Comparison of blood gas, electrolyte and metabolite results measured with two different blood gas analyzers and a core laboratory analyzer.

Metin Uyanik1, Erdim Sertoglu, Huseyin Kayadibi, Serkan Tapan, Muhittin A Serdar, Cumhur Bilgi, Ismail Kurt.   

Abstract

BACKGROUND: Blood gas analyzers (BGAs) are important in assessing and monitoring critically ill patients. However, the random use of BGAs to measure blood gases, electrolytes and metabolites increases the variability in test results. Therefore, this study aimed to investigate the correlation of blood gas, electrolyte and metabolite results measured with two BGAs and a core laboratory analyzer.
METHODS: A total of 40 arterial blood gas samples were analyzed with two BGAs [(Nova Stat Profile Critical Care Xpress (Nova Biomedical, Waltham, MA, USA) and Siemens Rapidlab 1265 (Siemens Healthcare Diagnostics Inc., Tarrytown, NY, USA)) and a core laboratory analyzer [Olympus AU 2700 autoanalyzer (Beckman-Coulter, Inc., Fullerton, CA, USA)]. The results of pH, pCO₂, pO₂, SO₂, sodium (Na⁺), potassium (K⁺), calcium (Ca⁺²), chloride (Cl⁻), glucose, and lactate were compared by Passing-Bablok regression analysis and Bland-Altman plots.
RESULTS: The present study showed that there was negligible variability of blood gases (pCO₂, pO₂, SO₂), K⁺ and lactate values between the blood gas and core laboratory analyzers. However, the differences in pH were modest, while Na⁺, Cl⁻, Ca²⁺ and glucose showed poor correlation according to the concordance correlation coefficient.
CONCLUSIONS: BGAs and core laboratory autoanalyzer demonstrated variable performances and not all tests met minimum performance goals. It is important that clinicians and laboratories are aware of the limitations of their assays.

Entities:  

Keywords:  Blood gas analyzer; electrolytes; interchangeability; metabolites; method comparison

Mesh:

Substances:

Year:  2014        PMID: 25431133     DOI: 10.3109/00365513.2014.981854

Source DB:  PubMed          Journal:  Scand J Clin Lab Invest        ISSN: 0036-5513            Impact factor:   1.713


  9 in total

1.  Application of Sigma Metrics and Performance Comparison Between Two Biochemistry Analyser and a Blood Gas Analyser for the Determination of Electrolytes.

Authors:  Yasemin Ustundag-Budak; Kagan Huysal
Journal:  J Clin Diagn Res       Date:  2017-02-01

2.  Plasma sodium measurements by direct ion selective methods in laboratory and point of care may not be clinically interchangeable.

Authors:  Brigit A Weld; Thomas J Morgan; Jeffrey J Presneill; Steven Weier; David Cowley
Journal:  J Clin Monit Comput       Date:  2016-09-28       Impact factor: 2.502

3.  Evaluation of the analytical performances of Cobas 6500 and Sysmex UN series automated urinalysis systems with manual microscopic particle counting.

Authors:  Ebubekir Bakan; Zafer Bayraktutan; Nurcan Kilic Baygutalp; Mehmet Ali Gul; Fatma Zuhal Umudum; Nuri Bakan
Journal:  Biochem Med (Zagreb)       Date:  2018-06-15       Impact factor: 2.313

4.  Validation of the blood gas analyzer for pH measurements in IVF culture medium: Prevent suboptimal culture conditions.

Authors:  Juan D N Diaz de Pool; Sjoerd A A Van Den Berg; Gonneke S K Pilgram; Bartholomeus E P B Ballieux; Lucia A J Van Der Westerlaken
Journal:  PLoS One       Date:  2018-11-12       Impact factor: 3.240

5.  Performance Evaluation of the i-SmartCare 10 Analyzer and Method Comparison of Six Point-of-Care Blood Gas Analyzers.

Authors:  Sang-Mi Kim; Hyung-Doo Park
Journal:  Ann Lab Med       Date:  2022-07-01       Impact factor: 4.941

6.  Blood-gas vs. Central-Laboratory analyzers: interchangeability and reference intervals for sodium, potassium, glucose, lactate and hemoglobin.

Authors:  Kocijancic Marija; Kraus Frank Bernhard; Ludwig-Kraus Beatrice
Journal:  Heliyon       Date:  2021-10-30

7.  Pseudohyperkalemia Associated With Leukemia.

Authors:  Biraj Shrestha; Swarup Sharma Rijal; Arpan Pokhrel; Anish Paudel; Krantikiran Baral; Bidhya Poudel; Sijan Basnet; Anthony Donato
Journal:  Cureus       Date:  2022-04-09

8.  A method comparison study of a point-of-care blood gas analyser with a laboratory auto-analyser for the determination of potassium concentrations during hyperkalaemia in patients with kidney disease.

Authors:  Mogamat-Yazied Chothia; Patricia Kassum; Annalise Zemlin
Journal:  Biochem Med (Zagreb)       Date:  2020-08-05       Impact factor: 2.313

9.  Comparison of electrolyte and glucose levels measured by a blood gas analyzer and an automated biochemistry analyzer among hospitalized patients.

Authors:  Huo-Chun Yi; Wen-Sheng Shi; Yin-Hui Zhang; Xiao-Zhen Zhu; Yang Yu; Xin-Xin Wang; Zhang Dai; Yong Lin
Journal:  J Clin Lab Anal       Date:  2020-03-09       Impact factor: 2.352

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.