| Literature DB >> 30022887 |
Ebubekir Bakan1, Zafer Bayraktutan1, Nurcan Kilic Baygutalp2, Mehmet Ali Gul1, Fatma Zuhal Umudum1, Nuri Bakan1.
Abstract
INTRODUCTION: Automated urinalysis systems are valuable tools in clinical laboratories, especially those with a high work load. The objective of this study was to compare the analytical performance of Sysmex UN series urine analyser, which may become a new one in our laboratory, with the Cobas 6500 automated urine analyser, which is used in our laboratory for a long time. For comparisons, manual microscopical examination was accepted as reference method.Entities:
Keywords: automated urinalysis; manual microscopy; urinalysis
Mesh:
Year: 2018 PMID: 30022887 PMCID: PMC6039169 DOI: 10.11613/BM.2018.020712
Source DB: PubMed Journal: Biochem Med (Zagreb) ISSN: 1330-0962 Impact factor: 2.313
Urine sediment categories for semi-quantitative reference values
| 0 - 4 | 5 - 10 | 11 - 20 | 21 – 50 | ≥ 51 | |
| 0 - 4 | 5 - 10 | 11 - 20 | 21 – 50 | ≥ 51 | |
| Negative | Few | Moderate | High | - | |
| Negative | Positive | - | - | - | |
| RBC - red blood cells. WBC - white blood cells. HPF - high power field (x 400). LPF - low power field (x 100). | |||||
Diagnostic accuracy of automated urine analysers compared to manual microscopy
| 86.1 | 91.8 | 86.8 | 91.5 | |
| 45.3 | 95.2 | 81.0 | 79.5 | |
| 98.2 | 98.2 | 98.2 | 98.9 | |
| 99.2 | 94.9 | 89.7 | 99.6 | |
| WBC - white blood cells. RBC - red blood cells. NPV - negative predictive value. PPV - positive predictive value. Manual urine sediment analysis was used as reference. | ||||
Inter- and intra-assay coefficient of variations for WBC and RBC counts in low and high control materials in Sysmex UF-5000
| 38.5 | (19.2 - 57.8) | 9.2 | 3.2 | |
| 779.4 | (623.5 - 935.3) | 3.2 | 3.2 | |
| 40.5 | (20.2 - 60.8) | 3.2 | 4.1 | |
| 200.2 | (160.2 - 240.2) | 6.2 | 3.2 | |
| WBC - white blood cells. RBC - red blood cells. CV - coefficient of variation. LL -lower limit. UL - upper limit. | ||||
The results of the Passing-Bablok regression analysis and Bland Altman plots for WBC and RBC counts
| 1 | 1.00 (0.92 to 1.07) x | 0.530 | 0.95 | 0.95 | 1.00 | 0.340 | - 1.6 | - 35.09 | 31.82 | |
| 2 | 0.87 (0.85 to 0.91) x | 0.380 | 0.98 | 0.99 | 0.99 | < 0.001 | - 6.0 | - 28.09 | 16.12 | |
| 1 | 0.89 (0.84 to 0.98) x | 0.700 | 0.95 | 0.95 | 1.00 | 0.034 | - 4.4 | - 44.15 | 35.45 | |
| 1 | 0.95 (0.92 to 0.99) x | 0.190 | 0.96 | 0.96 | 1.00 | 0.725 | - 0.5 | - 25.85 | 24.94 | |
| Comp - comparisons: 1 - Cobas u701, 2 - Sysmex UF-5000, 3 - Manual microscopy. 95% CI – 95% confidence intervals. A – regression slope. B – regression intercept. Cusum - Cusum test for linearity. CCC – concordance correlation analysis. r - Pearson correlation coefficient. Cb - cias correction factor Cb (accuracy) (> 0.99 excellent agreement; 0.99 - 0.95 substantial agreement; 0.90 - 0.94 moderate agreement; < 0.9 poor agreement). LoA LL - lower limit of agreement. LoA UL: upper limit of agreement. WBC - white blood cells. RBC - red blood cells. | ||||||||||
Figure 1Comparison of manual analysis and Cobas u701 for white blood cells (WBC). a) Passing-Bablok regression analysis demonstrates similar performance of manual microscopy and Cobas u701 for WBC count. The regression line equation is shown in the box. Solid line - regression line. Dashed lines - 95% CI for the regression line. Dotted line - identity line (X = Y). b) Comparison of Cobas u701 with manual microscopy for WBC count using Bland-Altman analysis. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation. (B)
Figure 2Comparison of manual analysis and Sysmex UF 5000 for white blood cells (WBC). a) Passing-Bablok regression analysis demonstrates similar performance of manual microscopy and Sysmex UF-5000 for WBC count. The regression line equation is shown in the box. Solid line - regression line. Dashed lines - 95% CI for the regression line. Dotted line - identity line (X = Y). b) Comparison of Sysmex UF-5000 with manual microscopy for WBC count using Bland-Altman analysis. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation. (B)
Figure 3Comparison of Cobas u701 and Sysmex UF 5000 for white blood cells (WBC). a) Passing-Bablok regression analysis demonstrates similar performance of Cobas u701 and Sysmex UF-5000 for WBC count. The regression line equation is shown in the box. Solid line - regression line. Dashed lines - 95% CI for the regression line. Dotted line - identity line (X = Y). b) Comparison of Cobas u701 with Sysmex UF-5000 for WBC count using Bland-Altman analysis. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation. (B)
Figure 4Comparison of Cobas u701 and Sysmex UF 5000 for red blood cells (RBC). a) Passing-Bablok regression analysis demonstrates similar performance of Cobas u701 and Sysmex UF-5000 for RBC count. The regression line equation is shown in the box. Solid line - regression line. Dashed lines - 95% CI for the regression line. Dotted line - identity line (X = Y). b) Comparison of Cobas u701 with Sysmex UF-5000 for RBC count using Bland-Altman analysis. Solid line (mean) – mean difference. Dashed lines (SD) – standard deviation. (B)
Comparison of the numbers of WBC counted by the manual method and the urine analyzers
| 260 | 18 | 2 | 2 | 1 | 283 | ||
| 21 | 21 | 9 | 1 | 0 | 52 | ||
| 1 | 10 | 20 | 12 | 1 | 44 | ||
| 0 | 2 | 5 | 18 | 7 | 32 | ||
| 2 | 0 | 3 | 2 | 38 | 45 | ||
| 284 | 51 | 39 | 35 | 47 | 456 | ||
| 280 | 3 | 0 | 0 | 0 | 283 | ||
| 3 | 43 | 5 | 1 | 0 | 52 | ||
| 0 | 0 | 33 | 10 | 1 | 44 | ||
| 0 | 0 | 0 | 30 | 2 | 32 | ||
| 0 | 0 | 0 | 2 | 43 | 45 | ||
| 283 | 46 | 38 | 43 | 46 | 456 | ||
| 261 | 17 | 2 | 2 | 1 | 283 | ||
| 19 | 19 | 8 | 0 | 0 | 46 | ||
| 2 | 10 | 18 | 8 | 0 | 38 | ||
| 1 | 5 | 9 | 22 | 6 | 43 | ||
| 1 | 0 | 2 | 3 | 40 | 46 | ||
| 284 | 51 | 39 | 35 | 47 | 456 | ||
| WBC - white blood cells. HPF - high power field (x 400). | |||||||