Literature DB >> 33553552

A comparison of automated urine analyzers cobas 6500, UN 3000-111b and iRICELL 3000 with manual microscopic urinalysis.

Piraya Tantisaranon1, Kanyarat Dumkengkhachornwong1, Peechana Aiadsakun2, Areerat Hnoonual3.   

Abstract

OBJECTIVES: Microscopic examination is essential in urine analysis. This is a simple way to collect informative data but it is also labor-intensive, time-consuming, and requires experienced staff for accurate results and interpretation. Several automated urine analyzers have been introduced for urine analysis in medical laboratories. The aim of this study was to assess and compare the performance of the most common three automated urine analyzers, Cobas 6500, UN3000-111b and iRICELL 3000.
DESIGN: and
Methods: A total of 100 routine urine samples were used in the study. Results from the three machines were compared with the routine procedure results including physical, chemical and sediment analysis.
RESULTS: There was good correlation of urine physical and chemical analyses between the three analyzers with an overall concordance level of more than 80%. For sediment analysis, the degree of concordance between manual analysis and the three instruments was very good to good for white blood cells, red blood cells and epithelial cells, and moderate for bacteria. There were fair to good agreements between manual microscopy and the three instruments, Cobas 6500, UN3000-111b and iRICELL 3000, for casts (Cohen's kappa 0.42, 0.38 and 0.62, respectively).
CONCLUSIONS: The three automated urine analyzers showed similar performances and good correlation with manual microscopy. The results of this study indicate that automated urine analyzers could be used for initial urine testing to reduce high workloads and to save time, but manual microscopic analysis by experienced staff is still necessary to classify urine sediments for confirmation, especially in pathologic specimens.
© 2021 The Author(s).

Entities:  

Keywords:  Automated urine analyzer; Microscopic; Urinalysis; Urine sediment

Year:  2021        PMID: 33553552      PMCID: PMC7847969          DOI: 10.1016/j.plabm.2021.e00203

Source DB:  PubMed          Journal:  Pract Lab Med        ISSN: 2352-5517


  13 in total

1.  European urinalysis guidelines.

Authors: 
Journal:  Scand J Clin Lab Invest Suppl       Date:  2000

Review 2.  Advances in Urine Microscopy.

Authors:  Gavin J Becker; Giuseppe Garigali; Giovanni B Fogazzi
Journal:  Am J Kidney Dis       Date:  2016-01-22       Impact factor: 8.860

3.  Preliminary evaluation of the Iris IQ 200 automated urine analyser.

Authors:  Lia Alves; Frederic Ballester; Jordi Camps; Jorge Joven
Journal:  Clin Chem Lab Med       Date:  2005       Impact factor: 3.694

4.  Urine sediment examination: a comparison of automated urinalysis systems and manual microscopy.

Authors:  Tzu-I Chien; Jau-Tsuen Kao; Hui-Lan Liu; Po-Chang Lin; Jhih-Sian Hong; Han-Peng Hsieh; Miao-Ju Chien
Journal:  Clin Chim Acta       Date:  2007-05-26       Impact factor: 3.786

5.  Comparison of LabUMat-with-UriSed and iQ200 fully automatic urine sediment analysers with manual urine analysis.

Authors:  Okhan Kadir Akin; Muhittin A Serdar; Zeynep Cizmeci; Ozlem Genc; Suleyman Aydin
Journal:  Biotechnol Appl Biochem       Date:  2009-06       Impact factor: 2.431

6.  Comparison of three automated systems for urine chemistry and sediment analysis in routine laboratory practice.

Authors:  Yasemin U Budak; Kağan Huysal
Journal:  Clin Lab       Date:  2011       Impact factor: 1.138

7.  Performance Evaluation and Comparison of the Fully Automated Urinalysis Analyzers UX-2000 and Cobas 6500.

Authors:  Bongkot Wesarachkitti; Varanya Khejonnit; Busadee Pratumvinit; Kanit Reesukumal; Suriya Meepanya; Chanutchaya Pattanavin; Preechaya Wongkrajang
Journal:  Lab Med       Date:  2016-05

8.  Cutoff values for bacteria and leukocytes for urine sediment analyzer FUS200 in culture-positive urinary-tract infections.

Authors:  Derya Kocer; Fatma M Sarıguzel; Cıgdem Karakukcu
Journal:  Scand J Clin Lab Invest       Date:  2014-04-03       Impact factor: 1.713

9.  Comparison of Cobas 6500 and Iris IQ200 fully-automated urine analyzers to manual urine microscopy.

Authors:  Ebubekir Bakan; Nurinnisa Ozturk; Nurcan Kilic Baygutalp; Elif Polat; Kadriye Akpinar; Emrullah Dorman; Harun Polat; Nuri Bakan
Journal:  Biochem Med (Zagreb)       Date:  2016-10-15       Impact factor: 2.313

10.  The comparison of automated urine analyzers with manual microscopic examination for urinalysis automated urine analyzers and manual urinalysis.

Authors:  Fatma Demet İnce; Hamit Yaşar Ellidağ; Mehmet Koseoğlu; Neşe Şimşek; Hülya Yalçın; Mustafa Osman Zengin
Journal:  Pract Lab Med       Date:  2016-03-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.