Giovanni Leo; Jeremy N Ruskin; Vivek Y Reddy
Journal:  J Cardiovasc Electrophysiol       Date:  2010-02-01

9.  Contact Force in Atrial Fibrillation: Role of Atrial Rhythm and Ventricular Contractions: Co-Force Atrial Fibrillation Study.

Authors:  Andrea Sarkozy; Dipen Shah; Johan Saenen; Juan Sieira; Thomas Phlips; Wim Boris; Mehdi Namdar; Christiaan Vrints
Journal:  Circ Arrhythm Electrophysiol       Date:  2015-09-17

10.  Randomized, Controlled Trial of the Safety and Effectiveness of a Contact Force-Sensing Irrigated Catheter for Ablation of Paroxysmal Atrial Fibrillation: Results of the TactiCath Contact Force Ablation Catheter Study for Atrial Fibrillation (TOCCASTAR) Study.

Authors:  Vivek Y Reddy; Srinivas R Dukkipati; Petr Neuzil; Andrea Natale; Jean-Paul Albenque; Josef Kautzner; Dipen Shah; Gregory Michaud; Marcus Wharton; David Harari; Srijoy Mahapatra; Hendrik Lambert; Moussa Mansour
Journal:  Circulation       Date:  2015-08-10       Impact factor: 29.690

View more
  2 in total

1.  Effect of Baseline Impedance in Radiofrequency Delivery on Lesion Characteristics and the Relationship Between Impedance and Steam Pops.

Authors:  Lijuan Qu; Min Guo; Meng Sun; Rui Wang; Nan Zhang; Xin Li
Journal:  Front Cardiovasc Med       Date:  2022-04-27

2.  Bipolar ablation's unique paradigm: Duration and power as respectively distinct primary determinants of transmurality and steam pop formation.

Authors:  Mathews John; Ashley Rook; Allison Post; Alton Mersman; Whitney Allen; Christina Schramm; Mehdi Razavi
Journal:  Heart Rhythm O2       Date:  2020-06-24
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.